产品概述
Allows you to monitor hydrogen in the power production process.
High Pressure Hydrogen TC sensor with external temperature sensor adaptor with nitrogen purge and protection cap. Sensor Cable to be ordered: 32505.03. Maximum pressure: 170 bar
主要特点
- Patented thermal conductivity (TC) sensor design:Selective and precise measurement of hydrogen
- Continuous measurements in gaseous or liquid samples:No specific sample preparation required
- Rugged construction can handle harsh plant conditions and high sample pressures:Flexibility of installation for better process control
- Very low maintenance:Limited exposure to radiations in nuclear power plant
- Very repeatable:Tight control of hydrogen in critical processes
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | (Sample temp. 20 - 50°C within ± 5°C of calibration temperature) The greater of Membrane 29561A : ±1% of reading, or ±2 ppb, or ±0.03 cc/kg, or ±1.5 bar Membrane 2952A : ±1% of reading, or ±8 ppb, or ±0.1 cc/kg, or ±6 mbar Membrane 2935A : ±1% of reading, or ±25 ppb, or ±0.4 cc/kg, or ±20 mbar |
| Accuracy 2 | (Sample temp. 0 - 50°C independent of calibration temperature) The greater of Membrane 2952A : ±3% of reading, or ±60 ppb, or ±0.6 cc/kg, or ±20 mbar Membrane 2935A : ±3% of reading, or ±150 ppb, or ±2.5 cc/kg, or ±50 mbar |
| Application | Power |
| Certifications | 2004/108/EC - EN 61326-1 |
| Cycle Time | 17 s |
| Flow Rate | Typical Flow Rate Membrane 29561A : 220 mL/min Membrane 2952A : 200 mL/min Membrane 2935A : 100 mL/min |
| Protection Class | IP65 |
| Radiation Resistance Level | Membrane 29561A : 105 rad Membrane 2952A : 108 rad Membrane 2935A : 108 rad |
| Range | Range at 25 °C Membrane 29561A : 0 - 2 ppm, or 0 - 25 cc/kg, or 0 - 1.5 bar Membrane 2952A : 0 - 10 ppm, or 0 - 120 cc/kg, or 0 - 6 bar Membrane 2935A : 0 - 20 ppm, or 0 - 220 cc/kg, or 0 - 12 bar |
| Response Time | 17 s |
| Sample Pressure | 0 - 170 bar |
| Temperature Range | -5 - 50 °C |
| Weight | 0.95 kg |
| What's included? | Sensor only. Controller, sensor cable, flow-chamber or process connection must be ordered separately. |
资料下载
- ORBISPHERE and Hydrogen Measurement for Stator Coolant; Hydrogen Measurement in Hydrogen-Cooled Turbo Generators
- Dissolved Oxygen and Hydrogen Analysis in Reactor Coolant Systems: Orbisphere 510 with Channel Compensation
- Analysis Of Waste-Gas / Off-Gas Oxygen and Hydrogen in Nuclear Power Plants
- Introduction to Hach Services
- Power Industry Laboratory & Process Analysis (JP)
- 有機溶媒用 溶存酸素計(DO計)
- Orbisphere Hydrogen Sensors
- User Manual : Orbisphere Model 31xxx Thermal Conductivity Sensors
Related Products
HACHNP6000sc Phosphate Analyzers
View →
HACHSC1000 relay module for DIN-rail mounting, 4 Relays, max. 240V
View →
A
HACHCOVER-SERVICE REPLACEMENT
View →
A
HACHStainless Steel Chain Mount Kit
View →
HACHNH6000sc Ammonia Analyzers
View →
A
HACHAccu4 Spare Chamber with Plumbing, High Pressure
View →
HACHEZ Series Arsenic Analyzers
View →
A
HACHInterface, 4-20mA, 10PPM FS, Out
View →