Technical Data
CompareORP / Redox
| Measuring Principle | Sensor ORP / Redox |
| Application | Hygiene and sterile applications (sterilizable, autoclavable): • Fermenter • Biotechnology • Pharma |
| Characteristic | Digital ORP electrode for hygienic production processes with ion trap for long-term stable reference |
| Measuring range | –1 500 to 1 500 mV |
| Measuring principle | Gel compact electrode with ceramic junction and ion trap Measuring element platinum cap |
| Design | All shaft lengths with temperature sensor Advanced gel technology |
| Material | Sensor shaft: Glass to suit process Metal lead: Ag/AgCl Open aperture: Ceramic junction, zirconium dioxide ORP measuring element: Platinum O-ring: FKM Process coupling: PPS fiber-glass reinforced Nameplate: Ceramic metal oxide |
| Dimension | Diameter: 12 mm (0.46 inch) Shaft lengths: 120, 225 and 360 mm (4.68, 8.77 and 14.04 inch) |
| Process temperature | 0 to 100 °C (32 to 212 °F) 0 to 140 °C (32 to 284 °F) (140 °C (284 °F) or sterilization only) |
| Process pressure | 0.8 to 14 bar (11.6 to 203.1 psi) absolute |
| Temperature sensor | NTC 30k |
| Ex certification | With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in hazardous areas Zone 0, Zone 1 and Zone 2 |
| Connection | Inductive, digital connection head with Memosens 2.0 technology |
| Ingress protection | IP68 |
Field of application
Memosens CPS62E is the digital specialist for demanding processes and hygienic applications. Its contamination-resistant gel guarantees stable measurement even in flowing media, in low conductivities and at high temperatures. Thanks to Memosens 2.0 digital technology, CPS62E combines maximum process integrity with simple operation. It resists moisture and enables lab calibration. It offers extended storage of calibration and process data providing the perfect basis for predictive maintenance.
Process technology and monitoring of processes with:
- Quickly changing ORP values
- High proportion of electrode poisons such as H2S
Chemical processes
Hygienic and sterile applications:
- Fermenters
- Biotechnology
- Life sciences
With ATEX, IECEx, CSA C/US, NEPSI, Japan, INMETRO approvals for use in hazardous areas, zones 0, 1 and 2.
Benefits
Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and providing a future-proof basis for predictive maintenance and enhanced IIoT services.
Sensor resists strong acids and bases and high organic loads.
Maintains high accuracy even after CIP/SIP or autoclaving.
Free from animal-based materials and from acrylamide.
Upside-down mounting option allows for flexible installation.
Maximum process safety is ensured by non-contact, inductive signal transmission.
Minimized process downtime and extended sensor lifetime result in reduced operating costs.
ORP / Redox:
| Measuring Principle | Sensor ORP / Redox | |
| Application | Hygiene and sterile applications (sterilizable, autoclavable): • Fermenter • Biotechnology • Pharma |
|
| Characteristic | Digital ORP electrode for hygienic production processes with ion trap for long-term stable reference | |
| Measuring range | –1 500 to 1 500 mV | |
| Measuring principle | Gel compact electrode with ceramic junction and ion trap Measuring element platinum cap |
|
| Design | All shaft lengths with temperature sensor Advanced gel technology |
|
| Material | Sensor shaft: Glass to suit process Metal lead: Ag/AgCl Open aperture: Ceramic junction, zirconium dioxide ORP measuring element: Platinum O-ring: FKM Process coupling: PPS fiber-glass reinforced Nameplate: Ceramic metal oxide |
|
| Dimension | Diameter: 12 mm (0.46 inch) Shaft lengths: 120, 225 and 360 mm (4.68, 8.77 and 14.04 inch) |
|
| Process temperature | 0 to 100 °C (32 to 212 °F) 0 to 140 °C (32 to 284 °F) (140 °C (284 °F) or sterilization only) |
|
| Process pressure | 0.8 to 7 bar (11.6 to 101.5 psi) absolute | |
| Temperature sensor | NTC 30k | |
| Ex certification | With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in hazardous areas Zone 0, Zone 1 and Zone 2 | |
| Connection | Inductive, digital connection head with Memosens 2.0 technology | |
| Ingress protection | IP68 | |
Related Models
| Model | Name |
|---|---|
| CPS62E-AA7PTU2 | ORP电极 |
| CPS62E-NA7PTU2 | ORP电极 |
Case Studies
市政污水处理污水处理厂 · ORP应用案例
项目背景
市政污水处理客户在污水处理厂环节,长期面临「出水水质监测精度不足、仪表易受污染导致数据漂移,影响达标排放」的困扰,对ORP的稳定性与测量精度提出了更高要求。
解决方案
阿仪科技技术团队结合现场工况,为客户选配了适配的ORP,并完成选型、安装、调试与整定的一站式交付。
实施效果
- 测量精度与响应速度显著提升,数据稳定可靠;
- 仪表适应市政污水处理复杂工况,维护量大幅降低;
- 助力客户实现连续在线监测与降本增效,获得客户高度认可。
食品饮料食品生产线 · ORP应用案例
项目背景
食品饮料客户在食品生产线环节,长期面临「卫生级测量要求高,在线仪表需耐 CIP/SIP 清洗且精度稳定」的困扰,对ORP的稳定性与测量精度提出了更高要求。
解决方案
阿仪科技技术团队结合现场工况,为客户选配了适配的ORP,并完成选型、安装、调试与整定的一站式交付。
实施效果
- 测量精度与响应速度显著提升,数据稳定可靠;
- 仪表适应食品饮料复杂工况,维护量大幅降低;
- 助力客户实现连续在线监测与降本增效,获得客户高度认可。
化工制药反应釜与储罐 · ORP应用案例
项目背景
化工制药客户在反应釜与储罐环节,长期面临「介质腐蚀性强、工况复杂,仪表需耐腐蚀且防爆」的困扰,对ORP的稳定性与测量精度提出了更高要求。
解决方案
阿仪科技技术团队结合现场工况,为客户选配了适配的ORP,并完成选型、安装、调试与整定的一站式交付。
实施效果
- 测量精度与响应速度显著提升,数据稳定可靠;
- 仪表适应化工制药复杂工况,维护量大幅降低;
- 助力客户实现连续在线监测与降本增效,获得客户高度认可。
电力能源电厂锅炉与汽机 · ORP应用案例
项目背景
电力能源客户在电厂锅炉与汽机环节,长期面临「高温高压工况下测量不稳定,维护停机成本高」的困扰,对ORP的稳定性与测量精度提出了更高要求。
解决方案
阿仪科技技术团队结合现场工况,为客户选配了适配的ORP,并完成选型、安装、调试与整定的一站式交付。
实施效果
- 测量精度与响应速度显著提升,数据稳定可靠;
- 仪表适应电力能源复杂工况,维护量大幅降低;
- 助力客户实现连续在线监测与降本增效,获得客户高度认可。