Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

DPMP-CLPHD

DPMP-CLPHD
Dual Parameter Monitoring Panel with SC4500, CL17sc and digital pH Sensor
HACH DPMP-CLPHD
The Dual Parameter Chlorine panels offers CL17sc chlorine analyzer, Digital pH sensor and SC4500 controller in the convenience of one panel. These panel configurations are pre-engineered, ready to ins

产品概述

Expertise + Simplicity

The Dual Parameter Chlorine panels offers CL17sc chlorine analyzer, Digital pH sensor and SC4500 controller in the convenience of one panel. These panel configurations are pre-engineered, ready to install and simple solutions that are designed to work together.

主要特点

  • Quality from start to finish​:Our panel configurations allow you to maximize the inherent quality of the instruments, providing peace of mind while leveraging Hach’s 100 years of water quality innovations.
  • Up and running in no time:Our preassembled, ready-to-use solutions streamline the process between purchasing and startup.
  • Ease of installation:When you buy a water quality monitoring station from Hach, you’ll receive an expertly designed, preassembled panel that makes installation straightforward and user friendly.

技术参数

参数名称说明
Accuracy± 5% or ± 0.04 mg/L (whichever is greater) from 0 - 5 mg/L Cl₂ ± 10% from 5 - 10 mg/L Cl₂
Air PurgeOptional with 3/8-inch quick-connect fitting and tubing; 0.003 m3/min at 1.38 bar (20 psig) maximum
AlertsLow and high chlorine. Dirty cell. Low and high sample flow.
CertificationsSC4500 Controller is CE compliant and is TÜV listed to UL and CSA safety standards
ControllerSC4500
Controller CompatibilitySC200, SC1000, SC4500. All controllers sold separately
Cycle Time2.5 minutes
Dimensions13" x 28" x 8.6 "(WxHxD)
Dimensions (H x W x D)711 mm x 330 mm x 218 mm
Drain½-inch ID flexible hose
Enclosure RatingIP66
Inlet¼-inch OD polyethylene tube, quick-disconnect fitting
Inlet Pressure0.3 - 5.2 bar (4.5 - 75 psig) supplied to Y-strainer 0.1 - 0.3 bar (1.5 - 5 psig) supplied to analyzer
Installation KitPressure Regulator Installation Kit
InterferencesOther oxidizing agents such as bromide, chlorine dioxide, permanganate, and ozone will cause a positive interference. Hardness must not exceed 1,000 mg/L CaCO₃.
Light SourceLED, measurement at 510 nm, 1 cm light pathlength
Lower Limit of Detection (LOD)0.03 mg/L
MountingWall or rack
Operating Humidity0 - 90% relative humidity, non-condensing
Operating Temperature Range41 - 104 °F (5 - 40 °C)
PanelYes
ParameterChlorine, pH
Power Requirements (Voltage)100 - 240 VAC ±10%, 50/60 Hz; 1 A
Process ConnectionSample Inlet ¼" OD tube Sample Drain ¼" OD tube
RangeCL17sc 0 -10 mg/L Cl₂ pHD -2.0 - 14.0 pH
Range0 - 10 mg/L free or total residual chlorine
Reagent Consumption0.5 L of buffer solution and 0.5 L of indicator solution in 31 days
Sample Flow Rate600 - 800 mL/min
Sample Pressure4.5 - 75 psi (31 -- 517 kPa)
Sample RequirementsY-strainer filtration with 40-mesh screen or higher
Sample Temperature41 - 104 °F (5 - 40 °C)
SensorsCL17sc pHD
Solution TypeDrinking water, raw and treated
Weight42.5 lb (19.3 kg)
What's included?SC4500 Controller, Claros-enabled, 5x mA Output, 2 digital Sensors, 100-240 VAC, US plug (LXV525.99EA1551) CL17sc Colorimetric Chlorine Analyser with Pressure Regulator Installation Kit, without Reagents (8572400) Digital pH Sensor, PPS, Water Distribution Monitoring Panel Mounting (DPD1R1-WDMP) User manual

应用领域

Municipal Water

资料下载