产品概述
Hach has transitioned the CL17 to the newly upgraded CL17sc, which delivers improved usability and performance. Hach will continue to support the legacy CL17 with service and parts.
Dependable online analyzer for colorimetric DPD analysis of free chlorine, total chlorine, or permanganate in disinfection applications.
*This product has not been evaluated to test for chlorine and chloramines in medical applications in the United States. For medical applications, please see the Hach CM130.
主要特点
- EPA-approved Free or Total Chlorine Analysis:Hach’s CL17 Chlorine Analyzer is compliant with US EPA Regulation 40 CFR 141.74, and both Method 4500-CL G and Method 334.0 can be used for measurement of residual chlorine in drinking water.
- Accurate, Reliable Results.:The Hach CL17 is the industry standard for online chlorine monitoring. It comes factory-calibrated and does not require recalibration unless specified by your regulatory agency. The CL17 analysis method is not affected by changes in chlorine concentration, sample pH, temperature, flow, or pressure, resulting in better accuracy than other online chlorine testing methods.
- Simple, predictable maintenance.:The CL17 can operate unattended for up to 30 days. Monthly preventative maintenance can be performed in 15 minutes or less and requires no special tools.
: :
技术参数
| 参数名称 | 说明 |
|---|---|
| Accessories | Pressure Regulator Installation Kit Maintenance Kit |
| Accuracy | ± 5 % or ±0.03 mg/L (ppm) as CL₂, whichever is greater |
| Air Purge | 0.1 cfm (0.17 m³/h) instrument quality air at max. 20 psig (ca. 1.4 bar) with 1/4-inch OD tube, quick disconnect fitting |
| Alarm | Two alarms selectable for sample concentration alarm, analyzer system warning, or analyzer system shut-down alarm. Each is equipped with an SPDT relay with contacts rated for 5A resistive load at 230 V AC. |
| Certifications | Europe, CE Approved with: EN 61326-1 CISPR 11 EN 61010-1 IEC 60529 North America: UL 61010A-1 CAN/CSA C22.2 No. 1010.1-92 |
| Communication Capabilities | 4-20mA Output |
| Cycle Time | 2.5 min |
| Description | CL17 Chlorine Analyzer with Pressure Regulator and Reagents for Free Chlorine |
| Dimensions (H x W x D) | 454 mm x 314 mm x 179 mm |
| Display | LCD, 3-1/2 inch digit measurement readout and six-character alphanumeric scrolling text line |
| Drain | 1/2-inch ID flexible hose, hose barb |
| Enclosure Construction | ABS plastic, two clear polycarbonate windows, IP62-rated with the gasketed door latched. |
| Inlet | 1/4-inch OD polyethylene tube, quick-disconnect fitting |
| Inlet Pressure | 1 to 5 psig (0.07 to 0.34 bar), 1.5 psi is optimum |
| Interferences | Other oxidizing agents such as bromide, chlorine dioxide, permanganate and ozone will cause a positive interference. Hexavalent chromium will cause a positive interference: 1 mg/L Cr⁶⁺ = approximately 0.02 mg/L as Cl₂. Hardness must not exceed 1,000 mg/L as CaCO₃. |
| Light Source | Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000 hours estimated minimum life |
| Lower Limit of Detection (LOD) | 0.03 mg/L (ppm) |
| Mounting | Wall mount |
| Operating Humidity | Up to 90 % at 40 °C (104 °F) maximum |
| Operating Temperature Range | 5 - 40 °C (41 - 104 °F) |
| Parameter | Chlorine, Free & Total |
| Pressure Range | Inlet Pressure to Sample Conditioning: 1.5 - 75 psi (0.1 - 5.2 bar) |
| Pressure Regulator Included? | Yes |
| Range | 0 - 5 mg/L Cl₂ |
| Recorder Outputs | One 0/4-20 mA with an output span programmable over any portion of the 0 to 5 mg/L range. Recommended load impedance 3.6 to 500 ohms, 130 V isolation from earth ground. |
| Repeatability | 0.01 ppm or 3.0 % , whichever is greater |
| Sample Flow Rate | 200 - 500 mL per minute minimum |
| Sample Temperature | 5 - 40 °C (41 - 104 °F) |
| Warranty | 12 months |
| Weight | 10.49 kg |
| What's included? | CL17 Chlorine Analyzer, 1 Month Supply of Free Chlorine Reagents, 100 mL of Sulfuric acid, Maintenance kit, Installation kit with pressure regulators, and Manual. (The power cord is ordered separately.) |
资料下载
- Chlorination, Chloramination And Chlorine Measurement
- Part One: CL17 for Permanganate Measurement in Drinking Water (Permanganate Only)
- Part Two: CL17 for Permanganate Measurement in Drinking Water (Two Oxidants)
- Monitoring Chlorine in Turbid Cooling Tower Water with the Hach AT1000 Titrator and CL17 Analyzer
- Monitoring Chlorine in Turbid Cooling Tower Water with the Hach AT1000 Titrator and CL17 AnalyserDOC040.52.10112
- Monitoreo del cloro en el agua turbia de la torre de refrigeración con el titulador AT1000 y el analizador CL17 de Hach
- Chlorine Analyzer Calibration Verification - Ensuring Confidence in Industrial Water Quality
- Amperometric Probes or DPD Analyzers-Which Best for On-Line
Похожие
9187sc Chlorine Dioxide Analyzer with sc100 controller and RS485
HACHElectrolyte for Chlorine/Ozone
HACHAT1000 Titration Bundle, Total Chlorine (Backward)
HACHSpecCheck набор вторичных гелевых стандартов, хлор DPD - низкий диапазон (LR)
HACHВводный курс по анализатору CL17sc (один пользователь)
HACH8450 Free Chlorine Probe, 25 ft
HACHpHD pH-датчик, CLF10sc датчик свободного остаточного хлора, с контроллером SC4500, 2 канала
HACH