Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

2987900Descontinuado

2987900
CLF10 sc sensor de cloro residual libre, controlador sc200 y panel de acero inoxidable, solo muestreo puntual, 24 V CCDescontinuado
HACH 2987900
Hach's answer to reagentless amperometric chlorine measurement.

产品概述

Hach's answer to reagentless amperometric chlorine measurement.

The CLF10 sc analyzer uses Hach's exclusive self diagnostics to alert users when the process has changed or the instrument needs servicing. Diagnostic features include the Cal Watch algorithm for warning of pH and chlorine calibration deviation and a non-contacting flow sensor for notification of insufficient sample flow. Free chlorine measurement with an amperometric analyzer, such as the CLF10 sc, does not require reagents, eliminating the need for routine reagent replacement and waste stream management. The CLF10sc free chlorine analyzer allows for real-time control of disinfection processes by providing continuous readings that indicate when treatment conditions have changed. Product includes free chlorine sensor, chlorine sensor block, sc200 controller, sc Gateway, stainless steel panel and pH flow cell for grab samples.

主要特点

  • Compatible with Hach's "Plug and Play" Digital Controllers:The CLF10 sc and CLT10 sc analyzers leverage Hach's exclusive self diagnostics to alert users when the process has changed or the instrument needs servicing. Diagnostic features include the Cal Watch algorithm for warning of pH and chlorine calibration deviation and a non-contacting flow sensor for notification of insufficient sample flow.
  • EPA compliant according to Method 334.0:Chlorine measurement with an amperometric analyzer, such as the CLF10 sc or CLT10 sc, does not require reagents, eliminating the need for routine reagent replacement and waste stream management.
  • Exclusive Self Diagnostics:The CLF10 sc and CLT10 sc analyzers allow for real-time control of disinfection processes by providing continuous readings that indicate when treatment conditions have changed.
  • No Reagent Replacement, No Waste Stream:The CLF10 sc and CLT10 sc analyzers can be used with any Hach sc digital controller. Whether you're measuring turbidity or chlorine, you only need to learn one controller for all your water analysis measurement points. Hach sc controllers have no complicated wiring or setup procedures. Just plug in any Hach digital sensor and it's ready to use without software configuration.
  • Real-time Process Control:In accordance with EPA Method 334.0, the CLF10 sc and CLT10 sc analyzers can be used for reporting chlorine residual measurements. Additionally, Hach has created a suite of laboratory products and methods to help with startup and quality control procedures required in Method 334.0. (See page 4 for a partial listing of accessories.)

技术参数

参数名称说明
AccuracyFree Chlorine: ±3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2Free Chlorine: ±10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3Total Chlorine: ±10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4Total Chlorine: ±20 % of the reference test** (DPD) at stable pH greater than 8.5
Cable Connection5 pin, M12 connector
Cable Length1 m (between gateway and sc controller)
Calibration Method(zero and slope) calibration
Calibration Method 2(zero and slope) calibration
CertificationsCE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits) Additional certifications when connected to an sc controller
Connection Drain LinepH Flow Cell Outlet: 1/2-inch ID
Connection Sample Line1/4 inch OD
Controller Compatibilitysc controller platform
DimensionsSensor Only: 195 |mm x 25 mm (7.68 in x 0.98 in)
InterferencesMonochloramine, chlorine dioxide, ozone, and chalk deposits
LengthSensor: 195 mm
Lower Limit of Detection (LOD)30 ppb (0.03 ppm)
MaterialCorrosion-resistant materials, fully-submersible (stainless steel, PEEK, PVC, silicon rubber and polycarbonate)
Measurement MethodReagentless, electrochemical, three-electrode amperometric system
MountingPanel or Wall Mounting
Operating Temperature Range0 - 45 °C
pH Range4 - 9
Pressure Range0.5 bar, no pressure impulses and/or vibrations
Range0 - 20 ppm Cl₂
Response TimeTotal Chlorine:100 sor less for 90% change (T90) at a stable temperature and pH
Response TimeFree Chlorine: 140 s or less for 90% change (T90) at a stable temperature and pH
Sample Flow Rate30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature5 - 45°C(41 - 113°F)
Ship Wt. (lbs)9.1
Warranty1 Year
Weight5.5 kg Approximately 5.5 kg (12 lbs) Panel and empty panel-mounted components only
What's included?SC200 Controller, Mounting Hardware, SC200 Basic User Manual, CL10sc Panel, 1 m digital extension cable, Panel Manual, Chlorine Sensor Manual.