Product Overview
-
Multiparameter Benchtop Photometer and pH Meter - HI83300
-
Order No.: HI83300-02
-
HI83300 is a compact multiparameter photometer for laboratory or field use. This photometer is one of the most advanced photometers available, featuring an innovative optical design that utilizes a reference detector and focusing lens to eliminate errors caused by light source variations and glass cuvette defects. The meter features 60 different programmed methods, measures 37 key water quality parameters, and also provides an absorbance measurement mode for performance verification and for users who wish to develop their own concentration versus absorbance curves.
-
To save valuable laboratory benchtop space, HI83300 doubles as a professional pH meter with its digital pH/temperature electrode input. Now one meter can be used for both photometric and pH measurements.
-
Advanced optical system
-
Unmatched performance in a benchtop photometer
-
Digital pH electrode input
-
Save valuable benchtop space with a meter that serves as both a photometer and a laboratory pH meter
-
Absorbance measurement mode
-
Allows performance verification using CAL Check standards
-
From aluminum to zinc, the HI83300 benchtop photometer measures 37 different key water quality parameters using 60 different methods. This photometer features an innovative optical system that uses LEDs, narrow band interference filters, focusing lens, and a silicon photodetector for absorbance measurement and a reference detector to maintain light source consistency, ensuring accurate and repeatable photometric readings every time.
-
The digital pH electrode input allows users to measure pH with a conventional glass electrode. The digital pH electrode has a built-in microchip in the probe that stores all calibration information. Storing calibration information in the probe allows for hot swapping of pH electrodes without the need for recalibration. All pH measurements are automatically compensated for temperature changes via a built-in thermistor located at the tip of the sensing bulb for fast and accurate temperature measurement.
-
HI83300 offers an absorbance measurement mode that allows the use of CAL Check standards to verify system performance. The absorbance mode allows users to select one of five wavelengths of light (420 nm, 466 nm, 525 nm, 575 nm, and 610 nm) to measure and plot their own concentration versus absorbance relationships. This is useful for users employing their own chemical methods and for educators teaching the concept of absorbance using the Beer-Lambert law.
-
Two USB ports are provided for data transfer to a flash drive or computer and for use as a power supply for the meter. For added convenience and portability, the meter can also operate on an internal 3.7 VDC lithium polymer rechargeable battery.
| SKU | HI83300-02 |
|---|---|
| Product Name | Multiparameter Benchtop Photometer and pH meter - HI83300 |
| Quote Required | Yes |
| pH Range | Photometer |
| pH Resolution | Photometer |
| pH Accuracy | Photometer |
| pH Calibration | Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01) |
| pH Temperature Compensation | Automatic (-5.0 to 100.0 oC; 23.0 to 212.0 oF); limits reduced based on the pH electrode used |
| pH CAL Check (electrode diagnostics) | clean electrode and check buffer/check probe displayed during calibration |
| pH Method | Photometer |
| pH-mV Range | ±1000 mV |
| pH-mV Resolution | 0.1 mV |
| pH-mV Accuracy | ±0.2 mV |
| Dissolved Oxygen Range | 0.0 to 10.0 mg/L (as O2) |
| Dissolved Oxygen Resolution | 0.1 mg/L |
| Dissolved Oxygen Accuracy | ±0.4 mg/L ±3% of reading |
| Dissolved Oxygen Measurement Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method |
| Absorbance Range | 0.000 to 4.000 Abs |
| Absorbance Resolution | 0.001 Abs |
| Absorbance Accuracy | +/-0.003Abs @ 1.000 Abs |
| Alkalinity Range | Freshwater |
| Alkalinity Resolution | 1 mg/L |
| Alkalinity Accuracy | ±5 mg/L ±5% of reading |
| Alkalinity Method | Colorimetric method |
| Aluminum Range | 0.00 to 1.00 mg/L (as Al3+) |
| Aluminum Resolution | 0.01 mg/L |
| Aluminum Accuracy | ±0.04 mg/L ±4% of reading |
