Семейство продуктов: CLF10sc and CLT10sc Amperometric Chlorine Analyzers
CLF10sc Free Chlorine Analyzer (Panel Only) with pHD Differential Sensor
HACH LXV45A.99.13022
CLT10 sc / CLF10 sc: Total and free reagentless chlorine sensors

产品概述

CLT10 sc / CLF10 sc: Total and free reagentless chlorine sensors

The CLF10 sc Free Chlorine Analyzer measures free chlorine continuously for real time process control. It is based on amperometric technology and is fully compatible with all of Hach's digital controllers. The CLF10 sc can be used in most municipal and industrial applications and is best utilized where waste stream management is a constraint. This analyzer is equipped with a pH sensor for use with the unique Cal Watch advanced diagnostic feature.

主要特点

  • Exclusive Self Diagnostics:The CLF10 sc and CLT10 sc analyzers leverage Hach's exclusive self diagnostics to alert users when the process has changed or the instrument needs servicing. Diagnostic features include the Cal Watch algorithm for warning of pH and chlorine calibration deviation and a non-contacting flow sensor for notification of insufficient sample flow.
  • No Reagent Replacement, No Waste Stream:Chlorine measurement with an amperometric analyzer, such as the CLF10 sc or CLT10 sc, does not require reagents, eliminating the need for routine reagent replacement and waste stream management.
  • Real-Time Process Control:The CLF10 sc and CLT10 sc analyzers allow for real-time control of disinfection processes by providing continuous readings that indicate when treatment conditions have changed.
  • Compatible with Hach's "Plug and Play" Digital Controllers:The CLF10 sc and CLT10 sc analyzers can be used with any Hach sc digital controller. Whether you're measuring turbidity or chlorine, you only need to learn one controller for all your water analysis measurement points. Hach sc controllers have no complicated wiring or setup procedures. Just plug in any Hach digital sensor and it's ready to use without software configuration.
  • EPA Compliant According to Method 334.0:In accordance with EPA Method 334.0, the CLF10 sc and CLT10 sc analyzers can be used for reporting chlorine residual measurements. Additionally, Hach has created a suite of laboratory products and methods to help with startup and quality control procedures required in Method 334.0.
  • Extended up-time via various service options:The Hach ISO certified service is available close to you. It offers a range from one-off installation service to warranty extension up to 5 years. Thus, the lifetime of the system is increased and costs become transparent and foreseeable.

技术参数

参数名称说明
AccuracyFree Chlorine: ± 3 % of the reference test* (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2Free Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5
Cable Connection5 pin, M12 connector
Cable Length1 m (between gateway and sc controller)
Calibration Method1-point or 2-point (zero and slope) calibration
CertificationsCE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits) Additional certifications when connected to an sc controller
Connection Drain Line1/2 in ID
Connection Sample Line1/4 in OD
ControllerPanel only
Controller Compatibilitysc controller platform
DescriptionCLF10sc Free Chlorine Analyzer (Panel Only) with pHD Differential Sensor
DimensionsSensor Only: 195 |mm x 25 mm (7.68 in x 0.98 in)
Fitting TypeUSA
IncludesCL10sc Panel, 1 m digital extension cable, Panel Manual, pHD Sensor Manual, Chlorine Sensor Manual.
InterferencesFree chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits. Total chlorine: chlorine dioxide, ozone and chalk deposits
LengthSensor: 195 mm
Lower Limit of Detection (LOD)30 ppb (0.03 ppm)
MaterialCorrosion-resistant materials, fully-submersible (stainless steel, PEEK, PVC, silicon rubber and polycarbonate)
Measurement MethodReagentless, electrochemical, three-electrode amperometric system
MountingPanel or Wall Mounting
Operating Temperature Range0 - 45 °C
OptionspHD Differential Sensor
ParameterFree Chlorine
pH Monitoring Required?With differential pH sensor
pH Range4 - 9
Pressure Range0.5 bar, no pressure impulses and/or vibrations
Range0 - 20 ppm Cl₂
Repeatability30 ppb or 3%, whichever is greater
Response TimeTotal Chlorine: 100 s or less for 90% change (T90) at a stable temperature and pH
Response TimeFree Chlorine: 140 s or less for 90% change (T90) at a stable temperature and pH
Sample Flow Rate30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature5 - 45°C (41 - 113°F)
Sampling TimeContinuous
Storage Conditions-20 °C to 60 °C dry
Warranty12 months
WeightApproximately 5.5 kg (12 lbs) Panel and empty panel-mounted components only
What's included?CL10sc Panel, 1 m digital extension cable, Panel User Manual pHD electrode, PPS, for CL10sc (9181500), pHD Sensor User Manual Digital extension cable, 1m (6122400) Replacement Sensor for CLF10sc, SS Tip (8626200), Chlorine Sensor User Manual

应用领域

Discharge · Groundwater · Influent for Wastewater · Membrane Protection · Municipal Water · Regulations · Surface Water

同系列其他配置

型号
LXV45A.99.11022CLF10sc Free Chlorine Analyzer (Panel Only) Grab Sample
LXV45B.99.12022CLT10sc Total Chlorine Analyzer (Panel Only) with Combination pH Senso