Order Code:COV61-1E1
Name:COV61 maintenance kits COS61/COS61x
Configuration
| Code | Name | Order | Specification |
|---|---|---|---|
| 010 | For Sensor | 1 | COS61 |
| 020 | Optical Spotcap | E | Sensor with black cap |
| 030 | Number of Optical Spotcaps | 1 | 1 piece |
Configuration Explained
This is the fluorescent cap maintenance kit for the COS61 dissolved oxygen sensor, one piece, black cap, for old sensors made before 7 January 2022 with serial number ending O00, when the cap ages and the reading drifts down, one change restores it, no new probe, one spare in stock takes little space, the cap body follows the production batch, do not fit a new batch cap on an old meter, fitting takes minutes, suitable for periodic replacement at dissolved oxygen points in wastewater plant aeration tanks.