9187sc Chlorine Dioxide Analyzer with sc100 controllerDiscontinued
HACH 6043400
Ideal for drinking water plants, industrial rinsing, and cooling towers.

产品概述

Ideal for drinking water plants, industrial rinsing, and cooling towers.

Reagentless analysis of chlorine dioxide (ClO2). All-inclusive, pre-assembled panel reduces installation expense. Plug & Play with digital controllers simplifies setup. Built-in flow regulation reduces need for additional accessories. Integral temperature sensor provides more accurate readings. Includes 2 years of typical maintenance parts, reduces operating costs.

主要特点

  • Selective Membrane Avoids Interference from other species:The Hach 9187sc Amperometric Chlorine Dioxide sensor uses a membrane that is selective for chlorine dioxide species. There are no interferences from bromine, chlorine or hydrogen peroxide. The only potential interference is from ozone, which is rarely present in water treated with chlorine dioxide.
  • Flexible Options:Options available for the 9187sc ClO2 sensor include: Acidification Unit, which can be used intermittently or continuously for cleaning and is fully programmable, and Intermittent Flow Unit, which is fully programmable and saves resources while limiting excess flow to drain.
  • Easy Setup and Maintenance:The all-inclusive 9187sc ClO2 system comes pre-assembled on a panel. The membrane system of the sensor means there are no chemicals used for measurement. Maintenance is minimal and two years of typical maintenance items are included with the system.
  • Compatible with Hach Multi-Sensor, Multi-Parameter Digital Controllers:The Hach 9187sc ClO2 sensor can be used with Hach's new sc100 or sc1000 Digital Controllers. The sc100 controller accepts up to two sensors. The sc1000 accepts up to eight sensors. Multiple controllers can be networked to accommodate many more sensors and parameters, reducing the cost per measuring point. Just plug in any Hach "Plug & Play" digital sensor and it's ready to use without software configuration. "Plug & Play" connectivity means there's no complicated wiring or set up. Network the 9187sc ClO2 sensor with any of Hach's digital sensors for measuring dissolved oxygen, turbidity, ORP, conductivity, and many other parameters.

技术参数

参数名称说明
Accuracy±5 % or ±10 ppb ClO2, whichever is greater
CalibrationComparison of lab method with process sample
Calibration Interval2 months
Calibration MethodElectrically or with de-chlorinated water
CertificationsUL, CSA (certified by ETL), CE
Connection Drain Line1/2-in. I.D. (supplied)
Connection Sample Line1/4-in. O.D.
Cycle Time90 % in T<90sec
Flow14 L/h (200 - 250 mL/min) au-regulated by flow thru cell
InterferencesOzone
Lower Limit of Detection (LOD)Minimum 10 ppm or 0.01 mg/L ClO2
Maintenance IntervalMeasurement Cell: 6 months for membrane and electrolyte, typical
MaterialElectrodes: gold cathode/silver anode
Measuring PrincipleAmperometric/Membrane (electrode, membrane, electrolyte)
MountingFlat, vertical surface (panel, stand, etc.)
Operating Temperature Range0 - 45 °C
Pressure Range0.1 to 2 bar in flow cell
Range02 mg/L (mg/L) as chlorine dioxide (ClO2) 2 ppm
Range- (0.010 mg/L -2 mg/L)
Relative Humidity3 - 90 % non-condensing
Response Time90 % in less than 90 seconds
Sample Input0.25 " outer diameter
Sample Output0.5 " inner diameter
Sample Temperature2 - 45°C (35.6-113 F)
Ship Wt. (lbs)6.5 kg(14.3 lbs.)
Standard deviation1.5
Temperature CompensationAutomatic over sample temperature range
Temperature CompensationAutomatic over sample temperature range
Weight6.5 kg
What's included?9187sc Chlorine Dioxide Sensor, 9187sc User Manual, sc100 Controller, sc100 User Manual, sc100 Mounting Panel