Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

LXV326.99.10002

LXV326.99.10002
TSS Vari sc Turbidity and Suspended Solids Inline probe, Stainless Steel
HACH LXV326.99.10002
Turbidity and suspended solids probe for use in VARIVENT® piping systems

产品概述

Turbidity and suspended solids probe for use in VARIVENT® piping systems

The TSS Vari sc is a digital process probe for the extremely accurate in pipe measurement of turbidity and suspended solids specially designed for use in food and pharmaceutical sectors. This in pipe probe is installed by Varivent access unit and can measure both on-line suspended solids and turbidity in one instrument. The two channel 90° scattered light turbidity method is in accordance with DIN EN ISO 7027. For the determination of suspended solids a scattered light measurement method at 120 ° is integrated. This eight channel measurement system, with an integrated bubble and temperature compensating software, enables the instrument to have a wide measuring range. The TSS Vari sc probes are compatible with the Hach SC200 and SC1000 controller platform. Furthermore, they are made of polished stainless steel with a scratch resistant and easy to clean sapphire window design to withstand harsh environments. Please contact Hach for further information. Request for a quote today!

主要特点

  • Unique Multi-Beam Pulsed Infrared Light System:TSS sc probes have a double optical system with two pulsating infrared LEDs and four receivers. This combined multiple beam alternating light method with beam focusing facilitates accurate color independent measurement of turbidity and suspended solids.
  • Automatic Compensation for Air Bubbles and Temperature Swings for Accuracy:Special software enables the system to recognize gas bubbles or temperature swings for a more accurate suspended solids or turbidity measurement.
  • Standard Compliant and Precise - Ease of Calibration:The turbidity measurement complies with the standard ISO 7027. Calibration is not necessary. If the probe is used to measure the solids content, a one-point calibration suffices. For special applications, curves can be defined using several calibration points.
  • Compatible with Hach SC controller platform:Measurement data is displayed and processed with the help of a digital Hach SC200 or SC1000 controller.
  • Two parameters in one instrument--Suspended Solids and Turbidity:

技术参数

参数名称说明
AccuracyTurbidity up to 1000 NTU: < 5% of the measured value ±0.01 NTU
Ambient Temperature0 - 60 °C (briefly 80 °C)
ApplicationVarivent piping systems
Automatic WiperNo
Cable Length10 m (optional extension cables available)
Cable Length10 m (optional extension cables available)
CalibrationTurbidity (TRB): Factory calibrated Solids (TS): To be calibrated by customer on site Zero point: Permanently calibrated in the factory
Calibration MethodTurbidity (TRB): Factory calibrated Solids (TS): To be calibrated by customer on site Zero point: Permanently calibrated in the factory
Controller CompatibilitySC200, SC1000, SC4500. All controllers sold separately
Diameter40 mm
FlowMax. 3 m/s (the presence of air bubbles affects the measurement)
IncludesTurbidity & Suspended Solids sensor, user manual
Installation styleVarivent Measuring tube
LengthTSS VARI sc 232 mm
Maintenance Interval1 h/month
MaterialBody: Stainless Steel (1.4571, AISI 316Ti, UNS S31635), Optics: 1.4404, Gasket: FKM
Max Temperature80
Measurement MethodCombined multiple beam alternating light method with infrared diode system and beam focusing. Turbidity (TRB): 2-channel 90º scattered light measurement in accordance with DIN EN ISO 7027, wavelength = 860 nm Solids (TSS): 90º and 120º scattered light measurement, eight channel multi-angle measurement system, with an integrated bubble and temperature compensating, wavelength = 860 nm
ModelTSS Vari sc
Mounting ConfigurationsVARIVENT
Operating Temperature Range0 - 80 °C
ParameterTurbidity, Suspended Solids
Pressure Range<16 bar or <160 m
RangeTurbidity (TRB): 0.001 - 9999 NTU Solids (TSS): 0.001 - 500 g/L
RepeatabilityTSS content: < 4 % Turbidity: < 3 %
Response Time1 s < T90 < 300 s (adjustable)
Response Time T901 - 300 s adjustable
Special NotesInstallation note: Distance sensor–wall > 50 cm (Turbidity) > 10 cm (TSS)
Weightapprox. 1.5 kg
What's included?Turbidity & Suspended Solids sensor, user manual