产品概述
Proven. Trusted. Reliable. The solution for exhaust gas cleaning systems.
Hach’s new PAH500 sensor was designed specifically for polycyclic aromatic hydrocarbon (PAH) measurement in maritime EGCS (scrubber) applications. The IMO’s decision to lower the Sulphur cap to 0.50% will be impactful in improving air quality, but this means as a ship owner, you need to be prepared to be compliant. Ship owners require maximum uptime for analytical equipment. The PAH500 was designed with this in mind.
A unique flow through design minimizes fouling while providing quick, accurate readings. Nitrogen purged, a quartz measurement tube, and a titanium body ensure that the PAH500 is robust and reliable in harsh EGCS applications. Primary phenanthrene standards and calibration kits are available for easy calibration and quick verification in the field by end-users, giving peace of mind that everything is operating as expected.
Along with the PAH500, Hach also offers DNV-GL approved solutions for pH, turbidity, and conductivity.
For over 85 years, Hach has offered water analysis solutions for process, lab and handheld applications. Within the maritime industry, Hach’s solutions can be found onboard vessels for gas scrubber applications, waste and potable water monitoring, ballast water disinfection monitoring, and recreational water monitoring. Hach has a global support network with locations in Europe, USA, Canada, Australia, and the Caribbean. With Hach, ship owners can:
- Maintain SOx compliance without hiring additional onboard expertise - Avoid fines by meeting USCG and IMO regulations - Provide safe water to passengers and ensure water is safe to discharge - Reduce treatment costs
主要特点
- Reliability:With hundreds of successful installations on inlet and outlet scrubber applications, you can have confidence in accurate readings, maximum uptime, premium service and support from a global water quality company. The PAH500 is the online oil-in-water sensor that delivers high sensitivity with low cost of ownership. Hach’s 85 years of water analysis leadership, experience, and innovation ensure that you’re getting the best instrument and peace of mind to go along with it.
- Certifications:We understand how important your business and your equipment are to you, that's why we take pride in receiving certifications from the following organisations: Maritime Type Approvals: DNV-GL, ABS, ClassNK, Japanese Government, and Korean Register** Marine Environmental Protection Committee: MEPC.259(68) from DNV-GL Certification demonstrates Hach's commitment to continual improvement and sustainable business performance. ** Certifications may be pending final reviews & certifications with certain Maritime Societies.
- Minimal Maintenance:With no tubes, pumps, or valves that can foul, maintenance is limited to occasional cleaning of the quartz tube and service once every other year to stay in compliance with IMO regulations (annual for USCG regulations). After all, why spend time on maintenance when you don’t have to?
- Built for Maritime:The PAH500 was purpose built for harsh maritime applications. It can detect and measure polycyclic aromatic hydrocarbons (PAHs) from 0 - 900 ppb (µg/L). Titanium housing and quartz measurement tube help to eliminate corrosion in exhaust gas cleaning system (EGCS) applications. The sensor is nitrogen purged from the factory, ensuring reliable operation in humid environments. The light source, filtering, and calibration for Phenanthrene in exhaust gas scrubbers mean that this sensor is designed specifically for your application. Hach offers a variety of products to support your EGCS application and comply with IMO regulations including: • Ultraturb sc (Turbidity) • pHD (pH) • Series 3700 (Conductivity) • SC1000 (Controller)
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | ± 5 µg/L for readings ≤ 100 µg/L ± 5% for readings > 100 µg/L |
| Cable | Analog cable, 4-conductors + shield, drain wire, 24 AWG, 90 °C, 600 VRMS, IEC 60332-1-2; UL94 VW-1, 1 - 3 m (standard: 1 m) |
| Cable Length | 1 m |
| Calibration | Calibration parameters: Polycyclic Aromatic Hydrocarbons (PAH's) Phenanthrene (PHE) Standard (traceable) |
| Certifications | EU CE Mark Maritime Type Approvals**: DNV-GL, ABS, Class NK, Japanese Government, Korean Register Marine Environment Protection Committee - MEPC.259(68) |
| Data Logger | 10 s - 15 min |
| Dimensions | Sensor Body (Ø x L): 76.2 x 145.1 mm (3.0 x 5.71 in) |
| Housing Material | Titanium |
| Measurement Method | Ultraviolet (LED) 255 nm excitation; 370 nm detection |
| Measurement Principle | UV Fluorescence |
| Measuring Interval | 60 s |
| Operating Temperature Range | 5 - 55 °C |
| Parameter | PAH (polycyclic aromatic hydrocarbon) |
| Power Consumption | 1 |
| Power Requirements (Voltage) | 12 VDC supplied by digital gateway |
| Range | 0 - 900 µg/L PAH / PHE (phenanthrene equivalents) |
| Temperature | 90 °C |
| Turbidity Compensation | Calculated, accurate for turbidity ≤ 150 NTU |
| Warranty | 12 months |
| Weight | 1 kg |
| What's included? | PAH500 Sensor, 1 m cable, mounting bracket, manual |
应用领域
Ballast Water · Discharge · Industrial Wastewater · Petrochemical
同系列其他配置
| 型号 | |
|---|---|
| LXV541.99.3001H | PAH500 Digital PAH Sensor, 15 m Cable, Bracket |
| LXV541.99.2001H | PAH500 Digital PAH Sensor, 7.7 m Cable, Bracket |
| LXV541.99.4001H | PAH500 Digital PAH Sensor, 30 m Cable, Bracket |
资料下载
- Introduction to Hach Services
- Care Advantage Flyer : Complete service solution for all your instruments
- Care Advantage Brochure : Complete service solution for all your instruments
- Statement of Compliance from DNV
- Type Approval Certificate from DNV
- PAH500 PAH Oil-in-Water Continous Online Monitoring Sensor
- Sensor PAH500 para PAH y monitorización en continuo de aceite en agua
- Capteur PAH500 pour la surveillance continue en ligne de la présence de HAP dans l’eau
Похожие
HACHSC1000 relay module for DIN-rail mounting, 4 Relays, max. 240V
HACHNP6000sc Phosphate Analyzers
Stainless Steel Chain Mount Kit
Mercury Thermometer, Celsius, -10 to 200°C
Mercury Thermometer, Fahrenheit, -30 to 120°F
Accu4 Spare Chamber with Plumbing, High Pressure
HACH