Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

EZ7050.XXXXXXXX

EZ7050.XXXXXXXX
EZ7050 Chemical Oxygen Demand Analyzer (COD, Permanganate Method)
HACH EZ7050.XXXXXXXX
Online, automatic, wet-chemical determination of COD values in waste water and surface water applications

产品概述

Online, automatic, wet-chemical determination of COD values in waste water and surface water applications

Bridging traditional chemistry with modern analytics The EZ7000 Series are wet-chemical COD analyzers bringing new levels of automation, reliability and performance in measuring COD values in waste water and surface water. The superior analytical performance is exemplary of their build quality, thanks to the use of high quality components, state of the art wet chemistry and standard smart software features. Compliance with international standard methods The analytical mainframe of the EZ7000 Series sets new standards in traditional COD analysis thanks to the flexible architecture, while compliance with international standard methods is assured. Prior to analysis, the sample is oxidized by means of a specific acid solution and heat, in accordance with the standard method applied. The EZ7000 Series of Online COD Analyzers are the answer to the needs of those users who require “true” COD values to quantify organic pollution in various water applications: - Wet-chemical COD analysis conform standard methods for dichromate or permanganate destruction - Built-in sample digestion/oxidation unit - Smart automatic features - Control and communication via industrial panel PC - Standard 4 - 20 mA signal output with alarm processing - Communication ports supporting connectivity to Modbus - Multiple stream analysis There are many additional options available. Please contact Hach for more details.

技术参数

参数名称说明
Alarm1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient Temperature10 - 30 °C ± 4 °C deviation (50 - 86 °F ± 7.2 °F deviation) at 5 - 95% relative humidity (non-condensing)
Analog OutputsActive 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
CalibrationAutomatic, 2-point; frequency freely programmable
CertificationsCE compliant / UL certified
Cooling WaterFlow rate approx. 5 L/h; temperature max. 30 °C; pressure max. 0.5 bar
Cycle Time40 minutes, including oxidation time of 30 minutes. Remark: standard method for Cr destruction requires 120 minutes.
Demineralized WaterFor rinsing
Digital OutputsOptional: Modbus (TCP/IP, RS485)
Dimensions (H x W x D)690 mm x 465 mm x 330 mm
DrainAtmospheric pressure, vented, min. Ø 64 mm
Earth ConnectionDry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument AirDry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
InterferencesChloride > 1 g/L, inorganic reducing agents such as nitrites, sulphides, and iron(II) will increase the result, aromatic hydrocarbons and pyridine are not oxidized to any appreciable extent. Some very volatile organic substances may escape the oxidation by evaporation. Straight chain aliphatic compounds are effectively oxidized by the silver sulphate/sulphuric acid solution. Fats, oil, proteins, surfactants and tar.
Lower Limit of Detection (LOD)≤ 1 mg/L
MaterialHinged part: Thermoform ABS, door: PMMA Wall section: Galvanized steel, powder coated
Measurement MethodRedox titration after oxidation by acid-potassium dichromate solution, conform with ISO 6060 method Or redox titration after oxidation by potassium permanganate solution, conform with ISO 8467 and JIS K0806 methods
Number of sample streams1, 2, 4, or 8
ParameterCOD
Power230 VAC, 50/60 Hz 120 VAC, 50/60 Hz Max. power consumption: 440 VA
PrecisionBetter than 5% full scale range for standard test solutions
Protection ClassAnalyzer cabinet: IP44 / Panel PC: IP65
Range1 - 20 mg/L O2
Range1 - 20 mg/L O₂
Reagent RequirementsKeep between 10 - 30 °C (50 - 86 °F)
Sample Flow Rate100 - 300 mL/min
Sample PressureBy external overflow vessel
Sample QualityMaximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature10 - 30 °C (50 - 86 °F)
ValidationAutomatic; frequency freely programmable
Warranty12 months
Weight25 kg (55 lbs.)
What's included?EZ7050 COD Analyzer, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x Combined Pt Electrode, Routine S7 Connection, 1 x empty 2.5L Reagent Container with Fittings (AgNO₃ Solution), 2 x empty 10L Reagent Containers with Fittings (Acid & Oxalate Solution) and 1 x empty 10L Reagent Dark Container with Fittings (KMNO₄ Solution)

应用领域

Municipal Water · Water Reuse