Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

31260HPDiscontinued

31260HP
Orbisphere High Pressure Hydrogen (H₂) TC sensor with nitrogen purge and protection cap (170 bar)Discontinued
HACH 31260HP
Allows you to monitor hydrogen in the power production process.

产品概述

Allows you to monitor hydrogen in the power production process.

High Pressure Hydrogen TC sensor with external temperature sensor adaptor with nitrogen purge and protection cap. Sensor Cable to be ordered: 32505.03. Maximum pressure: 170 bar

主要特点

  • Patented thermal conductivity (TC) sensor design:Selective and precise measurement of hydrogen
  • Continuous measurements in gaseous or liquid samples:No specific sample preparation required
  • Rugged construction can handle harsh plant conditions and high sample pressures:Flexibility of installation for better process control
  • Very low maintenance:Limited exposure to radiations in nuclear power plant
  • Very repeatable:Tight control of hydrogen in critical processes

技术参数

参数名称说明
Accuracy(Sample temp. 20 - 50°C within ± 5°C of calibration temperature) The greater of Membrane 29561A : ±1% of reading, or ±2 ppb, or ±0.03 cc/kg, or ±1.5 bar
Membrane 2952A : ±1% of reading, or ±8 ppb, or ±0.1 cc/kg, or ±6 mbar
Membrane 2935A : ±1% of reading, or ±25 ppb, or ±0.4 cc/kg, or ±20 mbar
Accuracy 2(Sample temp. 0 - 50°C independent of calibration temperature) The greater of Membrane 2952A : ±3% of reading, or ±60 ppb, or ±0.6 cc/kg, or ±20 mbar
Membrane 2935A : ±3% of reading, or ±150 ppb, or ±2.5 cc/kg, or ±50 mbar
ApplicationPower
Certifications2004/108/EC - EN 61326-1
Cycle Time17 s
Flow RateTypical Flow Rate Membrane 29561A : 220 mL/min
Membrane 2952A : 200 mL/min
Membrane 2935A : 100 mL/min
Protection ClassIP65
Radiation Resistance LevelMembrane 29561A : 105 rad
Membrane 2952A : 108 rad
Membrane 2935A : 108 rad
RangeRange at 25 °C
Membrane 29561A : 0 - 2 ppm, or 0 - 25 cc/kg, or 0 - 1.5 bar
Membrane 2952A : 0 - 10 ppm, or 0 - 120 cc/kg, or 0 - 6 bar
Membrane 2935A : 0 - 20 ppm, or 0 - 220 cc/kg, or 0 - 12 bar
Response Time17 s
Sample Pressure0 - 170 bar
Temperature Range-5 - 50 °C
Weight0.95 kg
What's included?Sensor only. Controller, sensor cable, flow-chamber or process connection must be ordered separately.