| Aluminum Method | Adaptation of the aluminon method |
| Ammonia Range | Low Range |
| Ammonia Resolution | 0.01 mg/L; 0.1 mg/L |
| Ammonia Accuracy | Low Range |
| Ammonia Method | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method |
| Anionic Surfactants Range | 0.00 to 3.50 mg/L (as SDBS) |
| Anionic Surfactants Resolution | 0.01 mg/L |
| Anionic Surfactants Accuracy | ±0.04 mg/L ±3% of reading |
| Anionic Surfactants Method | Adaptation of the USEPA method 425.1 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 5540C, Anionic Surfactants as MBAS. |
| Bromine Range | 0.00 to 8.00 mg/L (as Br2) |
| Bromine Resolution | 0.01 mg/L |
| Bromine Accuracy | ±0.08 mg/L ±3% of reading |
| Bromine Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method. |
| Calcium Range | Freshwater |
| Calcium Resolution | 1 mg/L |
| Calcium Accuracy | Freshwater |
| Calcium Method | Freshwater |
| Chloride Range | 0.0 to 20.0 mg/L (as Cl⁻) |
| Chloride Resolution | 0.1 mg/L |
| Chloride Accuracy | ±0.5 mg/L ±6% of reading at 25 °C |
| Chlorine Dioxide Range | 0.00 to 2.00 mg/L (as ClO2) |
| Chlorine Dioxide Resolution | 0.01 mg/L |
| Chlorine Dioxide Accuracy | ±0.10 mg/L ±5% of reading |
| Chlorine Dioxide Method | Adaptation of the Chlorophenol Red method. |
| Free Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) Ultra Low Range |
| Free Chlorine Resolution | 0.01 mg/L Ultra Low Range |
| Free Chlorine Accuracy | ±0.03 mg/L ±3% of reading Ultra Low Range |
| Total Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) Ultra Low Range |
| Total Chlorine Resolution | 0.01 mg/L Ultra Low Range |
| Total Chlorine Accuracy | ±0.03 mg/L ±3% of reading Ultra Low Range |
| Chlorine Method | Adaptation of the EPA 330.5 DPD method Free Chlorine (ULR) & Total Chlorine (UHR) |
| Chromium, Hexavalent Range | Low Range |
| Chromium, Hexavalent Resolution | 1 μg/L |
| Chromium, Hexavalent Accuracy | Low Range |
| Chromium, Hexavalent Method | Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method. |
| Color, Water Range | 0 to 500 PCU (Platinum Cobalt Units) |
| Color, Water Resolution | 1 PCU |
| Color, Water Accuracy | ±10 PCU ±5% of reading |
| Color, Water Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method. |
| Copper Range | Low Range |
| Copper Resolution | 0.001 mg/L; 0.01 mg/L |
| Copper Accuracy | Low Range |
| Copper Method | Adaptation of the EPA bicinchoninate method |
| Cyanuric Acid Range | 0 to 80 mg/L (as CYA) |
| Cyanuric Acid Resolution | 1 mg/L |
| Cyanuric Acid Accuracy | ±1 mg/L ±15% of reading |
| Cyanuric Acid Method | Adaptation of the turbidimetric method |
| Fluoride Range | Low Range |
| Fluoride Resolution | 0.01 mg/L; 0.1 mg/L |
| Fluoride Accuracy | Low Range |
| Fluoride Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, SPADNS method |
| Hardness, Total Range | Low Range |
| Hardness, Total Resolution | 1 mg/L |
| Hardness, Total Accuracy | Low Range |
| Hardness, Total Method | Adaptation of the EPA recommended method 130.1 |
| Hardness, Calcium Range | 0.00 to 2.70 mg/L (as CaCO3) |
| Hardness, Calcium Resolution | 0.01 mg/L |
| Hardness, Calcium Accuracy | ±0.11 mg/L ±5% of reading |
| Hardness, Calcium Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method |
| Hardness, Magnesium Range | 0.00 to 2.00 mg/L (as CaCO3) |
| Hardness, Magnesium Resolution | 0.01 mg/L |
| Hardness, Magnesium Accuracy | ±0.11 mg/L ±5% of reading |
| Hardness, Magnesium Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, EDTA Colorimetric method |
| Hydrazine Range | 0 to 400 μg/L (as N2H4) |
| Hydrazine Resolution | 1 μg/L |
| Hydrazine Accuracy | ±4% of full scale reading |
| Hydrazine Method | Adaptation of the ASTM Manual of Water and Environmental Technology, method D1385-88, p-Dimethylaminobenzaldehyde method |
| Iodine Range | 0.0 to 12.5 mg/L (as I2) |
| Iodine Resolution | 0.1 mg/L |
| Iodine Accuracy | ±0.1 mg/L ±5% of reading |
| Iodine Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method |
| Iron Range | Low Range |
| Iron Resolution | 0.001 mg/L; 0.01 mg/L |
| Iron Accuracy | Low Range |
| Iron Method | Low Range |
