Семейство продуктов:
3422 Series Analog Contacting Conductivity Sensors, Cell Constant k=10 cm⁻¹
Contacting Conductivity Sensor, High Conductivity (k=10), with ¾" PVDF Compression Fitting
Ultimate accuracy from ultra-pure to high conductivity applications.
产品概述
Ultimate accuracy from ultra-pure to high conductivity applications.
Contacting conductivity sensor with a nominal cell constant k = 10 cm-1, ¾-inch NPT PVDF compression fitting, 6 m analogue cable, and titanium electrode. Measuring range 0 -200000 µS/cm.
主要特点
- High Performance Design:These sensors are manufactured to exacting tolerances using high quality, rugged materials for demanding applications including ultra-pure water, clean-in-place (CIP), and boiler/condensate monitoring. Each sensor is tested to determine its unique, absolute four-digit cell constant. Also, each sensor has a PT1000 RTD temperature element built into its tip for exceptionally fast response to changes in temperature with ±0.1 ºC accuracy.
- Resistivity and Conductivity Measurement Capability:These enhanced performance sensors measure from 0.057 µS/cm or 18.2 MΩ (theoretically pure water) up to 200,000 µS/cm.
- Versatile mounting styles:Compression Fitting Sensors feature titanium electrodes and a compression fitting for universal installation with up to 102 mm insertion depth. The ½-inch or ¾-inch male NPT compression fitting are offered in PVDF or 316 stainless steel. A longer version of this sensor is available for use with a 316 stainless steel ball valve hardware assembly to insert/retract the sensor form the process without stopping the flow. The longer version can also be used for insertion through a compression fitting. Maximum insertion depth is 178 mm.
- Full-Featured “Plug and Play” with Digital Hach SC Controllers:There are no complicated wiring or set up procedures with any Hach SC controller. Just plug in any combination of Hach digital sensors and it's ready to use—it’s “plug and play.”
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | ±2 % of reading above 200 µS/cm |
| Cable Length | 6 m (20 ft) |
| Cell Constant k | 10 cm⁻¹ |
| Diameter | 12.7 mm |
| Electrode Material | Titanium |
| Flow | 0 - 3 m/s maximum, fully immersed |
| Installation style | Compression |
| Length | 193.5 mm |
| Material | Compression Fitting: ¾" NPT PVDF |
| Operating Temperature Range | -20 - 200 °C |
| Pressure Range | 1.7 bar at 150 °C |
| Range | 0 - 200000 µS/cm |
| Repeatability | ±0.5 % of reading |
| Response Time T90 | 30 s |
| Sensitivity | ±0.5 % of reading |
| Sensor Immersion Depth | 155 mm |
| Sensor Thread | ½" NPT |
| Sensor Type | Analog |
| Temperature Measurement | -20 - 200 °C |
| Temperature Sensor | PT1000 RTD |
| Transmission Distance | 100 m maximum; 1000 m maximum, when used with a termination box |
| Warranty | 12 months |
| What's included? | Includes: sensor with 6 m (20 ft) cable and manual |
应用领域
Boiler Water · Clean in Place (CIP) · Membrane Protection · Membrane Protection
同系列其他配置
| 型号 | |
|---|---|
| 3422E4A | Contacting, Conductivity Analog Sensor, Cell Constant 10.0, 3/4 inch N |
| 3422E4A20E | Contacting, Conductivity Analog Sensor, Cell Constant 10.0, Titanium E |
| 3422E6A | Contacting, Conductivity Analog Sensor, Cell Constant 10.0 |
资料下载
- Introduction to Hach Services
- Care Advantage Flyer : Complete service solution for all your instruments
- Care Advantage Brochure : Complete service solution for all your instruments
- Complete Water Analysis Solutions for the Food Industry
- Soluciones completas de análisis de agua INDUSTRIA ALIMENTARIA
- Contacting Conductivity/Resistivity Sensors
- C1D2 sc Sensor Control Drawing
- User Instructions: sc Sensors Class 1, Division 2 Safety Precautions
Похожие
HACH1990 µS/cm Conductivity Standard; (995 mg/L TDS, 1000 mg/L NaCl), 1L
Далее →
HACHRadiometer Analytical CDC565 Conductivity Cell (4-pole Pt sensor, k = 1.0 cm-1, epoxy body)
Далее →
HACHConductivity Solution, 1000-2000 µS/cm, 1 L
Далее →
HACHConductivity Solution, 2000-150,000 µS/cm, 1 L
Далее →
HACHConductivity Solution, 200,000-300,000 µS/cm, 1 L
Далее →
HACHConductivity Solution, 300,000-800,000 µS/cm, 1 L
Далее →
HACHCDW97-K10, Conductivity probe for salinity and brackish water
Далее →
A
HACHSolution, Conductivity Assembly
Далее →