Приборы · Инжиниринг · Комплектация · ПЛК
0755-82221901 sales@ainstru.com
109970
Dual Parameter Monitoring Panel with SC4500 and 2 x TU5300sc Low Range Turbidimeters
HACH 109970
The Dual Parameter Turbidity Panel offers the TU5300sc Low Range Laser Turbidimeter and the SC4500 Controller in one convenient package. These pre-engineered, ready-to-install panels are designed to w

产品概述

Expertise + Simplicity

The Dual Parameter Turbidity Panel offers the TU5300sc Low Range Laser Turbidimeter and the SC4500 Controller in one convenient package. These pre-engineered, ready-to-install panels are designed to work seamlessly together, providing a simple and effective solution for your turbidity measurement needs.

主要特点

  • Quality from start to finish​:Our panel configurations allow you to maximize the inherent quality of the instruments, providing peace of mind while leveraging Hach’s 100 years of water quality innovations.
  • Up and running in no time:Our preassembled, ready-to-use solutions streamline the process between purchasing and startup.
  • Ease of installation:When you buy a water quality monitoring station from Hach, you’ll receive an expertly designed, preassembled panel that makes installation straightforward and user friendly.
  • No surprises:Prognosys™ monitors your TU5 Series online instrument, proactively alerting you to maintenance needs before your measurement becomes questionable. And a Hach Service Agreement protects your investment and helps ensure that you stay in compliance and on budget.

技术参数

参数名称说明
Cable Length1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without accessories
Calibration MethodFor Formazin and Stablcal: 20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range Custom calibration for up to 6-point calibrations
CertificationsSC4500 Controller is CE compliant and is TÜV listed to UL and CSA safety standards
ControllerSC4500
Dimensions24" x 42" x 11.5" (WxHxD)
Dimensions (H x W x D)106 nm x 610 mm x 291 mm
Fitting TypeSample quick connector: ¼-in. for ¼-in. tubing
Light SourceClass 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
MaterialASA Luran S 777K / RAL7000, TPE RESIN Elastocon® STK40, Thermoplastic Elastomer TPS-SEBS
MountingWall or rack
Mounting ConfigurationsWall Mount
Operating HumidityRelative humidity: 5 - 95% at different temperatures, non-condensing
PanelYes
ParameterTurbidity
Power Requirements (Voltage)100 - 240 VAC ±10%, 50/60 Hz; 1 A
Process ConnectionSample Inlet ¼" OD tube Sample Drain ¼" OD tube
RangeTU5300sc 0 - 700 NTU / FNU / TE/F / FTU
Sample Flow Rate100 - 1000 mL/minute; optimal flow rate: 200 - 500 mL/min
Sample PressureMax. 75 psi (517 kPa)
Sample Temperature35 - 140 °F (2 - 60 °C)
Sensor to Controller Distance (Maximal)1000 m (3280 ft), per RS485
Sensor TypeDigital output
Sensors2 x TU5300sc
Solution TypeDrinking water, raw and treated
Storage Conditions-40 - 140 °F (-40 - 60 °C)
Stray Light<10 mNTU
UnitsNTU, FNU, TE/F, FTU, EBC
VerificationLiquid: Stablcal®, Formazin (0.1 to 40 NTU) Dry: Glass Rod at <0.1 NTU
Weight43.8 lb (19.5 kg)
What's included?SC4500 Controller, Claros-enabled, 5x mA Output, 2 digital Sensors, 100-240 VAC, US plug (LXV525.99EA1551) Flow Sensor for TU5300sc and TU5400sc Laser Turbidimeter (LQV160.99.00002) Two TU5300sc Low Range Laser Turbidimeters. Ensures precise and reliable turbidity measurements (LXV445.53.B3112) User manual

应用领域

Municipal Water

资料下载