Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

EZ sc Series Additional Analyzers

EZ sc Series Additional Analyzers
EZ sc Series Additional Analyzers
HACH EZ-sc-Series Add-Analyzers-US
Modelos configurables 3 models
EZ sc Series Additional Analyzers

产品概述

Reliable, Scalable Systems Designed for Peak Performance

The EZ1030sc Phenol analyzer offers a reliable, easy-to-use solution for precise water analysis. Ideal for industrial and wastewater applications, it ensures accurate phenol measurements to boost efficiency, maintain compliance, and protect processes. Designed for easy integration and low maintenance, the EZ1030sc helps you stay ahead of challenges. With the advanced SC4500 Controller, it streamlines operations and gives you superior control over water analysis. Key Features Flexible Customisation The EZ1030sc is designed to adapt seamlessly to the unique needs of your operation. The base model features a single stream with 4x or 8x mA outputs, but it can be customised to handle four or even eight streams for enhanced efficiency. With advanced communication options such as Modbus RTU, Modbus TCP/IP, Profibus DP, and Ethernet/IP, it ensures smooth integration into any system. Reliable and Easy to Use Built for demanding environments, the EZ1030sc Phenol analyzer delivers consistent, high-accuracy results. Its straightforward interface reduces complexity, enabling operators to perform analyses quickly and efficiently. With minimal maintenance requirements, this tool is a dependable partner for uninterrupted operation. Explore Tailored Solutions Every facility faces unique challenges, and the EZ1030sc Phenol analyzer is designed to adapt. Get in touch to discuss your specific needs, and we’ll work with you to customize this solution to optimise your processes, minimise downtime, and meet compliance with confidence.

主要特点

  • Trusted Equipment for Critical Operations:The EZ sc Series combines robust engineering with high-quality materials to ensure reliability 24/7. Designed for long-term performance, EZ sc Series minimizes breakdowns and supports sustained operational success.
  • Scalable Solutions for Growing Needs:Built to evolve alongside your needs, online analysers from the EZ sc Series offer adaptable parameters and easier integration. They ensure your facility remains future proof without requiring expensive upgrades. Easily adjust parameter testing ranges and expand sample streams to accommodate innovation and growth.
  • Real-Time Data for Faster Decisions:The EZ sc Series offers advanced connectivity and real-time monitoring, delivering immediate access to actionable insights. Facilities utilising this technology benefit from improved reaction times, empowering more informed decision-making and quicker responses during critical situations.
  • Simplified Troubleshooting with Expert Support:Built-in diagnostics combined with secure remote data access make the EZ sc Series simplify issue resolution. Supported by a highly responsive team of experts, it ensures uninterrupted functionality and minimises operational downtime. By addressing and identifying issues early, EZ sc Series of online analysers helps prevent costly disruptions and repairs while maintaining system efficiency.
  • Clear Workflows for Predictable Outcomes:Standardized tools and guided workflows ensure consistent implementation and operation. The EZ sc Series minimizes miscommunication, sets clear expectations, and reduces process inconsistencies, driving greater efficiency.

技术参数

参数名称说明
Alarm1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient Temperature50 - 86°F (10 - 30 °C) ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analog OutputsActive 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard: 4, optional: 8
Automatic cleaningYes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
CalibrationAutomatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
CertificationsCE, ETL certified to UL and CSA safety standards, UKCA
Cooling WaterFlow rate approx. 5 L/h; temperature max. 30 °C; pressure max. 0.5 bar
Cycle TimeNo Dilution Default: 10 min. Continuous: 6 min. Internal Dilution Default: 15 min. Continuous: 6 min.
Demineralized WaterInternal dilution option: Rinse water & Dilution water
Digital OutputsProfibus DP or Modbus RTU
Dimensions (H x W x D)688 mm x 460 mm x 340 mm
DrainAtmospheric pressure, vented, min. Ø 32 mm
Earth ConnectionDry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument AirDry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
InterferencesOxidizing agents, e.g., chlorine (Cl₂) and sulfur compounds, e.g., sulfide (S²⁻) interfere. Large quantities of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar interfere.
Lower Limit of Detection (LOD)No Dilution 0.01 - 0.2 mg/L Phenol: 0.01 mg/L 0.025 - 0.5 mg/L Phenol: 0.025 mg/L 0.025 - 1 mg/L Phenol: 0.025 mg/L 0.1 - 2 mg/L Phenol: 0.1 mg/L Internal Dilution 0.5 - 10 mg/L Phenol: 0.5 mg/L 1 - 20 mg/L Phenol: 1 mg/L 2.5 - 50 mg/L Phenol: 2.5 mg/L 5 - 100 mg/L Phenol: 5 mg/L 7.5 - 150 mg/L Phenol: 7.5 mg/L 10 - 200 mg/L Phenol: 10 mg/L
MaterialHinged part: Thermoform ABS, door: PMMA Wall section: Galvanized steel, powder coated
Measurement method4-aminoantipyrine colorimetric method
Measurement MethodColorimetric
Number of sample streamsStandard: 1 Optional accessory for 2, 4, or 8 sample streams
OutputsOptional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP
ParameterPhenol
Power100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA
PrecisionBetter than 2% full scale range for standard test solutions
Protection ClassIP44
RangeNo Dilution 0.01 - 0.2 mg/L Phenol 0.025 - 0.5 mg/L Phenol 0.025 - 1 mg/L Phenol 0.1 - 2 mg/L Phenol Internal Dilution 0.5 - 10 mg/L Phenol 1.0 - 20 mg/L Phenol 2.5 - 50 mg/L Phenol 5.0 - 100 mg/L Phenol 7.5 - 150 mg/L Phenol 10 - 200 mg/L Phenol
Reagent RequirementsKeep between 50 - 86 °F (10 - 30 °C)
Relays5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)
Sample Flow Rate100 - 300 mL/min
Sample PressureBy external overflow vessel (open to atmospheric pressure)
Sample QualityMaximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature50 - 86 °F (10 - 30 °C)
ValidationAutomatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Warranty12 months
WeightMax. 77 lb (35 kg)
What's included?EZ1030sc Phenol Analyzer, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution) and 1 x empty 2.5L Reagent Container with Fittings (for Color Solution)

应用领域

Municipal Water

同系列其他配置

型号
EZ1008.9700140TEZ1008sc Color Colorimetric Analyzer, 1 stream, 4x mA Output
EZ1022.9700140TEZ1022sc Peroxide Colorimetric Analyzer, 1 stream, 4x mA Output

资料下载

3 modelos · clic para contraer