产品概述
APA6000™ Low Range Hardness Analyzer
Measures low-range hardness using USEPA-approved calmagite chemistry. Range 50 to 10,000 ug/L. Accurately and continuously measures up to two sample streams. Operates unattended for one month. Makes your water softening system more efficient and less costly.
主要特点
- Accurate and Continuous Monitoring of Soluble Hardness:Hach's patented* APA 6000 Low-Range Hardness Analyzer accurately and continuously measures up to two sample streams for low levels (50 µg/L to 10 mg/L) of soluble hardness. Using a precise colorimetric method, the instrument performs an analysis every four minutes. It delivers continuous, accurate, and dependable performance with minimal cost for operation and maintenance. * U.S. Patent 5849592
- Optimize Softener Performance:The APA 6000 Low-Range Hardness Analyzer is designed specifically for reliable and continuous analysis at low concentrations. Use it to establish an automatic system for monitoring softener performance in industrial, ultrapure, and water treatment processes.
- Industrial and Ultrapure Water Applications:The APA 6000 Hardness Analyzer can be used to monitor both influent and effluent in industrial applications and water treatment processes including demineralizer effluent, boiler feed-water, boiler water, and process water. It is also appropriate for monitoring hardness in ultrapure processes used by pharmaceutical, electronic chip, and cosmetics manufacturers.
- Easy to Use:The interface of the APA 6000 Hardness Analyzer is menu-driven with on-screen prompting, self-test diagnostics, security provisions, and flexible data logging options. Simple menus (available in English, French, German or Spanish) guide the user through each step of operation.
- Automatic 2-Stream Analysis:Two separate sample streams can be monitored with the APA 6000 Hardness Analyzer when used with additional sample sequencing kits.
- Operation:
技术参数
| 参数名称 | 说明 |
|---|---|
| Height | 21 " (526 mm) |
| Width | (627 mm) |
| Accuracy | ± 5 % of reading or ± 0.03 mg/L whichever is greater |
| Alarm | Two SPDT relays with contacts rated for 5A resistive load at 230 Vac. Additional relays available with optional Signal Output Modules. |
| Certifications | NRTL certified to UL and CSA standards and CE approved |
| Cycle Time | 4 min |
| Drain Fitting | 3/4" NPT barbed hose fitting |
| Inlet Pressure | 0.5 - 30.0 psig (0.03 to 2.04 bar) |
| Length | 533 mm (522 mm) |
| Network Connectivity | AquaTrend™ network, using the Lonworks® protocol |
| Outputs | Two 4-20 mA outputs suitable for recorders or PID control. Ouput span programmable over any portion of the measuring range (130 V AC isolation from earth ground) |
| Parameter | Low Range Hardness |
| Power Requirements (Hz) | 50/60 Hz |
| Power Requirements (Voltage) | 90 - 240 V AC |
| Range | 0.05 - 10 mg/L hardness as CaCO3 |
| Repeatability | ± 3 % of reading or ± 0.03 mg/L, whichever is greater |
| Response Time | Less than 5 min for 90% response |
| Sample Flow Rate | 100 - 2000 mL/min max |
| Sample Temperature | 5 - 50 °C (41 - 122 °F) |
| Weight | 25.5 kg |
| What's included? | APA6000 LR Hardness analyzer, installation kit, one month's supply of reagents, a maintenance kit, Hach user manual. |
资料下载
- Complete Water Analysis for the UPSTREAM OIL & GAS INDUSTRY
- Introduction to Hach Services
- Complete Water Analysis Solutions for Power and Steam Generation
- Complete Water Analysis Solutions for Power and Steam Generation
- Soluciones completas de análisis de agua para Generación de energía y vapor
- Solutions complètes d’analyse de l’eau pour La production de vapeur et d’électricité
- Complete Water and Product Quality Analysis: Beverage Industry
- Complete Water Analysis Solutions for Power and Steam Generation
Related Products
HACHNP6000sc Phosphate Analyzers
View →
HACHSC1000 relay module for DIN-rail mounting, 4 Relays, max. 240V
View →
A
HACHCOVER-SERVICE REPLACEMENT
View →
A
HACHStainless Steel Chain Mount Kit
View →
HACHNH6000sc Ammonia Analyzers
View →
A
HACHAccu4 Spare Chamber with Plumbing, High Pressure
View →
HACHEZ Series Arsenic Analyzers
View →
A
HACHInterface, 4-20mA, 10PPM FS, Out
View →