产品概述
Hach’s CLF10 sc and CLT10 sc provide online amperometric solutions for real-time analysis of free or total chlorine in disinfection applications.
The CLF10 sc Free Chlorine Analyzer and CLT10 sc Total Chlorine Analyzer measure free or total chlorine continuously for real time process control. It is based on amperometric technology and is fully compatible with all of Hach's digital controllers. The CLF10 sc can be used in most municipal and industrial applications, and is best utilized where waste stream management is a constraint.
主要特点
- EPA-approved Free or Total Chlorine Analysis:The CLF10 sc and CLT10 sc chlorine analyzers are compliant with EPA Method 334.0, ensuring that you not only have precise process control, but reportable results as well.
- No Reagents to Replace, No Waste Stream:Hach’s amperometric chlorine technology does not use reagents, eliminating the need for a sanitary drain and saving you the cost of monthly reagent replacement.
- Chlorine Analysis Made Easy:The CLF10 sc and CLT10 sc are plug-and-play with Hach’s digital controllers, ensuring easy setup. Available with exclusive self-diagnostics, these instruments will alert the user to changes in the process or upcoming maintenance. Hach’s Cal Watch feature will provide a warning for pH or chlorine calibration changes, and a non-contacting flow sensor will notify users of insufficient sample flow.
: :
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | Free Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit) |
| Accuracy 2 | Free Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration) |
| Accuracy 3 | Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration) |
| Accuracy 4 | Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5 |
| Cable Connection | 5 pin, M12 connector |
| Cable Length | 1 m (between gateway and sc controller) |
| Calibration Method | 1-point or 2-point (zero and slope) calibration |
| Certifications | CE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits) Additional certifications when connected to an sc controller |
| Connection Drain Line | pH Flow Cell Outlet: 1/2-inch ID |
| Connection Sample Line | 1/4 inch OD |
| Controller | Single Digital |
| Controller Compatibility | sc controller platform |
| Description | CLT10sc with SC200 Signal Input, Combination pH Sensor |
| Diameter | 25 mm |
| Dimensions | Sensor Only: 195 |mm x 25 mm (7.68 in x 0.98 in) |
| Fitting Type | USA |
| Interferences | Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits. Total chlorine: chlorine dioxide, ozone and chalk deposits |
| Length | Sensor: 195 mm |
| Lower Limit of Detection (LOD) | 30 ppb (0.03 ppm) |
| Material | Corrosion-resistant materials, fully-submersible (stainless steel, PEEK, PVC, silicon rubber and polycarbonate) |
| Measurement Method | Reagentless, electrochemical, three-electrode amperometric system |
| Mounting | Panel or Wall Mounting |
| Operating Temperature Range | 0 - 45 °C |
| Parameter | Total Chlorine |
| pH Range | 4 - 9 |
| Power Requirements (Voltage) | 110 - 230 VAC |
| Pressure Range | 0.5 bar, no pressure impulses and/or vibrations |
| Range | 0 - 20 ppm Cl₂ |
| Repeatability | 30 ppb or 3%, whichever is greater |
| Response Time | Total Chlorine: 100 s or less for 90% change (T90) at a stable temperature and pH |
| Response Time | Free Chlorine: 140 s or less for 90% change (T90) at a stable temperature and pH |
| Sample Flow Rate | 30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour) |
| Sample Temperature | 5 - 45°C (41 - 113°F) |
| Storage Conditions | -20 °C to 60 °C dry |
| Weight | 13.83 kg Approximately 5.5 kg (12 lbs) Panel and empty panel-mounted components only |
| What's included? | SC200 Controller, Mounting Hardware, SC200 Basic User Manual, CL10sc Panel, 1 m digital extension cable, Panel Manual, Combination pH Sensor Manual, Chlorine Sensor Manual. |
应用领域
Discharge · Groundwater · Influent for Wastewater · Membrane Protection · Municipal Water · Regulations · Surface Water
资料下载
- Chlorination, Chloramination And Chlorine Measurement
- Reagentless Total Chlorine Probe for Wastewater Disinfection. Comparing the Hach® CLT10 sc and Wallace & Tiernan Micro/2000®
- Discovering the Unexpected in Disinfection Applications
- Introduction to Hach Services
- Chlorine Analyzer Customer Guide
- Máy phân tích Clo tổng & Clo tự do không cần hóa chất CLF10sc & CLT10sc
- CLF10sc & CLT10sc Free & Total Reagentless Chlorine Analyzers
- HACH SC200 Controller CSI Specification
Related Products
HACHRadiometer Analytical PHC3031-9 Combination pH Electrode for Penetration Applications (glass body, Ag/AgCl ref., screw cap)
Radiometer Analytical PHC4006-9 Combination Calomel pH Electrode (l=158 mm, d=6.5 mm, glass body, screw cap)
pHC2501-8 Combined pH Electrode, Red Rod, Double Junction, BNC plug (Radiometer Analytical)
HACHpH Electrode Storage Solution, 500 mL
HACHRadiometer Analytical PHC2601-8 Combination Red-Rod pH Electrode (sleeve junction, glass body, BNC)
Radiometer Analytical PHC2401-7 Combination Red-Rod pH Electrode (glass body, type 7 plug)
HACHRadiometer Analytical PHG201-8 Glass pH Electrode (d=7.5 mm, BNC)
HACH