Module for automatic compressed air cleaning, controlled by relays or the IQ SENSOR NET
Compressed air cleaning for sensors - this IQ SENSOR NET module make this possible, simple installatoin included:
•Fast function expansion - compressed air cleaning is quick and uncomplicated to install
•Low maintenance – due to automatic sensor cleaning
•Easy mounting - the modular design saves money, time and work
The expansion of your IQ SENSOR NET by automatically controlled cleaning functions: With the solenoid module MIQ/CHV PLUS this is as easy as the system expansion itself.
Compressed air cleaning is especially applied at heavily polluted locations, e.g. inlet of a wastewater treatment plant. Typical sensors for using compressed air cleaning are our ion selective sensors VARiON®, AmmoLyt® and NitraLyt® and the optical UV/UV-VIS probes NitraVis®, NiCaVis® and CarboVis®.
|
MIQ Module Coupling at Front |
Combined mechanical and electrical connection for rapid docking and removal of the MIQ/TC 2020 3G |
|
MIQ Module Coupling at Rear |
Combined mechanical and electrical connection, for rapid coupling to MIQ modules, up to 3 modules as a stack mounted unit possible |
|
Cable Feeds |
2 screw cable glands M 16 x 1.5 and 2 pressure hose nozzle |
|
Terminal Connections |
Screw terminal strips |
|
IQ Sensor Net Terminal Connections |
Terminal connections for the IQ Sensor Net are available on each module and can be used as required: |
|
Other Functions |
Two LEDs, yellow and red, for monitoring the operating voltage of the IQ Sensor Net; IQ Sensor Net connection, Integrated local identity function; Integrated switchable terminal resistor (SN terminator) |
|
Electric Supply |
Directly via the IQ Sensor Net |
|
Ambient Conditions |
Operating temperature: -4 °F … 131 °F (-20 °C … +55 °C); Storage temperature: -13 °F … 149 °F (-25 °C … +65 °C) |
|
Housing Material |
PC – 20 % GF (polycarbonate with 20 % fiberglass) |
|
Protection Rating |
IP 66 / corresponding to NEMA 4X (not for direct conduit connections). Conduits need to be connected with flexible adapters (CC-Box), respectively with adapters CC-PM.. |
|
Dimensions (W x H x D) |
5.67 x 5.67 x 2.05 in. (144 x 144 x 52 mm) |
|
Weight |
Approx. 1.1 pounds (0.5 kg) |
|
Certifications |
ETL, cETL (conforms with relevant UL and Canadian standards), CE |
|
Electromagnetic Compatibility |
EN 61326-1, Class B; FCC Class A |
|
Integrated Overvoltage Protection |
According to EN 61326-1 enhanced overvoltage protection for the entire system, implemented in each component |
|
Connection Medium Cable |
IQ Sensor Net cable SNCIQ or SNCIQ/UG (underground cable with additional PVC coating): 2-wire with shield; 2 x 0.75 mm2; Filler cord for easy connection of shield: 0.75 mm2; pressure resistant to 10 bar |
|
Connection Characteristics |
Energy and data transfer via 2 wire technique; resistant to reversed polarity; Comprehensive EMC shield control; cable topology within IQ Sensor Net system as required, e.g. in the form of a line, tree, star, multiple star; Total cable length: max. 1000 m/1094 yds (without signal amplifying), with signal amplifying module MIQ/JBR additional 1000 m/1094 yds (max 3000 m/3282 yds) |
|
Warranty |
3 years for defects of quality |