产品概述
Total chlorine confidence. From the global leader in chlorine analysis.
Meet your unique water cycle needs with the new Ultra Low Range CL17sc online total chlorine analyzer from Hach. Whether treating utility water, product water, or wastewater discharge, you’ll have the direct chlorine measurements you need to optimize your process. You’ll be able to protect valuable assets and report to regulatory agencies with confidence.
主要特点
- Monitor and Optimize your Dechlorination process:Chlorine breakthroughs can adversely affect your process, product, or equipment. You’ll be able to mitigate issues and optimize your process with the Hach Ultra Low Range CL17sc analyzer, which has the lowest limit of detection at less than 8 ppb. It's time to accurately measure your total residual chlorine and take control of your dechlorination process.
- Depend on direct measurements:Online residual chlorine measurement options for ultra-low range have been either inadequate or inefficient – until now. The Ultra Low Range CL17sc gives you direct total chlorine results, eliminating the need for ORP or amperometric measurement. The Ultra Low Range CL17sc uses colorimetric DPD Standard Method 4500-Cl G and consistently maintains accuracy without loss of sensitivity in your new target range.
- Protect your assets:Understand the true impact of chlorine exposure. The Ultra Low Range CL17sc allows for process control across your water cycle, including visibility into GAC exhaustion or channeling and dechlorinating agent dosage. It is the only instrument with a cumulative chlorine counter, which helps you forecast your RO membrane efficiency and its useful life.
- Smart Monitoring Enabled:The new Ultra Low Range CL17sc analyzer is Smart Monitoring enabled, providing real-time visibility into your processes so you can optimize and automate your chlorination/dechlorination. You’ll have peace of mind and more time to focus on high-value tasks.
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | ± 5% or ± 0.01 mg/L (whichever is greater) from 0 - 4 mg/L; ± 10% from 4 - 5 mg/L |
| Air Purge | Optional with 3/8-inch quick-connect fitting and tubing; 0.003 m3/min at 1.38 bar (20 psig) maximum |
| Alerts | Low and high chlorine. Dirty cell. Low and high sample flow. |
| Certifications | CE compliant with: EN 61326-1, CISPR 11, EN 50581 ACMA RCM South Korea KC Certificate North America: FCC Supplier’s Declaration of Conformance, IEC/EN 60529, ICES-003 |
| Controller Compatibility | SC200, SC1000, SC4500. All controllers sold separately |
| Cycle Time | 2.5 minutes |
| Dimensions (H x W x D) | 342 mm x 329 mm x 177 mm |
| Drain | ½-inch ID flexible hose |
| Enclosure Rating | IP66 |
| Inlet | ¼-inch OD polyethylene tube, quick-disconnect fitting |
| Inlet Pressure | 0.3 - 5.2 bar (4.5 - 75 psig) supplied to Y-strainer 0.1 - 0.3 bar (1.5 - 5 psig) supplied to analyzer |
| Installation Kit | Pressure Regulation Installation Kit |
| Interferences | Other oxidizing agents such as bromide, chlorine dioxide, permanganate, and ozone will cause a positive interference. |
| Light Source | LED, measurement at 510 nm, 1 cm light pathlength |
| Lower Limit of Detection (LOD) | 0.008 mg/L Cl₂ at 25 °C |
| Mounting | Wall mount |
| Operating Humidity | 0 - 90% relative humidity, non-condensing |
| Operating Temperature Range | 5 - 40 °C (41 - 104 °F) |
| Parameter | Chlorine, total residual |
| Power Requirements (Voltage) | 12 VDC, 400 mA maximum (supplied by the controller) |
| Range | 0.008 - 5 mg/L Cl₂ |
| Reagent Consumption | 0.5 L of buffer solution and 0.5 L of indicator solution in 31 days |
| Sample Flow Rate | 60 - 200 mL/min through the instrument |
| Sample Requirements | Y-strainer filtration with 40-mesh screen or higher |
| Sample Temperature | 5 - 40 °C (41 - 104 °F) |
| Warranty | 12 months |
| Weight | 4.1 kg (analyzer only) |
| What's included? | ULR CL17sc Chlorine Analyser with Pressure Regulation Installation Kit (9790300); SC4500 Controller for 2 digital sensors, Prognosys, 5x mA Output, 100-240 VAC, without power cord (LXV525.99A11551); Total Chlorine Indicator Solution (2263411); Total Chlorine Buffer Solution (2263511); DPD Compound (2297255); User Manual |
应用领域
Boiler Water · Discharge · Filtration · Groundwater · Membrane Protection · Municipal Water · Potable Water · Pre-Treatment · Product Loss · Surface Water
同系列其他配置
| 型号 | |
|---|---|
| 9792500 | Ultra Low Range CL17sc Total Chlorine Analyzer with Standpipe Installa |
| 9792600 | Ultra Low Range CL17sc Total Chlorine Analyzer with Pressure Regulatio |
| 9792700 | Ultra Low Range CL17sc Total Chlorine Analyzer with Standpipe Installa |
资料下载
- Total vs Free Chlorine monitoring for Dechlorination
- CL17 and ULR CL17sc vs ORP sensors : Is ORP Really Your Best Option for Dechlorination Measurement?
- Power Industry: Optimization of Dechlorination and CIP Processes for RO Filtration
- Semiconductor Application: ULR CL17sc Success at Microelectronics Manufacturing Facility
- Semiconductor Application: ULR CL17sc Success at Semiconductors Manufacturing Facility in TX
- Semiconductor Application: ULR CL17sc vs standard CL17sc and ORP at a Semiconductor Manufacturing Facility
- SBS Overfeed and Underfeed: Achieving a delicate balance to maintain RO membranes
- Ultrapure Water Dechlorination : Addressing GAC Bed Chlorine Leaks
Похожие
HACHChlorine Test Strips
HACHAT1000 Titration Bundle, Total Chlorine (Backward)
HACHChlorine Dioxide Color Disc Test Kit, Model CLD-2
HACHTesting Chlorine with DR300 Online Course
Chlorine, NEMA 4X 1/2 Din, Hart
MANUAL,SERV. CHLORINE ANALYZER
IC Eprom Chlorine Analyzer CL17, Rev P1.8