产品概述
Monitors oxygen in the beverage production process (bypass)
Sensor for the determination of dissolved oxygen (DO) in beer (bypass) and de-aerated water.
The M1100 Oxygen sensor has an unbeatable precision of 0.8 ppb and a limit of detection of 0.6 ppb. Such accurate measurement readings are essential to control low oxygen levels in beer. The absence of membrane and electrolyte means that the sensor accuracy is unaffected by process changes or pressure shocks. Maintenance and operating costs are also reduced. The sensor has been designed to ensure mechanical robustness and resistance to CIP processes to extend operational lifetime and optimise its total cost of ownership. The quick response time comes from a two second measurement frequency. Capable of measuring accurately at this frequency over a 12 month period.
主要特点
- Minimal Drift and Annual Calibration:The M1100 sensor provides immediate oxygen readings with a measurement frequency of two seconds. The instrument carries consistent readings with no drift for 12 months without calibration (when utilising standard weekly CIP processes), surpassing other optical sensors that display significant drift after only a few months in similar conditions.
- Minimal Maintenance Optical Technology:The M1100 pioneered the use of luminescent technology in brewing applications (bypass) and as a result, the sensor doesn’t require the replacement of membranes or any electrolytes. Additionally, the sensor’s accuracy is unaffected by process changes or pressure shocks, further reducing maintenance. Annual maintenance is limited to just a few minutes for a zero-point calibration. Chemicals are not required for this process, making the task easier and safer without reducing measurement precision.
- Low Level Oxygen Measurement with Accurate ppb:The M1100-L sensor has a lower detection level of 0.6 ppb. This highly accurate instrument’s readings are essential to control low oxygen levels in beer production, to decrease unnecessary line stoppages, and to increase production uptime.
- High Level Oxygen Measurement with Accurate ppm:The M1100-H sensor has a range of 0-40 ppm, and is ideal for use in wort applications, for example. Even in this harsh wort environment, the instrument maintains very good accuracy and minimal drift. Only a yearly maintenance and calibration will be required in most cases.
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | ± 0.8 ppb or 2 % whichever is greater |
| Ambient Temperature | -5 - 100 °C (23 - 212 °F) |
| Application | Beverage (bypass) |
| Calibration | Single point zero calibration |
| Certifications | 2004/108/EC - EN 61326-1 |
| Display Resolution | 0.1 ppb |
| Fitting Type | 12 mm fitting |
| Flow Rate | 50 - 300 mL/min |
| IP Rating | Sensor resistant to all common CIP methods |
| Lower Limit of Detection (LOD) | 0.6 ppb |
| Material | 316 L Stainless Steel (1.4435), Silicone |
| Operating Conditions | 0 - 95 % non-condensing relative humidity |
| Parameter | Oxygen |
| Power Supply | Universal 85-264 VAC @, 25 VA, 10-36 VDC, 25 W, |
| Range | 0 - 2000 ppb DO (dissolved oxygen) |
| Repeatability | ± 0.4 ppb or 1 % whichever is greater |
| Reproducibility | ± 0.8 ppb or 2 % whichever is greater |
| Response Time | (90%)< 10 s (gas phase) (90%)< 30 s (beer process) |
| Sample Pressure | 1 - 20 bar absolute |
| Temperature Range | Accurate from -5 - 50 °C (23 - 122 °F) Resistant - temperature from -5 - 100 °C (23 - 212 °F) |
| Warranty | 12 months |
| Weight | M1100 12 mm: 0.6 kg M1100 28 mm: 0.74 kg |
| Wetted Materials | 316 L Stainless Steel (1.4435), Silicone |
| What's included? | M1x00 sensor only, 410 or 510 controller, sensor cable, flow-chamber or process connection, must be ordered separately. |
应用领域
Clean in Place (CIP) · Production Water
同系列其他配置
| 型号 | |
|---|---|
| M1100-S00 | Orbisphere M1100 Luminescent dissolved oxygen sensor, 0-2 ppm, with 28 |
| M1100-S10H | Orbisphere M1100 Luminescent dissolved oxygen sensor, 0 - 40 ppm, with |
| M1100-S00H | Orbisphere M1100 Luminescent dissolved oxygen sensor, 0-40 ppm, with 2 |
| M1100-H0K | Orbisphere M1100 Luminescent dissolved oxygen sensor, 0-2 ppm, with 28 |
资料下载
- Orbisphere M1100: Customer Success Story-Carlsberg
- The Truth About Oxygen Testing
- Influence of Oxygen on Beer Foam Stability — Monitoring with Hach Analytical Solutions
- WHY MEASURE DISSOLVED OXYGEN IN THE BREWERY?
- Where to measure Dissolved Oxygen in the Brewery
- POR QUÉ MEDIR EL OXÍGENO DISUELTO DURANTE LA PRODUCCIÓN DE CERVEZA
- Dónde Medir El Oxígeno Disuelto Durante La Producción De Cerveza
- CÓMO MEDIR EL OXÍGENO DISUELTO DURANTE LA PRODUCCIÓN DE CERVEZA
Recomendados similares
HACHBench Service Partnership, LDO Model 2, 1 Service/Year
Replacement Protector for Clark DO sensor
HACHOrbisphere 510 Controller O₂ (EC), CO₂ (TC), Wall Mount, 100-240 VAC, 0/4-20mA, Ext. Press.
HACHField Service Partnership, 3100 LDO, 1 Service/Year
HACHBench Service Partnership, 3100 LDO, 1 Service/Year
HACHPocket Dissolved Oxygen (DO) Meter
HACHHach BioTector B7000i TOC Analyzer
HACH