Familia de producto: CL10sc Amperometric Chlorine Analyzer and Controller
CLF10 sc Free Chlorine Sensor, SC200 Controller and Stainless Steel Panel with pHD Differential Sensor, 24VDCDescontinuado
HACH 2987500
Hach’s CLF10 sc and CLT10 sc provide online amperometric solutions for real-time analysis of free or total chlorine in disinfection applications.

产品概述

Hach’s CLF10 sc and CLT10 sc provide online amperometric solutions for real-time analysis of free or total chlorine in disinfection applications.

The CLF10 sc Free Chlorine Analyzer and CLT10 sc Total Chlorine Analyzer measure free or total chlorine continuously for real time process control. It is based on amperometric technology and is fully compatible with all of Hach's digital controllers. The CLF10 sc can be used in most municipal and industrial applications, and is best utilized where waste stream management is a constraint.

主要特点

  • EPA-approved Free or Total Chlorine Analysis:The CLF10 sc and CLT10 sc chlorine analyzers are compliant with EPA Method 334.0, ensuring that you not only have precise process control, but reportable results as well.
  • No Reagents to Replace, No Waste Stream:Hach’s amperometric chlorine technology does not use reagents, eliminating the need for a sanitary drain and saving you the cost of monthly reagent replacement.
  • Chlorine Analysis Made Easy:The CLF10 sc and CLT10 sc are plug-and-play with Hach’s digital controllers, ensuring easy setup. Available with exclusive self-diagnostics, these instruments will alert the user to changes in the process or upcoming maintenance. Hach’s Cal Watch feature will provide a warning for pH or chlorine calibration changes, and a non-contacting flow sensor will notify users of insufficient sample flow.
  • :
  • :

技术参数

参数名称说明
AccuracyFree Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2Free Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5
Cable Connection5 pin, M12 connector
Cable Length1 m (between gateway and sc controller)
Calibration Method1-point or 2-point (zero and slope) calibration
CertificationsCE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits) Additional certifications when connected to an sc controller
Connection Drain LinepH Flow Cell Outlet: 1/2-inch ID
Connection Sample Line1/4 inch OD
ControllerSingle Digital
Controller Compatibilitysc controller platform
Controller OptionsSingle Input, 24 VDC
DescriptionCLF10sc Free Chlorine Sensor, SC200 Controller and Stainless Steel Panel with pHD Differential Sensor, 24VDC
DimensionsSensor Only: 195 |mm x 25 mm (7.68 in x 0.98 in)
Fitting TypeUSA
Fitting TypeUSA
InterferencesFree chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits. Total chlorine: chlorine dioxide, ozone and chalk deposits
LengthSensor: 195 mm
Lower Limit of Detection (LOD)30 ppb (0.03 ppm)
MaterialCorrosion-resistant materials, fully-submersible (stainless steel, PEEK, PVC, silicon rubber and polycarbonate)
Measurement MethodReagentless, electrochemical, three-electrode amperometric system
MountingPanel or Wall Mounting
Operating Temperature Range0 - 45 °C
OptionsSC200 Controller, Stainless Steel Panel, & pHD Differential Sensor
ParameterFree Chlorine
pH Monitoring Required?with pHD Differentinal
pH Range4 - 9
Power Requirements (Voltage)24 VDC
Pressure Range0.5 bar, no pressure impulses and/or vibrations
Range0 - 20 ppm Cl₂
Repeatability30 ppb or 3%, whichever is greater
Response TimeTotal Chlorine: 100 s or less for 90% change (T90) at a stable temperature and pH
Response TimeFree Chlorine: 140 s or less for 90% change (T90) at a stable temperature and pH
Sample Flow Rate30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature5 - 45°C (41 - 113°F)
Storage Conditions-20 °C to 60 °C dry
Warranty12 months
Weight12.61 kg Approximately 5.5 kg (12 lbs) Panel and empty panel-mounted components only
What's included?SC200 Controller, Mounting Hardware, SC200 Basic User Manual, CL10sc Panel, 1 m digital extension cable, Panel Manual, pHD Sensor Manual, Chlorine Sensor Manual.

同系列其他配置

型号
2981200CLF10 sc Free Chlorine Sensor, SC200 Controller and Stainless Steel Pa
2981800CLT10 sc Total Chlorine Sensor, SC200 Controller and Stainless Steel P
2982400CLF10 sc Free Chlorine Analyzer with SC200 Controller and pHD Differen
2983000CLT10 sc Total Chlorine Analyzer with SC200 Controller and pHD Differe
2981300CLF10 sc with SC200 Signal Input, Combination pH Sensor (Metric)
2981400CLF10 sc Free Chlorine Sensor, SC200 Controller, and Stainless Steel P
2981700CLT10 sc Total Chlorine Sensor, SC200 Controller, and Stainless Steel
2981900CLT10 sc with SC200 Signal Input, Combination pH Sensor (Metric)
2982000CLT10 sc Total Chlorine Sensor, SC200 Controller, and Stainless Steel
2982500CLF10sc Free Chlorine Analyzer with sc200 Controller and Combination p
2982600CLF10 sc Free Chlorine Analyzer with SC200 Controller (Metric)

资料下载

12 modelos · clic para contraer