产品概述
Repeatable dissolved ozone measurements.
Ozone analyzers perform super benefits. The analyzer is great for readings in low conductivity water with no interferences from oxidants or pH. An ion-selective membrane separates electrolyte from sample water, so no reagents are needed. All-inclusive, pre-assembled panel reduces installation expense. Plug & Play with digital controllers simplifies setup. Built-in flow regulation reduces need for additional accessories. Integral temperature sensor provides more accurate readings. Includes 2 years of typical maintenance parts.
主要特点
- Few Interferences from Oxidants or pH over a Wide Measurement Range:Bromine, chloramines, chlorine, chlorine dioxide, hydrogen peroxide, and pH do not interfere with ozone measurements using the Hach 9185sc Amperometric Ozone Sensor, which can measure from 0 to 2 mg/L ozone.
- Ideal for Low-Conductivity Samples:The 9185sc ozone system has a low minimum detection limit of 0.005 mg/L ozone. Because the amperometric cell is separated from the sample by the membrane and is immerged into an electrolytic medium, the 9185sc ozone sensor is suitable for use in water with low conductivity.
- Reliable Amperometric Measurements:The sensor used in the 9185sc system is a two-electrode amperometric sensor (gold cathode and silver anode) with a membrane selective to ozone. Membranes are pre-mounted on retaining caps to simplify and reduce maintenance time.
- Flexible Options:Options available for the 9185sc ozone system include: Acidification Unit, which can be used intermittently or continuously for cleaning and is fully programmable, and Intermittent Flow Unit, which is fully programmable and saves resources while limiting excess flow to drain.
- Easy Setup and Maintenance:The all-inclusive 9185sc ozone system comes pre-assembled on a panel. The membrane system of the sensor means there are no chemicals used for measurement. Maintenance is minimal and two years of typical maintenance items are included with the system.
- Compatible with Hach Multi-Sensor, Multi-Parameter Digital Controllers:The Hach 9185sc Amperometric Ozone Sensor can be used with Hach’s new sc100 or sc1000 Digital Controllers. The sc100 controller accepts up to two sensors. The sc1000 accepts up to eight sensors. Multiple controllers can be networked to accommodate many more sensors and parameters, reducing the cost per measuring point. Just plug in any Hach “plug and play” digital sensor and it’s ready to use without software configuration. “Plug and play” connectivity means there’s no complicated wiring or set up. Network the 9185sc ozone sensor with any of Hach’s digital sensors for measuring dissolved oxygen, turbidity, ORP, conductivity, and many other parameters.
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | 3 % or ± 10 ppb O3, whichever is greater |
| Calibration | Comparison of lab method with process sample |
| Calibration Interval | 2 months |
| Calibration Method | Electrically or with ozone-free water |
| Certifications | UL, CSA (certified by ETL), CE |
| Connection Drain Line | 1/2-in. I.D. (supplied) |
| Connection Sample Line | 1/4-in. O.D. |
| Cycle Time | 90 % in T<90sec |
| Flow | 14 L/h (200 - 250 mL/min) au-regulated by flow thru cell |
| Interferences | No interference from bromine, chloramines, chlorine, chlorine dioxide, or hydrogen peroxide |
| Lower Limit of Detection (LOD) | 5 ppb 0.005 mg/L or 0.005 mg/L O3 |
| Maintenance Interval | Measurement Cell: 6 months for membrane and electrolyte, typical |
| Material | Electrodes: gold cathode/silver anode |
| Measuring Principle | Amperometric/Membrane (electrode, membrane, electrolyte) |
| Mounting | Flat, vertical surface (panel, stand, etc.) |
| Operating Temperature Range | 0 - 45 °C |
| Pressure Range | 0.1 to 2 bar in flow cell |
| Range | 0 - 2 ppm (mg/L) ozone (O3) |
| Range | 0.005 - 2 ppm (0.005 mg/L - 2 mg/L) |
| Relative Humidity | 1 - 90 % non-condensing |
| Response Time | 90 % in less than 90 seconds |
| Sample Input | 0.25 " outer diameter |
| Sample Output | 0.5 " inner diameter |
| Sample Temperature | 2 - 45°C (35.6-113 F) |
| Ship Wt. (lbs) | 6.5 kg(14.3 lbs.) |
| Standard deviation | 1.0 |
| Temperature Compensation | Automatic over sample temperature range |
| Temperature Compensation | Automatic over sample temperature range |
| Weight | 6.5 kg (14.3 lbs) |
| What's included? | 9185sc Ozone Sensor, 9185sc User Manual, sc100 with Modbus, sc100 User Manual, sc100 Mounting Panel |