5500.KTO.PH.A2U
Familia de producto: 5500sc Phosphate Analyzers
5500sc Phosphate High Range Analyzer, 2 Channels, Reagents Included
HACH 5500.KTO.PH.A2U
The 5500sc Phosphate Analyzer needs only two liters of reagent to perform unattended for up to 90 days; twice as long as the Series 5000. The industry's only pressurized reagent delivery system elimin

产品概述

Lower Maintenance, Less Downtime

The 5500sc Phosphate Analyzer needs only two liters of reagent to perform unattended for up to 90 days; twice as long as the Series 5000. The industry's only pressurized reagent delivery system eliminates the frequent maintenance associated with pumps. Prognosys which is a predictive diagnostic system, allows you to be proactive in your maintenance. Warning LEDs, and high-visibility notification screens let you avoid unplanned downtime. No more dripping reagents on the instrument, the floor, or your clothing while fumbling with tubes and straws. Simply match the color-coded cap to the sealed reagent bottle and twist gently. Grab Sample In and Grab Sample Out features allow quick analysis of a grab sample poured into the analyzer, and facilitate taking a sample out of the online silica analyzer to verify in a lab test.

主要特点

  • 90 Days of Continuous Run Time:Two liters of reagent are required for the silica analyzer to perform unattended for up to 90 days; twice as long as the Series 5000.
  • Save Time on Maintenance:The industry's only pressurized reagent delivery system eliminates the frequent maintenance associated with pumps.
  • Avoid Downtime:Prognosys, the predictive diagnostic system, allows you to be proactive in your maintenance. Warning LEDs, and high-visibility notification screens let you avoid unplanned downtime.
  • Clean, Fast and Easy Reagent Change:No more dripping reagents on the instrument, the floor, or your clothing while fumbling with tubes and straws. Simply match the color-coded cap to the sealed reagent bottle and twist gently.
  • Verify Easily with Hach Lab Products so you don't Waste Time Second-guessing:Grab Sample In and Grab Sample Out features allow quick analysis of a grab sample poured into the analyzer, and facilitate taking a sample out of the analyzer to verify in a lab test.

技术参数

参数名称说明
AccuracyLR model: ±4 µg/L or ±4% of reading, whichever is greater; HR model: ±500 µg/L or ±5% of reading, whichever is greater
AlarmFour relays output; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum Connection: 18 to 14 AWG wire, 18 AWG stranded recommended
ApplicationPure water / Power
CertificationsCE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000 ) KC (EN 61326-1: 2006) C-tick (EN 61326-1: 2006) cETLus (UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012)
Controller CompatibilitySC200, SC1000
Enclosure RatingIP56 / NEMA 4X
Fitting TypeSample line and sample bypass drain: 6 mm (¼-in.) Air purge inlet: 6 mm (¼-in.) Chemical and case drains: 9.5 mm (3/8-in.)
Grab SampleGrab Sample In and Grab Sample Out capability
Light SourceClass 1M LED (light emitting diode) with a peak wavelength of 810 nm
Lower Limit of Detection (LOD)LR model: 4 µg/L HR model: 200 µg/L
Manual LanguagesEnglish French Spanish B. Portuguese Chinese Japanese Korean Thai German Italian C. Portuguese Czech Danish Dutch Polish Swedish Finnish Bulgarian Hungarian Romanian Lithuanian Russian Turkish Slovak Slovenian Croatian Greek Estonian
Measuring PrincipleColorimetric
MountingWall, panel or table
Number of Channels2
Operating Humidity5 - 95 % non-condensing (indoor use only)
Operating Temperature Range5 - 45 °C
Outputs4 - 20 mA
ParameterHigh Range Phosphate
Pollution Degree2/II
Power Requirements (Hz)50/60 Hz
Power Requirements (Voltage)100 - 240 VAC
Protection ClassI
Range200 - 50000 µg/L as PO₄
Reagent Consumption2 L of each reagent every 90 days with 15 minute cycle time
RelaysFour; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum
RepeatabilityLR model: ±1% HR model: ±500 µg/L or ±5% of reading, whichever is greater
Response TimeTypically, 9.5 minutes at 25 °C (77 °F); changes with temperature
Sample Flow Rate55 - 300 mL/min
Sample Pressure2 - 87 psi (to Preset Pressure Regulator)
Sample Temperature5 - 50°C (41 - 122°F)
Storage Conditions-20 - 60 °C
Warranty12 months
Weight20 kg (45 lb) without reagents and standards, 36.3 kg (80 lb) with reagents
What's included?List of content: 5500sc analyzer, reagents set for up to 90 days of use, installation kit, Hach user manuals

应用领域

Boiler Water · Steam Cycle · Water Cycle

同系列其他配置

型号
5500.KTO.PL.A2U5500sc Phosphate Low Range Analyzer, 2 Channels, Reagents Included
5500.KTO.PL.A1U5500sc Phosphate Low Range Analyzer, 1 Channel, Reagents Included
5500.KTO.PL.A4U5500sc Phosphate Low Range Analyzer, 4 Channels, Reagents Included
5500.KTO.PH.A1U5500sc Phosphate High Range Analyzer, 1 Channel, Reagents Included
5500.KTO.PH.A4U5500sc Phosphate High Range Analyzer, 4 Channels, Reagents Included
5500.KTO.PH.D1U5500sc Phosphate High Range Analyzer, DC Powered, 1 Channel, Reagents
5500.KTO.PL.D1U5500sc Phosphate Low Range Analyzer, DC Powered, 1 Channel, Reagents I
5500.KTO.PL.D2U5500sc Phosphate Low Range Analyzer, DC Powered, 2 Channels, Reagents
5500.KTO.PL.D4U5500sc Phosphate Low Range Analyzer, DC Powered, 4 Channels, Reagents
5500.KTO.PH.D2U5500sc Phosphate High Range Analyzer, DC Powered, 2 Channels, Reagents
5500.KTO.PH.D4U5500sc Phosphate High Range Analyzer, DC Powered, 4 Channels, Reagents