The Liquiline Control CDC90 plug & play system automates Memosens pH and ORP measuring points in all industries. It cleans, calibrates and adjusts up to two sensors automatically, reducing maintenance effort, improving working safety in harmful areas and helping to ensure the yield and quality of your product. It can easily be integrated into existing plant infrastructures and allows convenient remote control of your measuring points via any process control system or mobile device.
Compare

pH

Measuring Principle Potentiometric
Application Increased occupational safety by minimizing maintenance and servicing measures.
Customizable cleaning and calibration cycles for up to two measuring points.
Minimization of operating costs and buffer and cleaning agent consumption through smart media distribution and dosing.
Installation Complete automated cleaning and validation system including a selectable retractable holder
Characteristic Fully assembled system with powder-coated or stainless steel pneumatic control unit.
All non-ex applications
Measuring principle Automated cleaning and calibration of Memosens pH and ORP sensors installed in retractable assemblies.
Design Complete cleaning and calibration system including dosing pumps and canisters for buffer and cleaning solutions for connection to various retractable assemblies and sensors
Dimension 268 x 1170 x 770 mm (l x h x w)
10.55 x 46.06 x 28.74 inch (l x h x w)
Connection DCS/PLS connection on Modbus TCP, Ethernet IP, Profinet, ProfibusDP and current in/ output
Input 1 to 2x Memosens digital sensor
1x 0/4 to 20mA input
4x digital input
Output / communication DCS/PLS connection via Modbus TCP, EtherNet/IP, PROFINET, PROFIBUS DP, digital and analog signals
Certificates, declarations and approvals UKCA, CSA general purpose

The Liquiline Control CDC90 is plug & play and automatically performs all tasks at Memosens pH and ORP measuring points in all industries. The system automatically performs cleaning, verification, calibration and adjustment of up to two sensors, reducing maintenance workload, improving workability in harmful conditions and helping to increase product yield and quality. The system can be easily integrated into existing plants, and measuring points can be conveniently controlled remotely via any process control system or remote device.

Benefits
Optimized cleaning and calibration cycles provide reliable, reproducible measured values, helping you to increase product yield and quality while reducing raw material consumption.
In the event of electrode clogging or contamination, the system automatically starts cleaning and calibration, always providing reliable measured values and ensuring product safety.
Minimizes maintenance workload at hard-to-reach positions, under demanding conditions or in harmful environments, increasing the safety of service personnel.
Offers various certified communication methods for seamless integration into your process control system, e.g.: 0/4…20mA, PROFIBUS DP, Modbus TCP, EtherNet/IP, PROFINET and OPC UA as well as web server technology.
Pre-configured cleaning and calibration programs enable fast startup and commissioning of the Liquiline Control CDC90 at low operating cost.

Applications
The Liquiline Control CDC90 automatically performs all tasks at pH and ORP measuring points in all industries. Suitable for the following applications:
Electrodes must be cleaned and/or calibrated frequently.
Measuring points are difficult to access.
Measuring points are in harmful environments.
The process requires repeated high-precision pH measurement.
The process requires frequent measurement.
Measuring point performance must be frequently checked and documented.

Technical documentation

   

Technical data

 

 

Documents/Manuals/Software

 

 

Accessories/Spare parts

 

Source: Endress+Hauserpublic information from the website

市政污水处理

市政污水处理污水处理厂 · PH应用案例

项目背景

市政污水处理客户在污水处理厂环节,长期面临「出水水质监测精度不足、仪表易受污染导致数据漂移,影响达标排放」的困扰,对PH的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的PH,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应市政污水处理复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。
食品饮料

食品饮料食品生产线 · PH应用案例

项目背景

食品饮料客户在食品生产线环节,长期面临「卫生级测量要求高,在线仪表需耐 CIP/SIP 清洗且精度稳定」的困扰,对PH的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的PH,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应食品饮料复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。
化工制药

化工制药反应釜与储罐 · PH应用案例

项目背景

化工制药客户在反应釜与储罐环节,长期面临「介质腐蚀性强、工况复杂,仪表需耐腐蚀且防爆」的困扰,对PH的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的PH,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应化工制药复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。
电力能源

电力能源电厂锅炉与汽机 · PH应用案例

项目背景

电力能源客户在电厂锅炉与汽机环节,长期面临「高温高压工况下测量不稳定,维护停机成本高」的困扰,对PH的稳定性与测量精度提出了更高要求。

解决方案

阿仪科技技术团队结合现场工况,为客户选配了适配的PH,并完成选型、安装、调试与整定的一站式交付。

实施效果

  • 测量精度与响应速度显著提升,数据稳定可靠;
  • 仪表适应电力能源复杂工况,维护量大幅降低;
  • 助力客户实现连续在线监测与降本增效,获得客户高度认可。