| Magnesium Range | 0 to 150 mg/L (as Mg2+) |
| Magnesium Resolution | 1 mg/L |
| Magnesium Accuracy | ±5 mg/L ±3% of reading |
| Magnesium Method | Adaptation of the Calmagite method |
| Manganese Range | Low Range |
| Manganese Resolution | 1 μg/L; 0.1 mg/L |
| Manganese Accuracy | Low Range |
| Manganese Method | Low Range |
| Molybdenum Range | 0.0 to 40.0 mg/L (as Mo6+) |
| Molybdenum Resolution | 0.1 mg/L |
| Molybdenum Accuracy | ±0.3 mg/L ±5% of reading |
| Molybdenum Method | Adaptation of the mercaptoacetic acid method |
| Nickel Range | Low Range |
| Nickel Resolution | 0.001 mg/L; 0.01 g/L |
| Nickel Accuracy | Low range |
| Nickel Method | Low Range |
| Nitrate Range | 0.0 to 30.0 mg/L (as NO3-- N) |
| Nitrate Resolution | 0.1 mg/L |
| Nitrate Accuracy | ±0.5 mg/L ±10% of reading |
| Nitrate Method | Adaptation of the cadmium reduction method |
| Nitrite Range | Freshwater Low Range |
| Nitrite Resolution | Freshwater |
| Nitrite Accuracy | Freshwater Low Range |
| Nitrite Method | Adaptation of the EPA Diazotization method 354.1 |
| Oxygen, Scavenger Range | 0 to 1000 μg/L (as DEHA) 0.00 to 1.50 mg/L (as Carbohydrazide) 0.00 to 2.50 mg/L (as Hydroquinone) 0.00 to 4.50 mg/L (as ISO-ascorbic acid) |
| Oxygen, Scavenger Resolution | 1 μg/L (DEHA); 0.01 mg/L |
| Oxygen, Scavenger Accuracy | ±5 μg/L ±5% of reading (DEHA) |
| Oxygen, Scavenger Method | Adaptation of the iron reduction method |
| Ozone Range | 0.00 to 2.00 mg/L (as O3) |
| Ozone Resolution | 0.01 mg/L |
| Ozone Accuracy | ±0.02 mg/L ±3% of reading |
| Ozone Method | Colorimetric DPD Method |
| Phosphate Range | Freshwater Low Range |
| Phosphate Resolution | Freshwater |
| Phosphate Accuracy | Freshwater Low Range |
| Phosphate Method | Freshwater Low Range |
| Potassium Range | 0.0 to 20.0 mg/L (as K) |
| Potassium Resolution | 0.1 mg/L |
| Potassium Accuracy | ±3.0 mg/L ±7% of reading |
| Potassium Method | Adaptation of the Turbidimetric Tetraphenylborate method |
| Silica Range | Low Range |
| Silica Resolution | 0.01 mg/L; 1 mg/L |
| Silica Accuracy | Low Range |
| Silica Method | Low Range |
| Silver Range | 0.000 to 1.000 mg/L (as Ag) |
| Silver Resolution | 0.001 mg/L |
| Silver Accuracy | ±0.020 mg/L ±5% of reading |
| Silver Method | Adaptation of the PAN method |
| Sulfate Range | 0 to 150 mg/L (as SO42-) |
| Sulfate Resolution | 1 mg/L |
| Sulfate Accuracy | ±5 mg/L ±3% of reading |
| Sulfate Method | Turbidimetric - Sulfate is precipitated with barium chloride crystals |
| Zinc Range | 0.00 to 3.00 mg/L (as Zn) |
| Zinc Resolution | 0.01 mg/L |
| Zinc Accuracy | ±0.03 mg/L ±3% of reading |
| Zinc Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method |
| Input Channels | 1 pH electrode input and 5 photometer wavelengths |
| pH Electrode | digital pH electrode (not included) |
| Photometer/Colorimeter Light Source | 5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters |
| Photometer/Colorimeter Light Detector | silicon photodetector |
| Bandpass Filter Bandwidth | 8 nm |
| Bandpass Filter Wavelength Accuracy | ±1 nm |
| Cuvette Type | round, 24.6 mm |
| Number of Methods | 128 max. |
| GLP | calibration data for connected pH electrode |
| Display | 128 x 64 pixel LCD with backlight |
| Logging Type | log on demand with user name and sample ID optional input |
| Logging Memory | 1000 readings |
| Connectivity | USB-A host for flash drive; micro-USB-B for power and computer connectivity |
| Power Supply | 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included) |
| Battery Type/Life | 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement |
| Environment | 0 to 50.0 oC (32 to 122.0 oF); 0 to 95% RH, non-condensing |
| Weight | 1.0 kg (2.2 lbs.) |
| Dimensions | 206 x 177 x 97 mm (8.1 x 7.0 x 3.8") |
| Ordering Information | HI83300 is supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual. |
Кейсы
Городские сточные воды — Hanna: пример комплектации со склада
О проекте
Объект находится на входную, биологическую и выходную стадии городских очистных. Клиент хотел корректно оснастить существующее производство; позиция — Hanna, цель — портативные и лабораторные измерения воды приборами Hanna.
