产品概述
Universal self-cleaning filtration for digestate samples - compatible with the EZ7200 Series
About the EZ9100 Series The successful implementation of online analyzers in process control strategies could not have been achieved without the development of a new generation of automatic sampling and sample preconditioning systems. The EZ9100 Series are the result of many years of field experience in combining analyzers with filtration units. Simplified operation by self-cleaning action All preconditioning systems are designed for fully automatic operation and require virtually no human intervention. Basically all systems are incorporated either with a blow-back action by instrument air, or a specific cleaning cycle. This fundamental design principle does not merely allow trouble-free sampling, moreover it contributes to high up-times. The EZ9130 is a filtration system developed specifically for sampling wet type anaerobic digesters to make solid-free samples available for online analysis. Devoid of overly complex engineering, the filtration can handle the difficult properties of sludge and wastewaters charged with high levels of insoluble constituents: - Self-cleaning sample filtration with various pore sizes - Large-bore pneumatic ball valves for sample and drain - Automatic cleaning by solenoid controlled instrument air - PLC or analyzer controlled cleaning frequency
技术参数
| 参数名称 | 说明 |
|---|---|
| Cleaning | Automatic blowback by solenoid controlled instrument air |
| Dimensions (H x W x D) | 1150 mm x 750 mm x 200 mm |
| Drain | Overflow sample D=50 mm; overflow Static Pressure Regulator 3/8" OD |
| Earth Connection | Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |
| Instrument Air | Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air |
| Material | Filter: SS 316L Piping: PVC Manual ball valves: PVC Tubing: Norprene, PFA, PE Panel: weather resistant Trespa |
| Number of sample streams | 1 stream Optional: 1 to 8 streams |
| Pore Size | 500 µm |
| Power | 24 VDC, powered by analyzer |
| Protection Class | IP55 |
| Required Fast Loop | 2 m/s |
| Rinse water pressure | 3/8" BSPF, 4 bar max. |
| Sample Temperature | Max. 65 °C (149 °F) |
| Size | 500 µm |
| Special Features | Heavy Duty (Anasense) |
| Special Remarks | Backflush valve controlled by Analyzer |
| Weight | 18 kg (39.7 lbs.) |
同系列其他配置
| 型号 | |
|---|---|
| EZ9010.XX110A01 | EZ9010 Filtration System, for immersion, pore size 50 µm, 3 m tubing |
| EZ9010.XX310A01 | EZ9010 Filtration System, for immersion, pore size 50 µm, 3 m tubing |
| EZ9020.XX110001 | EZ9020 Filtration System, fast loop, pore size 50 µm |
| EZ9120.XX350031 | EZ9120 Heavy-duty Filtration System for sludge applications, pore size |
| EZ9110.XX310031 | EZ9110 Heavy-duty Filtration System, pore size 50 µm |
| EZ9130.XX330091 | EZ9130 Heavy-duty Filtration System for digestate samples, pore size 2 |
| EZ9200.XX070A01 | EZ9200 Microfiltration System, for immersion, pore size 0.04 µm, 3 m t |
| EZ9250.XX070001 | EZ9250 Microfiltration System, for bypass, pore size 0.04 µm |
资料下载
- EZ Series Analyzer Service
- Introduction to Hach Services
- Care Advantage Flyer : Complete service solution for all your instruments
- Care Advantage Brochure : Complete service solution for all your instruments
- Analyzer Seri EZ Online Solusi Lengkap
- EZ Series Online Analyzers
- Analizadores en continuo de la serie EZ
- Analisadores on-line da série EZ
Recomendados similares
HACHSC1000 relay module for DIN-rail mounting, 4 Relays, max. 240V
HACHNP6000sc Phosphate Analyzers
Stainless Steel Chain Mount Kit
Mercury Thermometer, Celsius, -10 to 200°C
Mercury Thermometer, Fahrenheit, -30 to 120°F
COVER-SERVICE REPLACEMENT
HACHBattery cable, SD900
HACH