Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

9525.99.10CCDG

9525.99.10CCDG
Product family: 9525sc Degassed Conductivity System
Hach 9525sc DCCP System with Specific Conductivity (SC), Cation Conductivity (CC), Calculated pH, Degassed Cation Conductivity
HACH 9525.99.10CCDG
This upgraded version of our degassed cation conductivity panel (DCCP) system measures specific conductivity (SC), cation conductivity (CC), and calculated pH in addition to degassed cation conductivi

产品概述

Simple to Integrate. Simple to Operate.

This upgraded version of our degassed cation conductivity panel (DCCP) system measures specific conductivity (SC), cation conductivity (CC), and calculated pH in addition to degassed cation conductivity (DCC) after CO₂ removal from the sample stream. This comprehensive solution saves you time on design, installation, training, maintenance, operation, and plant start-up, while helping protect your most critical assets and equipment.

- Designed with operator safety in mind - Compact, wall-mounted design - Operates in accordance with ASTM D4519

主要特点

  • Reliable Measurements:A new way of monitoring based on 9523sc + DCCP calculated pH, delivering reliable results that save you critical time and effort.
  • Save Time on Maintenance and Set-Up:The degassed conductivity system is easy to install and requires low annual maintenance as well as easy set-up with cation conductivity and specific conductivity instrumentation.
  • Flexible and Compact Design:The 9523sc cation conductivity analyzer can be combined with the degassed conductivity system in a single panel or separated in order to meet the need of each customer.
  • Reduce Training Complexity:A single platform minimizes time required to teach and learn product operations, getting new systems in use faster.

技术参数

参数名称说明
Accuracy< 2%
Ambient Temperature2 to 50 °C (36 to 122 °F)
Cell Constant k0.01 cm-1
Enclosure RatingIP66 / NEMA 4X
Flow100 - 150 cc/min (mL/min); 6 - 9 L/h
Includes 9523scYes
Measuring Range Conductivity0.01 - 200 µS/cm
Operating Temperature RangeRecommended operating temperature: 23 - 27 °C (75 - 80 °F)
ParameterDegassed Cation Conductivity, Specific Conductivity, Cation Conductivity, Calculated pH
Power Requirements (Hz)60 Hz
Power Requirements (Voltage)110 - 120 VAC
Regenerative CoolingNo
Storage Conditions0 - 50 °C (32 - 120 °F) 0 - 85 % relative humidity, non-condensing
Warranty12 months
Weight50 kg (110 lbs)
What's included?Kit includes (1 ea.): 9525.99.0010 9525sc Degas System, 9523.99.09P2 Hach 9523sc Specific and Cationic Conductivity Analyzer, and pH Calculator, with 5x 4-20 mA Outputs, 100-240 V AC, 9525.99.7040 Degas System Extension Panel Kit

同系列其他配置

型号
9525.99.0010DGHach 9525sc DCCP System, Degassed Conductivity only
9525.99.0011Hach 9525sc DCCP System, Degassed Conductivity only, with Regenerative
9525.99.10CCRHach 9525sc DCCP System with Specific Conductivity (SC), Cation Conduc

资料下载