产品概述
The original maintenance-free optical oxygen sensor for power plants.
The Orbisphere K1100 is a high accuracy Luminescent Dissolved Oxygen Sensor optimized for power industry application. Since 1978, Hach Orbisphere LDO analyzers have set the industry standard for oxygen measurement by delivering confidence to every water chemistry manager. Together with the Orbisphere 410 single channel controller and the Orbisphere 510 multichannel controller, offers a new way of monitoring oxygen in power plants. Orbisphere dissolved oxygen sensors set the industry standards for oxygen measurement by offering peace of mind to every water chemist. The absence of membrane and electrolyte means that the sensor accuracy is unaffected by process changes such as changes in flow. The K1100 dissolved oxygen analyzer maintains this tradition and offers significant operating and cost benefits.
主要特点
- One calibration per year:One zero point calibration per year is all that is needed with the K1100 sensor. Designed for minimal drift, luminescent technology makes the K1100 sensor the most stable sensor with the longest calibration interval in the industry.
- No membranes = two minutes of maintenance:With no membranes to replace and no electrolyte solution to replenish, the K1100 requires only two minutes of maintenance per year. Corrosive or hazardous chemicals are not required, making the annual task faster, easier and safer without reducing measurement precision.
- Low cost retrofit:The complete system consists of a controller, a flow chamber, and the K1100 Luminescent Dissolved Oxygen sensor. The sensor is compatible with Hach Orbisphere 28 mm flow chambers, eliminating the need for engineering changes. Installation is fast and easy and does not require special preparation.
- A new level of confidence:The K1100 optical sensor is the first to use luminescent measurement technology to measure both ppb and ppm oxygen levels in power plants. Since 1978, Hach Orbisphere sensors have set the industry standard for oxygen measurement by delivering confidence to every water chemistry manager. The K1100 maintains this tradition and offers significant operating and cost benefits.
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | ± 0.8 ppb or 2 % whichever is greater |
| Ambient Temperature | -5 - 50 °C (23 - 122°F) |
| Analog Outputs | 3 Smart 0/4 to 20 mA (500 Ohms) programmable as linear or tri-linear, configurable to send diagnostics or alarm informations. |
| Application | In-line |
| Calibration | Single point zero calibration with standard 99.999% nitrogen (quality 50) or equivalent oxygen free gas |
| Certifications | 2004/108/EC - EN 61326-1 |
| Communication Capabilities | 3 x 0/4-20 mA; RS485; Ethernet |
| Display Resolution | 0.1 ppb |
| Lower Limit of Detection (LOD) | 0.6 ppb |
| Mounting | 28 mm Orbisphere fitting |
| Parameter | Oxygen |
| Power Requirements (Voltage) | 100 - 230 VAC |
| Range | 0 - 2000 ppb dissolved O2 (DO) |
| Relays | Measurement board: 3 measurement alarm relays (1A-30 VAC or 0.5A-50 VDC), configurable to send diagnostics information. Main board: 1 system alarm relay (1A-30 VAC or 0.5A-50 VDC). |
| Repeatability | ± 0.4 ppb or 1 % whichever is greater |
| Reproducibility | ± 0.8 ppb or 2 % whichever is greater |
| Response Time | (90%) <10 s (gas phase); <30 s (liquid phase) |
| Sample Flow Rate | 50 - 300 mL/min |
| Sample Pressure | 1 - 20 bar absolute (14.5 - 290 psia) |
| Sample Temperature | -5 - 50°C (23 - 122°F) |
| Sensors | K1100 LDO |
| Storage Conditions | -5 °C - 100 °C |
| Temperature Range | Accurate from -5 - 50 °C (23 - 122 °F) Resistant - temperature from -5 - 100 °C (23 - 212 °F) |
| Warranty | 12 months |
| Weight Sensor | 0.6 |
| What's included? | K1100 Sensor only Controller, sensor cable, flow chamber or process connection must be ordered separately |
应用领域
Boiler Water · Steam Cycle · Ultrapure Water
资料下载
- Complete Water Analysis for the UPSTREAM OIL & GAS INDUSTRY
- Optical Dissolved Oxygen Measurement in Power and Boiler Applications
- Practice report: Dissolved Oxygen Measurement in Power Plants
- Mesure de l’oxygène dissous dans les centrales électriques
- Medición de oxígeno disuelto en centrales de energía
- Connah's Quay Opts for Almost Zero Maintenance DO Monitors
- Connah’s Quay opta por monitores de OD que casi no necesitan mantenimiento
- Application Note: Increasing LDO Longevity in Brewing
Похожие
Replacement Protector for Clark DO sensor
HACHBench Service Partnership, LDO Model 2, 1 Service/Year
HACHField Service Partnership, 3100 LDO, 1 Service/Year
HACHBench Service Partnership, 3100 LDO, 1 Service/Year
HACHPocket Dissolved Oxygen (DO) Meter
HACHField Service Partnership, 2586sc Oxygen Scavenger Analyzer, 2 Services/Year
HACHOrbisphere 510 Controller O₂ (EC), CO₂ (TC), Wall Mount, 100-240 VAC, 0/4-20mA, Ext. Press.
HACH