Orbisphere Hydrogen (H₂) TC Sensor with purge gas nitrogen, protection cap and ext. temp. Sensor (10 bar)Снято с производства
HACH 31270
Allows you to monitor hydrogen in the power production process.

产品概述

Allows you to monitor hydrogen in the power production process.

Standard Hydrogen TC sensor with external temperature sensor adaptor with nitrogen purge and protection cap. Used with Orbisphere 3654. Sensor Cable to be ordered: 32505.03. External Temperature sensor included. Maximum pressure: 10 bar

主要特点

  • Patented thermal conductivity (TC) sensor design:Selective and precise measurement of hydrogen
  • Continuous measurements in gaseous or liquid samples:No specific sample preparation required
  • Rugged construction can handle harsh plant conditions and high sample pressures:Flexibility of installation for better process control
  • Very low maintenance:Limited exposure to radiations in nuclear power plant
  • Very repeatable:Tight control of hydrogen in critical processes

技术参数

参数名称说明
Accuracy(Sample temp. 0 - 50°C independent of calibration temperature) The greater of Membrane 29561A: ±1% of reading, or ±2 ppb, or ±0.03 cc/kg, or ±1.5 bar Membrane 2952A: ±1% of reading, or ±8 ppb, or ±0.1 cc/kg, or ±6 mbar Membrane 2935A: ±1% of reading, or ±25 ppb, or ±0.4 cc/kg, or ±20 mbar
Accuracy 2(Sample temp. 0 - 50°C independent of calibration temperature) The greater of Membrane 2952A: ±3% of reading, or ±60 ppb, or ±0.6 cc/kg, or ±20 mbar Membrane 2935A: ±3% of reading, or ±150 ppb, or ±2.5 cc/kg, or ±50 mbar
ApplicationWaste gas offgas, reactor coolant
Certifications2004/108/EC - EN 61326-1
Cycle Time20 sec
Flow RateTypical Flow Rate Membrane 29561A: 220 mL/min Membrane 2952A: 200 mL/min Membrane 2935A: 100 mL/min
MaterialPFA
MembraneThickness 25 µm
Protection ClassIP65
Radiation Dose Limits105
Radiation Resistance LevelMembrane 29561A: 105 rad Membrane 2952A: 108 rad Membrane 2935A: 108rad
RangeRange at 25 °C Membrane 29561A: 0 - 2 ppm, or 0 - 25 cc/kg, or 0 - 1.5 bar Membrane 2952A: 0 - 10 ppm, or 0 - 120 cc/kg, or 0 - 6 bar Membrane 2935A: 0 - 20 ppm, or 0 - 220 cc/kg, or 0 - 12 bar
Response Time17 s
Sample Pressure0 - 20 bar
Temperature Range-5 - 50 °C
Weight0.95 kg
What's included?Sensor only. Controller, sensor cable, flow-chamber or process connection must be ordered separately.