Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

PMP11-10N2/0

PMP11-10N2/0
Product family: Cerabar PMP11
Pressure transmitterIn Stock (19)
Endress+Hauser PMP11-10N2/0
Cerabar PMP11;Approval Non-hazardous area;Output 4-20mA;Electrical Connection Valve plug ISO4400 M16, IP65 NEMA Type 4X Encl.;Sensor Range 1bar/100kPa/15psi gauge, overload: 4bar/400kPa/60psi;Calibrat

Order Code:PMP11-10N2/0

Corresponding ext. order code:PMP11-AA1U1HBWJJ

Name:Cerabar PMP11

Configuration

CodeNameOrderSpecification
010ApprovalAANon-hazardous area
020Output14-20mA
040Electrical ConnectionUValve plug ISO4400 M16, IP65 NEMA Type 4X Encl.
070Sensor Range1H1bar/100kPa/15psi gauge, overload: 4bar/400kPa/60psi
090Calibration; UnitBSensor range; mbar/bar
110Process ConnectionWJJThread ISO228 G1/2 DIN3852,shape E,316L, flush mount

Configuration Explained

This is a PMP11 economy pressure transmitter, metallic sensor, dual use for gases and liquids, for non-hazardous areas, 24V two-wire 4-20mA output, Hirschmann connector M16, IP65, overpressure limit 1 bar, calibrated to the sensor range in mbar/bar, process connection G1/2 with DIN3852 sealing ring, 316L flush mounting, very responsive at low pressure, suitable for pressure monitoring of low pressure gas lines, water treatment and fans.

35 models · click to collapse