Electromagnetic Flowmeter(Promag 10)

[Function Key Description]]

1, the operation interface has three buttons, namely:"""-"and"E";

2,"","-"can switch menus up and down, or set values in the value input interface (when there is a shadow in the number input interface, press one""or"-", the shadow disappears and a single cursor appears to modify the current value, such as a number.Word"1"Press One"", become a numberWord"2");"E"is the confirmation key, used to enter the corresponding menu or confirm the value (if you need to enter multiple digits,such"178", lose first.in1,Econfirmationin"7", again.Eenter third place to losein"8", again.Efor confirmation).3. Press simultaneously""and"-"is the return key. If you accidentally enter the wrong menu during the setting process, press it several times to return to the measurement main interface, and then re-enter the correct menu for setting.

[unit setting]]

1. After the instrument is powered on, it enters the measurement mode through the startup procedure,"E"to enter the function menu:

2,"E"confirmation, setting system units,""or"-"Prompt for Passwordcode"10","E"confirmation, and then""or"-"modify the instantaneous volumetric flow unit,"E

img1

"Confirmation:

3,"E"go to the next setting,""or"-"modify the cumulant unit, after modification"E

img2

"Confirmation;4, multiple times at the same time"

img3

"and"-"exit to the function menu:

[range setting]]

1, by function menu""or"-""CURRENT OUTPUT":

2,"E"to displayShown"VALUE 20mA"namely20mAthe corresponding value,""or"-"to modify (the value needs to be consistent with the upper computer settings), after"E

img4

"Confirmation:

3, simultaneously"+

img5img6img7

"and"-"exit to the function menu:

[Process parameter setting] (small signal removal and empty pipe detection function)

1, by function menu"+"or"-"to showShown"PROCESS PARAMETER"

2,"E"Enter the process parameter setting,"+"or"-"Setfixed"LOW FLOW CUTF"value (I. e. small signal cut-off value), it is recommended that the small signal cut-off value is the range.1%~5%:

3,"E"to the nextItem"EPD"(ATC detection function),"+"or"-"will"EPD"Item Selection"ON","E

img8

"determination:

4, multiple times at the same time"+"and"-"to exit to measurement mode:

img9img10img11

After the above settings, the basic setting of the instrument is completed.