9494600
Семейство продуктов: IO9000 Module
IO9004 Module (connects through auxiliary port), includes multiple 0/4-20 mA outputs and inputs
HACH 9494600
Относится IO9000 Module(IO9000Module)
The IO9004 Module allows the use of anlaog inputs and outputs with the AS950 sampler. With the IO9004 module, the AS950 provides more input and output options than any other Hach sampler before it.

产品概述

The IO9004 Module allows the use of anlaog inputs and outputs with the AS950 sampler. With the IO9004 module, the AS950 provides more input and output options than any other Hach sampler before it.

主要特点

  • Maximum Versatility:The IO9000 module allows the use of analog inputs and outputs with the AS950 sampler. With the IO9000 module, the AS950 provides more input and output options than any other Hach sampler before it. For added versatility the IO9000 is available in two versions; IO9001 is available for simple applications that only require 1 high voltage relay while the IO9004 is fully loaded. See specifications for more detail.
  • Outputs:Analog 0/4-20 mA outputs can be used to send information to a SCADA system for parameters such as level, velocity, flow and pH. Examples of alarms and events include: sensor outside of limits (flow related, water quality related, internal temperature or ambient temperature), program started, program completed, sample being taken, bottle full (single bottle), pump running forward, pump running in reverse, pump error, distributor error, power failure, or main battery low.
  • Inputs:Analog 0/4-20mA inputs can be used to interface third party sensors to the AS950. Once logged, the analog process measurement can be used like a Hach sensor in applications such as setpoint alarms or event-based sampling. In addition, the third party measurements can be monitored on the user interface along with Hach sensors and peripherals.
  • Relays:Like the digital outputs, these relay outputs are used to indicate alarms or events. The difference is that relays can be used for switching AC mains line voltage to control higher power functions. Examples for use are for a warning light or sound signal, switching a diversion valve or gate or a control signal to another machine.

技术参数

参数名称说明
Altitude2000 Meters
Analog Inputs(2) 0/4-20 mA inputs generic to map external processes to logable measurements (ex. third party ultrasonic level). Current Input Mode: • Accuracy: 0.5% of FSR over -40 to 50 °C, 0.3% of FSR over -10 to 50 °C • Resolution: <0.05% of FSR • Loop Power Supply Voltage: 18 V max. (referenced to (-) input) • Burden: 108 Ω max. + 0.4 V max. Voltage Pulse Input Mode: • Input Resistance: 11 kΩ typical • Pulse High Level: 4 V min. to 15 V max. • Pulse Low Level: 0 V min. to 2 V max. • Hysteresis: 0.5 V typical • Pulse Width (high state or low state): 50 ms min. Wire size: 24 AWG to 16 AWG solid or stranded copper
Analog Outputs(3) 0/4-20 mA outputs that map logged analogue measurements (ex. Level, velocity, flow, pH) - Can function as loop-powered or self-powered. • Accuracy: 0.8% of FSR over -40 to 50 °C, 0.4% of FSR over -10 to 50 °C • Resolution: <0.05% of FSR • External Loop Power Supply Voltage: 18 V max. • Internal Loop Power Supply Voltage: 14 V min., 18 V max. • Loop Minimum Resistance: 0 Ω • Burden Voltage: 3.6 V max. at 25 mA Wire size: 24 AWG to 16 AWG solid or stranded copper
Cable Length1.22 m (4 ft)
Calibration2 point linear user calibration via controller UI
CertificationsCETLus
Digital Outputs(4) Low-voltage contact closure mapped to alarm events. • ±30 VDC max., 150 mADC • 20 VAC-RMS, 150 mAAC-RMS Wire Size: 24 AWG to 16 AWG solid or stranded copper
DimensionsEnclosure with supplied mounting brackets: 24.7 cm (9.71") Length, 30 cm (11.82") Width, 13 cm (5.12") Depth
Enclosure RatingIP66 / NEMA 4X
Housing MaterialPolycarbonate (UL94 VO), hardware 18-8SS
InputsEnclosure Inputs Cable Entry: - High Voltage Ports: 4 Cable Bushings (accomodate cable diameters 0.170° - 0.450°) - Low Voltage Ports: 4 Cable Bushings (accomodate cable diameters 0.170° - 0.450°) All ports accept user installed ½" (16 mm) Trade Size Conduit
Inputs and OutputsAnalogue: Output= 3, Input= 2; Digital: Output= 4, Input= 4
ModelIO9004 - Multiple IN/OUT
MountingMounting brackets and hardware included.
Operating ConditionsTemperature: -40 - 50 °C Humidity: 100% (non-condensing)
Power inputPowered by controller
Relay Functions(4) Form C (SPDT) • Voltage Rating (Contacts Open): 20 VAC-RMS to 230 VAC-RMS • Current Rating (Contacts Closed): 6 mAAC-RMS to 500 mAAC-RMS Wire Size: 20 AWG to 14 AWG solid or stranded copper
Storage ConditionsTemperature: -40 to 80 °C, Humidity: 100% (non-condensing)
Weight2.5 kg

应用领域

Discharge · Flow Monitoring · Groundwater · Industrial Wastewater · Influent for Wastewater · Municipal Water · Pre-Treatment · Primary Treatment · Production Water · QA/QC Lab · Regulations · Secondary Treatment · Source Water · Storm Water · Surface Water · Water Cycle · Water Reuse