Технические вызовы
Реальная проблема в следующем: высокое содержание взвесей и коррозионных компонентов, большие суточные колебания расхода, и при этом показатели стоков колеблются, датчики зарастают илом и биоплёнкой. Всё это означает, что для конфигурации Hanna главное — подбор по набору параметров, точности и портативности.
Конфигурация
Поставка и монтаж организованы через «склад»: комплектация со склада с быстрой поставкой. На объекте это означало подбор по набору параметров, точности и портативности; построчно сверялось: стабильное соответствие нормам сброса класса 1A и передача данных онлайн.
Работа после ввода
После пусконаладки результат — результат сразу на месте, меньше повторных лабораторных проверок; показания теперь используются для настройки процесса.
Пищевая промышленность — Hanna: пример комплектации со склада
Ситуация на объекте
Клиент работает в отрасли «Пищевая промышленность», производство расположено на участки дозирования, пастеризации, розлива и CIP-мойки. Главная задача — портативные и лабораторные измерения воды приборами Hanna.
В чём сложность
Сложность в том, что сиропы, молочные продукты и напитки, склонные к отложениям, при жёстких санитарных требованиях; кроме того, частые мойки требуют приборов, стойких к горячим щелочным циклам CIP/SIP. Поэтому ключевое здесь — подбор по набору параметров, точности и портативности.
Что мы сделали
Мы использовали подход «склад» — комплектация со склада с быстрой поставкой. Основа реализации — подбор по набору параметров, точности и портативности; поэтапно проверялось: требования пищевой безопасности и гигиенического исполнения 3-A / EHEDG.
Результат работы
В эксплуатации система ведёт себя как задумано — результат сразу на месте, меньше повторных лабораторных проверок, а объём текущего обслуживания заметно снизился.
Химия и фармацевтика — Hanna: пример комплектации со склада
Контекст применения
реакторы, мерные и товарные резервуары — типовые условия отрасли «Химия и фармацевтика». Проект: Hanna; главное — портативные и лабораторные измерения воды приборами Hanna.
Ограничения процесса
коррозия и взрывозащита сокращают срок службы приборов и удорожают простои, а условия «сильно агрессивные среды с растворителями и кристаллизацией, частично во взрывоопасных зонах» только усиливают это. Поэтому конфигурация Hanna должна включать подбор по набору параметров, точности и портативности.
Подход к поставке
Формат поставки — «склад»: комплектация со склада с быстрой поставкой. В работе главное — подбор по набору параметров, точности и портативности, а также требования: сертификация Ex и прослеживаемость данных по GMP.
Результаты на объекте
Эксплуатация наглядно показывает результат — результат сразу на месте, меньше повторных лабораторных проверок; доверие к точке измерения заметно выросло.
Энергетика — Hanna: пример комплектации со склада
О клиенте
Типовая площадка в отрасли «Энергетика» — питательная вода котлов, конденсат турбин, системы серо- и азотоочистки. Точка измерения — Hanna; основной акцент — портативные и лабораторные измерения воды приборами Hanna.
Ключевые трудности
Условия: высокие температура и давление, пар и минерализованная вода; вдобавок дрейф при высоких температуре и давлении, крайне дорогие внеплановые остановы — типовое исполнение здесь долго не выдерживает. Решение опирается на следующее: подбор по набору параметров, точности и портативности.
Реализация
Работы шли по каналу «склад»: комплектация со склада с быстрой поставкой. Техническая основа — подбор по набору параметров, точности и портативности; также учтено: соответствие отраслевым нормам и показателям готовности блока.
Выгода клиента
На практике это означает результат сразу на месте, меньше повторных лабораторных проверок, что снизило нагрузку на ремонтный персонал.
Похожие
HANNA InstrumentsПортативный кондуктометр с несколькими диапазонами
HANNA InstrumentsПортативный фотометр аммонийного азота низкого диапазона
HANNA InstrumentsИзмеритель солёности для морских аквариумов
HANNA InstrumentsПортативный турбидиметр с несколькими диапазонами
HANNA InstrumentsПортативный анализатор цветности мёда
HANNA InstrumentsПортативный фотометр свободного хлора
HANNA InstrumentsПортативный фотометр остаточного и общего хлора
HANNA Instruments