Приборы · Инжиниринг · Комплектация · ПЛК
0755-82221901 sales@ainstru.com

Proline Promag 10D 诊断代码 321

未标注

TOL. COIL CURR.

诊断行为
Alarm
十六进制码
—
手册 / 页码
BA00082DEN_1917.pdf · P78

可能原因

Sensor: Coil current is out of tolerance.

维修指导

  1. # Warning! Switch off power supply before manipulating the coil current cable, coil current cable connector or measuring electronics boards! Remote version:
  2. Check wiring of terminals 41/42 → 46
  3. Check coil current cable connector. Compact and remote version: Replace measuring electronics boards if necessary
← 111 CHECKSUM TOTAL.
351 CURRENT RANGE →

同机型同类代码(信号 ?)

001 CRITICAL FAIL. 011 AMP HW EEPROM 012 AMP SW EEPROM 101 GAIN ERROR AMP 111 CHECKSUM TOTAL. 351 CURRENT RANGE
该代码通常对应可更换部件
按经验,报 321 常与传感器、电缆、电子模块或密封件相关。我们可以按您的订货型号核对备件,并提供现场处理建议。
备件询价 / 技术支持 返回全部 16 条代码