Familia de producto: Analizador de cloro libre en línea
9184sc Free Chlorine (AFC) Amperometric Sensor
HACH LXV430.99.00001
Smart amperometric Hypochlorous Chlorine Sensor: 9184sc

产品概述

Smart amperometric Hypochlorous Chlorine Sensor: 9184sc

The low minimum detection limit and wide measurement range of the Hach 9184sc Amperometric Free Chlorine Sensor makes it ideal for efficient residual chlorine monitoring. Continuously measures hypochlorous acid (HOCl).

主要特点

  • Low Minimum Detection Limit for Efficient Residual Chlorine Monitoring:The Hach 9184sc sensor has a minimum detection limit of 5 ppb or 0.005 mg/L hypochlorous acid in the form of HOCl.
  • Wide Measurement Range:The 9184sc sensor is particularly useful for drinking water distribution monitoring, chlorination applications, demineralizer systems, wastewater, or cooling water processes due to its 0 - 20 ppm measurement range.
  • Continuous Chlorine Measurement:Measurements are made continuously by the sensor and reported to the controller
  • Compatible with Hach's "Plug and Play" Digital Controllers:The 9184sc Analyzer is designed to be used with all Hach SC digital controllers. Set up is fast and easy, and the system is compatible with the full line of Hach digital sensors.
  • Integral temperature sensor provides more accurate readings:
  • Low minimum detection limit with a minimum of 5 ppb or 0.005 mg/L Hypochlorous acid in the form of HOCl:

技术参数

参数名称说明
Accuracy2 % or ±10 ppb TFC, whichever is greater at pH <7.5
Accuracy 22 % or ±10 ppb HOCl, whichever is greater at pH <8
Calibration Interval2 months
Calibration MethodZero Calibration: electrically (automatic) or with de-chlorinated water Calibration: Comparison of lab method with process sample pH Calibration: Single- or two-point using standards or comparison of lab method with process sample
Description9184sc Hypochlorous Acid Chlorine Amperometric Sensor
Enclosure RatingIP66 / NEMA 4X
Flow0.25 L/h (200 - 250 mL/min) auto--regulated by flow through cell
InterferencesNo interference from chloramines. Chlorine dioxide and ozone interfere with measurement.
Lower Limit of Detection (LOD)0.005 mg/L HOCl
Maintenance Interval6 months Measurement Cell: for membrane and electrolyte, typical
Maintenance Interval 21 1.5 pH Cell: typical
Measuring IntervalContinuous
Measuring PrincipleAmperometric/Membrane (electrode, membrane, electrolyte)
MountingFlat, vertical surface (panel, stand, etc.)
Operating Temperature Range0 - 45 °C
ParameterActive Free Chlorine: Hypochlorous Acid Only
pH Range4 - 8, acidification unit available for pH > 8
Pressure Range0.1 - 2 bar in flow cell
Range0 - 20 mg/L hypochlorous acid HOCl
Relative Humidity0 - 90 % non-condensing
Response Time90 % in less than 90 seconds
Sample Input0.25 " outer diameter
Sample Output0.5 " inner diameter
Sample Temperature2 - 45 °C
Standard deviation0.7
Storage Conditions-20 - 60 °C
Temperature CompensationAutomatic over sample temperature range
Temperature CompensationAutomatic over sample temperature range
Warranty12 months
Weight6.5 kg
What's included?9184sc Chlorine Sensor Panel, Membranes and Electrolyte and Manual. Does not include pH sensor.