Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

19-BAS-002

19-BAS-002
Venturi Sampler w/ air and wash (2-Stream), for BioTector B7000/B7000i/B7000i Dairy
HACH 19-BAS-002
Combine the TOC Analysis Power of BioTector with the Sampling Ease of the Venturi Vacuum Sampler

产品概述

Combine the TOC Analysis Power of BioTector with the Sampling Ease of the Venturi Vacuum Sampler

The BioTector Venturi Vacuum Sampler is an air operated, vacuum-pumping system capable of drawing samples over long distances and depths to the BioTector B7000 series TOC analyzers. Specifically designed for highly charged and dirty wastewater, it comes standard with a self cleaning feature, which prevents cross-contamination or carry-over and offers multi-channel operation.

主要特点

  • Covers Long Distances with Multiple Channels:The Venturi Vacuum Sampler is a pump capable of drawing samples from a horizontal distance of up to 80 meters and a vertical distance of up to 7 meters. It can be equipped with 1, 2, 3, 4, or 6 channels allowing you to make upstream adjustments to a downstream analyser and requires pressurized air and hot water connections for operation.
  • Boasts a Simple, Worry-Free Design:The Venturi Vaccum Sampler creates a vaccum generated by a flow of pressurized air to draw sample content to the BioTector TOC analyzer. The Venturi Sampler is powered by and integrated next to the TOC analyzer, allowing for ease of installation and access. Unlike mechanical sample pumps, the Venturi Vacuum Sampler requires infrequent maintanance and attention, even with highly charged wastewater containing large amounts of things like oil, grease, or solids.
  • Requires Little Maintenance While Ensuring Sample Quality:After each measurement cycle the sample is pushed out of the Venturi Vacuum Sampler and associated sampling lines and all components that come in contact with the sample are backwashed with a hot water microbubble mixture. These components are then air dried and are ready to receive the next sample, avoiding any potential cross contamination or dilution and providing a truly representative sample for analysis.
  • Handles Low Flow with Ease:Where mechanical pumps would be in danger of damages by running dry, the Venturi Vacuum Sampler can be adjusted to low flow rates in the respective sampling channel without the need of a barrier system and comes with an automatic no sample fill fault indication.

技术参数

参数名称说明
Ambient Temperature5 - 40 °C
Dimensions (H x W x D)2010 mm x 390 mm x 160 mm
Housing MaterialPLC Enclosure: Polystyrene with glazed polycarbonate door
Instrument AirInstrument air: 6 bar (50 L/min flow rate when venturi pump is running at 6 bar) (~100 L/min flow rate when venturi sampler is purging at 3.5 bar) Air quality: -20 °C dew point (free of water, oil and dust)
MaterialSample chamber: PFA Fittings and valves: SS-316, EPDM, PTFE Venturi pump: SS-316 Sample tubing: PFA (1/2" OD x 3/8" ID)
Number of sample streams2
Operating SystemPLC with Keyboard Automatic 'no sample fill' fault indication Direct control from the BioTector through the 6 core cable
Power Consumption24 W (1 Amp 24 V DC), power supplied by BioTector analyzer
Requirement rinse water pressure3 bar (~10 L/h at 3 bar) Water can be heated up to 90 °C for hot water backwash Water regulator and/or water heater to be supplied by the user
Sample Inlet Temperature2 - 90 °C
Sample PressureAmbient pressure
Sample QualityOnly for non-foaming samples!
Sampling PrincipleVacuum sampling
Service Interval6 months (typical)
Suction Height7 m (at sea level; measured from the bottom of the sampler to the lowest liquid level)
Suction Hose LengthMax. distance of sample transport: 80 m
Weight20 - 30 kg
What's included?Sampler only. Hoses need to be ordered separately.

资料下载