RANGE PULSE

诊断行为
Warning
十六进制码
—
手册 / 页码
BA00082DEN_1917.pdf · P79

可能原因

Pulse output: the pulse output frequency is out of range.

维修指导

  1. Increase the setting for pulse weighting
  2. When selecting the pulse width, choose a value that can still be processed by a connected counter (e.g. mechanical counter, PLC etc.). Determine the pulse width:
  3. Variant 1: Enter the minimum duration that a pulse must be present at the connected counter to ensure its registration.
  4. Variant 2: Enter the maximum (pulse) frequency as the half "reciprocal value" that a pulse must be present at the connected counter to ensure its registration. Example: The maximum input frequency of the connected counter is 10 Hz. The pulse width to be entered is: 1 / (2 · 10 Hz) = 50 ms
  5. Reduce flow.
← 351 CURRENT RANGE
401 EMPTY PIPE →