Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

31291TCDescontinuado

31291TC
Orbisphere Hydrogen (H₂) TC Sensor with argon purge, protection cap, ext. temp. sensor (20 bar)Descontinuado
HACH 31291TC
Allows you to monitor hydrogen in the power production process.

产品概述

Allows you to monitor hydrogen in the power production process.

Standard Hydrogen TC sensor with argon purge and protection cap. With with external temperature sensor adapter. Sensor cable to be ordered: 32505.03. Maximum pressure: 20 bar

This sensor is the successor model for 31261, 31261GP, and 31251.

主要特点

  • Unique thermal conductivity (TC) sensor design:
  • Selective and precise measurement of hydrogen in gaseous or liquid samples:
  • No specific sample preparation required, the gas concentration is measured directly in the sample:
  • Rugged construction can handle harsh plant conditions and high sample pressures:
  • Flexibility of installation for better process control and very low maintenance:

技术参数

参数名称说明
Accuracy(Sample temp. 20 - 50°C within ± 5°C of calibration temperature) The greater of Membrane 29561A: ±1% of reading, or ±2 ppb, or ±0.03 cc/kg, or ±1.5 bar Membrane 2952A: ±1% of reading, or ±8 ppb, or ±0.1 cc/kg, or ±6 mbar Membrane 2935A: ±1% of reading, or ±25 ppb, or ±0.4 cc/kg, or ±20 mbar
ApplicationPower
Purge connectionArgon Purge
RangeRange at 25 °C Membrane 29561A: 0 - 2 ppm, or 0 - 25 cc/kg, or 0 - 1.5 bar Membrane 2952A: 0 - 10 ppm, or 0 - 120 cc/kg, or 0 - 6 bar Membrane 2935A: 0 - 20 ppm, or 0 - 220 cc/kg, or 0 - 12 bar
Sample Pressure0 - 20 bar
Temperature Range-5 - 50 °C
Warranty6 months
Weight0.95 kg
What's included?Sensor only. Controller, sensor cable, flow-chamber or process connection must be ordered separately.