产品概述

Regenerateable Encapsulated LCP or PPS Differential pH and ORP Sensors-ideal for aggressive applications.

LCP (Liquid Crystal Polymer), Convertible Mounting Style, 3m Analog Cable, 2-Wire with Built-In Transmitter, General Purpose Glass pH Electrode

主要特点

  • Exceptional Performance with the Differential Electrode Measurement Technique:This field-proven technique uses three electrodes instead of the two normally used in conventional pH sensors. Process and reference electrodes measure the pH differentially with respect to a third ground electrode. The end result is unsurpassed measurement accuracy, reduced reference junction potential, and elimination of sensor ground loops. These sensors provide greater reliability, resulting in less downtime and maintenance.
  • Lower Maintenance Needs with the Replaceable Salt Bridge:The salt bridge creates a barrier to contamination which minimizes the dilution of the internal standard cell solution. The result is lower maintenance needs and a longer time period between calibrations.
  • Built-in Preamp or Two-Wire Transmitter:The built-in preamp produces a strong signal, enabling you to locate the analyzer up to 3000 ft. (914 m) from the sensor. An optional built-in two-wire transmitter is available for applications requiring a 4-20 mA sensor signal. This option requires that the indicating instrument of the measuring system be capable of providing 24 VDC to power the sensor, and have adjustment means to calibrate for zero offset and span.
  • Versatile Mounting Styles:1.5 inch NPT threads are provided on both ends of the convertible mounting style sensor for either mounting into a pipe tee or attaching to the end of a pipe for immersion. The convertible style enables you to consolidate inventory, and thereby reduce associated costs. A union-mount style sensor and mounting tee are also available to conveniently install and remove the sensor for in-line service.
  • 3,000 ft. (914 m) transmission distance:
  • Also available with flat glass electrodes:

技术参数

参数名称说明
Body materialLCP
Cable Length3 m 10 ' 2-conductor (twisted pair)
Diameter55 mm
Drift0.03 pH per 24 hours, non-cumulative
Electrode MaterialGlass
FlowMax. 3 m/s
JunctionPVDF Junction
Length180 mm
MaterialGround Electrode: Titanium or Hastalloy C
Measuring Range ORP±2000 mV
Measuring Range pH0 - 14 pH
MountingConvertible
Pressure RangeMax. 10.7 bar sensor only (pressure range of mounting hardware to be checked separately)
Sensitivity< 0.5 mV
Sensitivity< 0.005 pH
Sensor Thread1.5" NPT at both ends
Special NotespH sensor with built-in two-wire transmitter providing 4-20 mA output
Temperature Range-5 - 95 °C (23 - 203 °F)
Temperature SensorNTC 300 Ω thermistor for automatic temperature compensation and analyzer temperature readout
Warranty12 months
Weight0.6 kg
What's included?Includes: sensor with 10 ft cable and manual

应用领域

Membrane Protection · Nutrient Monitoring · Nutrient Monitoring · Secondary Treatment · Secondary Treatment · Ultrapure Water

同系列其他配置

型号
6458P0pH Sensor, Union-mount, LCP (Liquid Crystal Polymer) Body Material, 2-
6428P010010000NpH Encapsulated Sensor, LCP Body, 2 Wire, Convertible, FKM/FPM Seals,
6428P020pH Encapsulated Sensor, LCP Body, 2 Wire, Convertible, FKM/FPM Seals,
6428P030pH Encapsulated Sensor, LCP Body, 2 Wire, Convertible, FKM/FPM Seals,
6428P050pH Encapsulated Sensor, LCP Body, 2 Wire, Convertible, FKM/FPM Seals,
6428P025pH Encapsulated Sensor, LCP Body, 2 Wire, Convertible, FKM/FPM Seals,

资料下载