产品概述
Turbidity and suspended solids in pipe probe for use in high temperature range up to 90°C in the industrial sector
The TSS HT sc Inline is a digital process probe for the extremely accurate in pipe measurement of turbidity and suspended solids in high temperature media up to 90°C. This in pipe probe is installed by ball valve safety fitting and can measure both on-line suspended solids and turbidity in one instrument. The two channel 90° scattered light turbidity method is in accordance with DIN EN ISO 7027. For the determination of suspended solids a scattered light measurement method at 120 ° is integrated. This eight channel measurement system, with an integrated bubble and temperature compensating software, enables the instrument to have a wide measuring range. The TSS HT sc Inline probes are compatible with the Hach SC200 and SC1000 controller platform. Furthermore, they are made of polished stainless steel with a scratch resistant and easy to clean sapphire window design to withstand harsh environments. Please contact Hach for further information. Request for a quote today!
主要特点
- Unique Multi-Beam Pulsed Infrared Light System:TSS sc probes have a double optical system with two pulsating infrared LEDs and four receivers. This combined multiple beam alternating light method with beam focusing facilitates accurate color independent measurement of turbidity from 0.001 to 4,000 FNU and suspended solids from 0.001 to 500 g/l.
- Automatic Compensation for Air Bubbles and Temperature Swings for Accuracy:Special software enables the system to recognize gas bubbles or temperature swings for a more accurate suspended solids or turbidity measurement.
- Standard Compliant and Precise - Ease of Calibration:The turbidity measurement complies with the standard ISO 7027. Calibration is not necessary. If the probe is used to measure the solids content, a one-point calibration suffices. For special applications, curves can be defined using several calibration points.
- Withstands Difficult Conditions and Harsh Environments:Probes are made of highly polished stainless steel with a scratch resistant sapphire window. They are design to withstand harsh environments and keep particles from sticking to the surface.
- High Temperature Probe Suitable for Temperatures of 90°C and Pressures up to 10 bar:TSS HT sc optics and electrical system are coated with special material to withstand operating temperatures of 90°C. Inline measurement does not require a flow-through cell or cooling lines.
- Decrease polymer cost, digester volume and sludge hauling:
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | Turbidity up to 1000 NTU: < 5% of the measured value ±0.01 NTU |
| Ambient Temperature | 0 - 60 °C (briefly 80 °C) |
| Application | High temperature |
| Automatic Wiper | No |
| Cable Length | 10 m (optional extension cables available) |
| Cable Length | 10 m (optional extension cables available) |
| Calibration | Turbidity (TRB): Factory calibrated Solids (TS): To be calibrated by customer on site Zero point: Permanently calibrated in the factory |
| Calibration Method | Turbidity Formazin or StablCal® Standard (at 800 NTU). Requires a calibration kit. Suspended Solids Sample specific, based on gravimetric analysis with a correction factor procedure. |
| Controller Compatibility | SC200, SC1000, SC4500. All controllers sold separately |
| Diameter | 40 mm |
| Flow | Max. 3 m/s (the presence of air bubbles affects the measurement) |
| Includes | Turbidity & Suspended Solids sensor, user manual |
| Installation style | Ball Valve Insertion, High Temperature |
| Length | 332 mm |
| Maintenance Interval | 1 h/month |
| Material | Body: Stainless Steel (1.4571, AISI 316Ti, UNS S31635), Optics: 1.4404, Gasket: FKM |
| Max Temperature | 90 |
| Max. cable length Sensor-to-Controller | 100 |
| Measurement Method | Combined multiple beam alternating light method with infrared diode system and beam focusing. Turbidity (TRB): 2-channel 90º scattered light measurement in accordance with DIN EN ISO 7027, wavelength = 860 nm Solids (TSS): 90º and 120º scattered light measurement, eight channel multi-angle measurement system, with an integrated bubble and temperature compensating, wavelength = 860 nm |
| Model | TSS HT sc |
| Mounting Configurations | Insertion |
| Operating Temperature Range | 0 - 90 °C |
| Parameter | Turbidity, Suspended Solids |
| Pressure Range | < 10 bar or < 100 m |
| Range | Turbidity (TRB): 0.001 - 9999 NTU Solids (TSS): 0.001 - 500 g/L |
| Repeatability | TSS content: < 4 % Turbidity: < 3 % |
| Response Time | 1 s < T90 < 300 s (adjustable) |
| Response Time T90 | 1 - 300 s adjustable |
| Special Features | Simple Inline installation |
| Special notes | Installation note: Distance sensor–wall > 10 cm (TSS) |
| Special Notes | Installation note: Distance sensor–wall > 50 cm (Turbidity) |
| Warranty | 12 months |
| Weight | Approx. 1.6 kg |
| Weight Sensor | 1.6 |
| What's included? | Turbidity & Suspended Solids sensor, user manual |
同系列其他配置
| 型号 | |
|---|---|
| LXV325.99.10002 | TSS HT sc Turbidity and Suspended Solids Immersion Probe, Stainless St |
| LXV325.99.20002 | TSS HT sc Turbidity and Suspended Solids TriClamp Insertion Probe, Sta |
资料下载
- Case Study: SOLITAX Sensors, LDO Probes; Accurate On-Line DO and TSS Monitoring Helps Optimize Petrochem Effluent Treatment
- Beer Recovery from Surplus Yeast Using the TSS Vari sc by Hach
- Measuring Turbidity and Suspended Solids in Power Plants
- Introduction to Hach Services
- Complete Water Analysis Solutions : Upstream and Oil & Gas Industry
- Complete Water Analysis Solutions - Midstream and Liquefied Natural Gas Production
- Care Advantage Flyer : Complete service solution for all your instruments
- Care Advantage Brochure : Complete service solution for all your instruments
Recomendados similares
HACHSonatax sc Sensor
HACHGelex Secondary Turbidity Standard, Stray Light
HACHGasket for mounting flange, Solitax sc, TSS sc
HACHWiper Set TSS sc
HACHStablcal® Turbidity Standards Calibration Kit, 2100AN / AN IS Turbidimeter, 500 mL bottles
HACHStablcal® Turbidity Standard, 0.10 NTU, 100 mL, for 2100N, 2100N IS, 2100AN, 2100AN IS
HACHStablcal® Turbidity Standard, 40 NTU, 3.78 L
HACH