Hach BioTector B3500e TOC Analyzer

Hach BioTector B3500e TOC Analyzer
Hach BioTector B3500e TOC Analyzer
HACH 3500e
Configurable Models 4 models
Hach BioTector B3500e TOC Analyzer

产品概述

Highest reliability in TOC analysis for water control meets low cost of ownership.

The Hach BioTector B3500e, with patented Two Stage, Advanced Oxidation Technology, provides accurate results you can be confident in, and gives you maximum reliability and uptime for effluent waste water applications. The analyzer requires maintenance only twice a year for standard items like pump tube replacement and calibration, making it the most affordable TOC analyzer option available. The Hach BioTector B3500e analyzer can be configured as: 1. A TIC & TOC (NPOC) system to measure the Total Inorganic Carbon (TIC) and Total Organic Carbon (TOC) content of a liquid sample. The TOC result obtained from a TIC & TOC system represents the Non-Purgeable Organic Carbon (NPOC) because VOC (POC) compounds may partly escape from the sample during TIC stripping. The TIC & TOC system is suitable for samples that contain very low or insignificant concentrations of volatile organic materials. 2. A TC system to measure the Total Carbon (TC) content of a sample. The TC result obtained from a TC system represents the sum of TIC, NPOC and VOC content. 3. A VOC system to measure the TIC, TOC, TC and VOC contents of a sample. The TOC result in a VOC system is calculated from the TC and TIC measurements as TOC = TC - TIC. Therefore, the TOC result includes the volatiles present in the sample. In other words, the TOC result represents the sum of NPOC and VOC (POC) content. Because the analyzer body has one of the most compact analyzer footprints available, plants can free up critical wall space that can be used for other instruments. With maximum reliability and uptime along with the lowest operating costs, the B3500e provides immediate, direct savings and is the industry’s choice when it comes to TOC analyzis for wastwater effluent applications.

For more details, to discuss specific application requirements, or to request additional BioTector versions, please contact us. Our team is ready to help identify the right solution for your plant.

The B3500e is tailor-made for monitoring final wastewater effluent and discharges to meet TOC requirement in water regulations. Being regulatory compliant will save you significant penalty costs while protecting the local environment.

主要特点

  • Ensure environmental compliance.:The B3500e is tailor-made for monitoring final wastewater effluent and discharges to meet TOC requirement in water regulations. Being regulatory compliant will save you significant penalty costs while protecting the local environment.
  • Save with low ownership costs.:Requiring you to replace the pump tube and calibrate only twice a year, the Hach BioTector B3500e provides low operating expenses.
  • Worry-free TOC with smart design.:The B3500e comes with a built in self-cleaning sample tube and reactor. This enables the B3500e to deliver trustworthy results even if your water contains some level of fat, oil, greases and sludge or has moderate pH swings.
  • Information you can rely on.:Using BioTector’s internationally proven and patented Two Stage Advanced Oxidation technology, the B3500e delivers maximum uptime, reliability and accuracy.
  • Access your results from everywhere.:With the B3500e TOC results can be viewed at your desk, your home, or while you are on the move with the BioTector Network Control Unit remote access.

技术参数

参数名称说明
Ambient Temperature5 - 45 °C
Automatic cleaningYes
CalibrationAutomatic with Self-Cleaning Technology
CommunicationModbus RTU, Modbus TCP/IP & Profibus (when the Profibus option is selected, the digital output signals are sent through the Profibus converter with its specific communication protocol)
Cycle TimeTypically 7 minutes, 30 seconds
Data StoragePrevious 9,999 reaction data Previous 99 fault events
DisplayHigh contrast 40 character x 16 line backlit LCD with LED backlight
EExp / Hazardous LocationNo
Exceedance TrackingNo
Humidity5 - 85 % (non-condensing)
LanguagesEnglish
Measurement MethodInfrared measurement of CO2 after oxidation
Number of Channels1
Number of sample streamsSingle process stream and manual grab sample
Number of streamsSingle stream
OptionsGrab Sample, Cleaning, Sample Sensor
Oxidation MethodPatented Two-Stage Advanced Oxidation Process (TSAO) using Hydroxyl Radicals
ParameterMeasures TOC, TIC and TC
Particle SizeUp to 100 µm
Power Requirements (Hz)50/60 Hz
Power Requirements (Voltage)115 VAC
Range250 mg/L C (ppm)
Range SelectionAutomatic or manual range selection
RepeatabilityStandard range ± 3 % of reading or ± 0.45 mg/L C (ppm), whichever is greater
Sample Inlet Temperature2 - 60 °C
Service Interval6 month service intervals
User InterfaceMicrocontroller with membrane keyboard
Warranty12 months
Warranty1
Weight46 kg
What's included?Includes: B3500e analyser, Tubing, Fuses, Ferrules, Drain, Acid & Base Dip Tubes, C02 Filter & B3500e User Manual

应用领域

Discharge · Regulations · Water Reuse

同系列其他配置

型号
BEAAAA152AAA2Hach BioTector B3500e Online TOC Analyzer, 0 - 250 mg/L C, 1 stream, 1
BEABAA152AAA2Hach BioTector B3500e Online TOC Analyzer, 0 - 1000 mg/L C, 1 stream,
BEABAA152AAB2Hach BioTector B3500e Online TOC Analyser, 0 - 1000 ppm, 1 stream, 115

资料下载

4 models · click to collapse