Приборы · Инжиниринг · Комплектация · ПЛК
0755-8222 1901 info@ainstru.com
HI5522-02

HI5522-02

实验室研究级台式pH/mV/ISE和EC/TDS/盐度/电阻率计

HI5522是一款先进的研究级台式pH/mV/ISE/EC/TDS/盐度/电阻率计,可通过大彩色LCD、电容触摸键和用于计算机连接的USB端口进行完全定制。HI5522具有丰富的功能,包括ISE增量方法、数据记录、报警限值、全面的GLP等,同时保留了常规操作专用键和引导用户完成设置选项的虚拟键的简单使用。HI5522通过独有的Hanna Instruments CAL Check™功能确保pH测量的可信度,该功能可提醒用户校准过程中的潜在问题,包括缓冲液是否被污染或探针是否需要清洁。
电容式触摸键盘
两个独立通道
ISE浓缩装置和增量方法的选择

SKU HI5522-02
Product Name Laboratory Research Grade Benchtop pH/mV/ISE and EC/TDS/Salinity/Resistivity Meter - HI5522
Quote Required Yes
pH Range -2.0 to 20.0 pH / -2.00 to 20.00 pH / -2.000 to 20.000 pH
pH Resolution 0.1 pH / 0.01 pH / 0.001 pH
pH Accuracy ±0.1 pH / ±0.01 pH / ±0.002 pH ± 1LSD
pH Calibration Up to five-point calibration, eight standard buffers available (pH 1.68, 3.00, 4.01, 6.86, 7.01,9.18, 10.01, 12.45), and five custom buffers
pH Temperature Compensation automatic or manual from -20.0 to 120.0 °C
mV Range ±2000.0 mV
mV Resolution 0.1 mV
mV Accuracy ±0.2 mV ± 1LSD
mV Relative Offset ±2000.0 mV
ISE Range e.g. 10-7 to 10M/ 0.005 to 105 ppm/ 5·10–7 to 5·107 conc
ISE Resolution 1 conc./ 0.1 conc. / 0.01 conc. / 0.001 conc
ISE Accuracy ±0.5% (monovalent ions); ±1% (divalent ions)
ISE Calibration Up to five-point calibration, seven fixed standard solutions available for each measurement unit, and five custom solutions
EC Range 0.000 to 9.999 μS/cm, 10.00 to 99.99 μS/cm, 100.0 to 999.9 μS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm actual EC*
EC Resolution 0.001 μS/cm, 0.01 μS/cm, 0.1 μS/cm, 1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm
EC Accuracy ±1% of reading (±0.01 μS/cm)
EC Calibration automatic standard recognition (0.000 μS/cm, 84.00 μS/cm, 1.413 mS/cm, 5.000 mS/cm, 12.88 mS/cm, 80.00 mS/cm, 111.8 mS/cm) or user standard; single point or multi-point calibration
TDS Range 0.000 to 9.999 ppm, 10.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 ppt, 10.00 to 99.99 ppt, 100.0 to 400.0 ppt actual TDS* (with 1.00 factor)
TDS Resolution 0.001 ppm, 0.01 ppm, 0.1 ppm, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt
TDS Accuracy ±1% of reading (±0.01 ppm)
EC/TDS Temperature Compensation Disabled / Linear / Non linear (natural water)
Temperature Correction Coefficient 0.00 to 10.00 %/°C
Reference Temperature 5.0 °C to 30.0 °C
Cell Constant 0.0500 to 200.00
Cell Type 4 cells
Salinity (% NaCl) Range 0.0 to 400.0%
Salinity (% NaCl) Resolution 0.01%
Salinity (% NaCl) Accuracy ±1% of reading
Salinity (PSU) Range 0.00 to 42.00 psu
Salinity (PSU) Resolution 0.01
Salinity (PSU) Accuracy ±1% of reading
Salinity (PSU) Calibration one-point (with HI7037 standard)
Salinity Range 0.00 to 80.00 ppt
Salinity Resolution (ppt) 0.01
Salinity Accuracy (ppt) ±1% of reading
Resistivity Range 1.0 to 99.9 Ω•cm; 100 to 999 Ω•cm; 1.00 to 9.99 KΩ•cm; 10.0 to 99.9 KΩ•cm; 100 to 999 KΩ•cm; 1.00 to 9.99 MΩ•cm; 10.0 to 100.0 MΩ•cm
Resistivity Resolution 0.1 Ω•cm; 1 Ω•cm; 0.01 KΩ•cm; 0.1 KΩ•cm; 1 KΩ•cm; 0.01 MΩ•cm; 0.1 MΩ•cm*
Resistivity Accuracy ±2% of reading (±1 Ω•cm)
Temperature Range -20.0 to 120.0 ºC**; -4.0 to 248.0 ºF**; 253.15 to 393.15 K**
Temperature Resolution 0.1 °C, 0.1 °F, 0.1 K
Temperature Accuracy ±0.2 °C; ±0.4 °F; ±0.2 K (without probe)
Input Channels 2 (pH/mV/ISE; Conductivity/Resistivity/TDS/ Salinity)
pH Electrode HI1131B glass body pH electrode with BNC connector and 1 m (3.3’) cable (included)
EC/TDS Probe HI76312 platinum, four-ring EC/TDS probe with and 1 m (3.3’) cable (included)
Temperature Probe HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included)
Logging Type Automatic, Manual, AutoHold
Logging Memory Up to 100 lots, 50,000 records max/lot; maximum 100,000 data points / channel
Logging Interval 14 selectable between 1 second and 180 minutes
Connectivity Opto-isolated USB
GLP yes (GLP Channel 1
Calibration Timeout Reminder yes
USP Compliant USP <645> Application
Profiles Up to 10 (5 for each)
Display Color Graphic LCD 240 x 320 pixels, 8 keys capacitive touch keyboard
Power Supply 12 VDC adapter (included)
Environment 0 to 50°C (32 to 122°F; 273 to 323 K), RH max 95% non-condensing
Dimensions 160 x 231 x 94 mm (6.3 x 9.1 x 3.7”)
Weight 1.2 Kg (2.6 lbs)
Ordering Information HI5522 is supplied with HI1131 pH electrode, HI7662-T temperature probe, HI76312 conductivity probe, HI76404W electrode holder, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI700601 electrode cleaning solution sachet (2), HI7082 3.5M KCl electrolyte solution (30 mL), 12 VDC power adapter and instructions.
Notes *Uncompensated conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation.