Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

LXV322.99.00002

LXV322.99.00002
TSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Level System
HACH LXV322.99.00002
TSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Blanket Level Instrument

产品概述

TSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Blanket Level Instrument

Compact instrument for analytical determination by continuous measurement mode with integration time setting. Adjustment of up to 5 measuring points, storage of measuring values with name, date and time. Shockproof. Combined multiple beam alternating light method with infrared diode system and beam focusing; wavelength 860 nm. The measurement ranges for turbidity and suspended solids are: 0.001 - 9999 FNU and 0.001 - 400 g/L respectively.

主要特点

  • Three Parameters in One Instrument:The Hach TSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Blanket Level Instrument's unique multi-beam alternating light method with infrared diode system gives it a broad measuring range for both turbidity and suspended solids in one portable handheld instrument.
  • Multiple Calibration Curves for Convenience:It stores up to four different calibration curves for suspended solids and one for turbidity. There's no need for constant re-calibration for different locations or types of sludge.
  • Easy Sludge Blanket Levels:The sludge blanket level can be measured by setting the units to mg/L, g/L or percent. Calibrate the instrument and then slowly lower the sensor into the clarifier. The suspended solids concentration will significantly increase as it reaches the sludge blanket level. Once the blanket level is reached, examine the sensor cable which is marked every meter to determine the sludge blanket level depth from the water surface.
  • Air Bubble Compensation for Accuracy:The instrument software automatically compensates for air bubbles to prevent its influence on the measurement. Turbidity is pre-calibrated with a calibration curve stored in the instrument. For suspended solids, each of the calibration curves can be closely adjusted to the laboratory results because each calibration curve can use up to three points per curve plus a zero value.
  • Durable Materials for Long Life:The probe is made of polished stainless steel with a scratch-resistant sapphire window to withstand harsh environments. Polished stainless steel helps keep particles from sticking to the surface of the probe. The internal memory stores up to 290 measuring values, including time, date, measurement, location, and homogeneity. Three operating modes are available: single, interval, or continuous.
  • Rechargeable Battery Power:With rechargeable batteries, the Hach TSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Blanket Level Instrument can provide up to one month of usage in typical measuring mode and up to four hours of continuous operation. Interval mode has automatic shut off between measurements for longer operation in the field. The battery status is always visible. The operating menu is user-friendly and is displayed on an easy-to-read LCD graphic display.

技术参数

参数名称说明
Cable Length10 m (32.8 ft.) cable with marks for distance
Current InputApprox. 60 mA
Diameter40 mm
DimensionsProbe (Ø X L) 40 mm x 29 cm (1.57 x 11.42 in.)|Meter (H X W X D): 110 x 230 x 40 mm (4.33 x 9.06 x 1.57 in.)|Case L: 450 mm (17.7 in)
DisplayAlphanumerical display, 4 rows of 16 characters
Kit?Yes
Length290 mm
MaterialSensor Body: stainless steel; Probe: stainless steel; Sensor Window: sapphire
Measurement ModesContinuous measurement, integration time settings 0 - 1000 seconds
Measuring PrincipleCombined multiple beam alternating light method with infrared diode system and beam focusing.; Turbidity: 2-channel 90° scattered light measurement in accordance with DIN/EN 27027/ISO 7027; additional measurement value verification through six-channel multiple-angle measurement; Solids: Modified absorption measurement; six-channel multiple-angle measurement;
Measuring Range TSS0.001 - 400 g/L (upper range depending on sample characteristics)
Measuring Range Turbidity0.001 - 9999 FNU
MemoryUp to 290 measuring points
Operating Pressure0 - 10 bar (0 - 145 psi)
Operating Temperature Range0 - 60 °C
PowerAutomatic shut down after 4 minutes (if no button has been pressed and no permanent measurement is activated)
Power SupplyMeter: 6 rechargeable NiMH batteries or 6 "AA" standard batteries
Precision Suspended Solids< 4 % ± 0.001 g/L , whichever is greater
Precision Turbidity< 3 % or ± 0.02 NTU , whichever is greater
Pressure RangeMax. 10 bar
Protection ClassProbe IP68; Control unit IP 55
RangeTSS 0.001 - 400g/L Turbidity 0.001 - 9999 FNU
Relative Humidity0 - 95 %
ResolutionSuspended Solids: 0.01 at 1 - 9.999 g/L
ResolutionSuspended Solids: 0.1 at 10 - 99.9 g/L
ResolutionSuspended Solids: 1 at greater than 100 g/L
Resolution turbidity0.01 @ 1 - 9.99 NTU
Resolution turbidity0.1 @ 10 - 99.9 NTU
Resolution turbidity1 at greater than 100 NTU
Resolution turbidity0.001 @ 0 - 0.999 NTU
Sensor Cable10 m
Warranty12 months
WeightProbe: 3.5 lbs (1.6 kg); Meter: 1.2 lbs (0.54 kg)
Wetted MaterialsProbe: stainless steel Sensor Window: sapphire
What's included?Turbidity, Suspended Solids and Sludge Level probe, meter, rechargeable batteries, user manual, carrying case. The battery charger must agree with local and national safety regulations and may not be part of the shipment. If the charger is not part of the shipment, charge the batteries with a charger available in your country. Make sure that the charger agrees the battery specifications.

应用领域

Municipal Water

资料下载