Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

CPS61E-1009/0

CPS61E-1009/0
Product family: Memosens CPS61E
pH electrodeIn Stock (3)
Endress+Hauser CPS61E-1009/0
Memosens CPS61E;Approval Non-hazardous area;Electrode Type Hygienic version, zero point pH 7,0, temperature sensor NTC 30k;Application Range 0-14 pH, 0...100oC, 0,8...7 bar (abs) , sterilizable up to

Order Code:CPS61E-1009/0

Corresponding ext. order code:CPS61E-AA6NTP2

Name:Memosens CPS61E

Configuration

CodeNameOrderSpecification
010ApprovalAANon-hazardous area
020Electrode Type6Hygienic version, zero point pH 7,0, temperature sensor NTC 30k
030Application RangeN0-14 pH, 0...100oC, 0,8...7 bar (abs) , sterilizable up to 140oC
040Reference SystemTPPressurized, 1 ceramic junction, ion trap, 3 M KCl, Ag/AgCl
050Shaft Length2120mm

Configuration Explained

This CPS61E pH electrode is the hygienic version, dedicated to food and bioreactor tanks, range 0 to 14pH, temperature 0 to 100°C, 0.8 to 7bar, sterilisable up to 140°C, pressurised reference, 1 ceramic diaphragm plus ion trap, 3M KCl, Ag/AgCl reference, measuring rod 120 mm, non-hazardous area, suitable for online pH measurement in fermenters and dosing lines.

4 models · click to collapse