产品概述
Highly accurate and customizable oxygen measurement for all environments
The Orbisphere GA2800 EX oxygen (O₂) Electrochemical (EC) sensor is designed for process monitoring as well as laboratory analysis in the liquid or gas phases across. With its large measuring possibilities the Orbisphere GA2800 EX O₂ EC-sensor can be used for a wide range of harsh environments from chemical or oil to petrochemical plants where oxygen measurement is critical. A small residual signal with unrivalled accuracy (±0.1 ppb) is made possible by the exclusive sensor design. Different pre-mounted membrane kits are available to fulfill any particular process requirements and O₂ measurement ranges required. The GA2800 EX O₂ electrochemical sensor is mechanically resistant to pressure up to 100 bar due to Hastelloy material and FFKM/FFPM O-rings.
主要特点
- Easy and fast maintenance:Traditional cleaning processes can take more than thirty minutes to complete. The GA2400 and GA2800 EX come with an innovative cartridge system that contains everything needed to complete membrane and electrolyte solution change in less than 5 minutes.
- Low level oxygen measurement and unrivaled accuracy:The Hach Orbisphere GA2400 and GA2800 EX utilize an electrochemical oxygen sensor with a lower detection level of 0.1 ppb and unrivaled highly accurate readings of ±1%. This feature allows users to ensure control of low oxygen levels, for product and equipment integrity and avoiding corrosion.
- Robust for harsh environments:The robust stainless steel or Hastelloy design makes the GA2400 and GA2800 EX sensor ideal for the most demanding applications. ATEX certification comes standard in all GA2800 units, making the sensor especially suitable for harsh chemical environments.
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | Accuracy = Trueness (ISO 57251) 2935A-A: The greater of ±1% of reading or ± 10 ppb, or ± 20 Pa 2952A-A: The greater of ±1% of reading or ± 2 ppb, or ± 5 Pa 2956A-A: The greater of ±1% of reading or ± 0.1 ppb¹, or ± 1 ppb², or ± 0.25 Pa 29552A-A: The greater of ±1% of reading or ± 2 ppb, or ± 5 Pa ¹ Accuracy is ± 0.1 ppb for 410, 510, 362x, 360x and 3655 instruments ² Accuracy is ± 1 ppb for 366x and 3650 instruments |
| Body Material | Hastelloy |
| Certifications | CE, Ex II 1 G, Ex ia IIC T6 |
| Flow Rate | Recommended flow rate in flow chamber 2935A-A: 25 mL/min 2952A-A: 50 mL/min 2956A-A: 180 mL/min 29552A-A: 50 mL/min |
| Material | O-Rings: FFKM/FFPM |
| Model | GA2800 EX |
| Pressure Range | 100 bar |
| Pressure resistance | 40 bar minimum with default PPS collar (100 bar with stainless steel collar) |
| Protection Class | IP66 |
| Range | Range at 25 °C (77 °F) 2935A-A: 10 ppb - 400 ppm or 20 Pa - 1,000 kPa 2952A-A: 1 ppb - 80 ppm or 5 Pa - 200 kPa 2956A-A: 0.1 ppb - 20 ppm or 0.25 Pa - 50 kPa 29552A-A: 2 ppb - 80 ppm or 5 Pa - 200 kPa |
| Response Time | Response time at 25 °C (77 °F) for a 90% signal change 2935A-A: 2.5 min 2952A-A: 38 s 2956A-A: 7.2 s 29552A-A: 90 s |
| Sensor material | Hastelloy |
| Temperature Compensation | -5 - 60 °C (23 - 140 °F) |
| Temperature resistance | -5 - 100 °C (23 - 212 °F) |
| Warranty | 12 months |
| Weight | 0.3 kg (0.66 lbs) |
| Wetted Materials | Hastelloy No O-rings are in contact with the sample |
| What's included? | Oxygen Sensor (GA280E-HKS), incl. plastic screw-on storage cap for sensor head protection; protection cap without grille (33051-S0); protection cap with grille (33051-SP) |
同系列其他配置
| 型号 | |
|---|---|
| GA2400-S00 | Orbisphere GA2400 Stainless Steel Oxygen Sensor (EC), 40 bar, EPDM O-r |
| GA2400-S00T | Orbisphere GA2400 Stainless Steel Oxygen Sensor (EC) for 6110 TPO Anal |
| GA2400-S0S | Orbisphere GA2400 Stainless Steel Oxygen Sensor (EC), 100 bar, EPDM O- |
| GA280E-HVS | Orbisphere GA2800 ATEX Oxygen Sensor (EC), Hastelloy, 100 bar, FKM/FPM |
| GA280E-SKS | Orbisphere GA2800 ATEX Oxygen Sensor (EC), Stainless Steel, 100 bar, F |
| GA280E-SVS | Orbisphere GA2800 ATEX Oxygen Sensor (EC), Stainless Steel, 100 bar, F |
资料下载
- Dissolved Oxygen Measurement in Ultrapure water for Semiconductor Manufacturing
- Residual Oxygen controlled blanketing in the Semiconductor Industry
- Dissolved Oxygen Measurement in Ultrapure Water for Semiconductor Manufacturing
- Dissolved Oxygen and Hydrogen Analysis in Reactor Coolant Systems: Orbisphere 510 with Channel Compensation
- Introduction to Hach Services
- Care Advantage Flyer : Complete service solution for all your instruments
- Care Advantage Brochure : Complete service solution for all your instruments
- Comparison Sheet - Orbisphere Dissolved Oxygen
Похожие
Replacement Protector for Clark DO sensor
HACHBench Service Partnership, LDO Model 2, 1 Service/Year
HACHField Service Partnership, 3100 LDO, 1 Service/Year
HACHBench Service Partnership, 3100 LDO, 1 Service/Year
HACHPocket Dissolved Oxygen (DO) Meter
HACHField Service Partnership, 2586sc Oxygen Scavenger Analyzer, 2 Services/Year
HACHCOD TNTplus Vial Tests, LR
HACH