Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

D3708E2T

D3708E2T
Product family: 3700 Digital Inductive Conductivity Sensors
Digital Electrodeless Conductivity Sensor, Sanitary Body Style, PFA Body Material
HACH D3708E2T
PFA, Sanitary (CIP) Flange Mounting Style, 6m (20 ft) Analog Cable, Electrodeless Conductivity Sensor, includes 6120800 Digital Gateway for Connecting to sc Controller and a 6122400 1m (3 ft) Digital

产品概述

PFA, Sanitary (CIP) Flange Mounting Style, 6m (20 ft) Analog Cable, Electrodeless Conductivity Sensor, includes 6120800 Digital Gateway for Connecting to sc Controller and a 6122400 1m (3 ft) Digital Extension Cable

主要特点

  • Wide Measuring Range:Hach’s Inductive Conductivity Sensors measure 200 up to 2,000,000 microSiemens/cm. A built-in Pt 1000 RTD compensates the measured conductivity for changes in process temperature.
  • Low Maintenance Design:The inductive sensor design eliminates polarization and electrode coating problems that commonly affect conventional contacting electrode-type conductivity sensors.
  • Versatile Mounting Styles:Sensors can be installed using a choice of four mounting styles—immersion, insertion, union, and sanitary.
  • Principal of Operation:Inductive conductivity sensors induce a low current in a closed loop of solution, then measure the magnitude of this current to determine the solution’s conductivity. The conductivity analyzer drives Toroid A, inducing an alternating current in the solution. This current signal flows in a closed loop through the sensor bore and surrounding solution. Toroid B senses the magnitude of the induced current which is proportional to the conductance of the solution. The analyzer processes this signal and displays the corresponding reading.
  • Withstands Harsh Environments:The inductive sensor is available in sanitary (CIP) flange style and convertible styles in PFA®, polypropylene, PEEK®, and PVDF material. Select sensors can withstand high pressures and temperatures.

技术参数

参数名称说明
Accuracy± 0.01 % of reading, all ranges
Body MaterialPFA
Cable Length22 ft
FlowRate 3 m (10 ft.) per second, maximum
MaterialPFA
MountingSanitary
Operating Temperature Range-10 - 200 °C
Range200 µS/cm - 2000 mS/cm
Sensor CablePolypropylene and PVDF Sensors: 5 conductor (plus two isolated shields) cable with XLPE (cross-linked polyethylene) jacket; rated to 150 °C (302ºF); 6 m (20 ft.) long
Sensor TypeDigital
Temperature SensorTemperature Compensator Pt 1000 RTD
Warranty12 months
Wetted MaterialsPolypropylene, PVDF, PEEK® or PFA®
What's included?Includes: sensor with cable, digital gateway, extension cable and manual

应用领域

Membrane Protection · Nutrient Monitoring · Nutrient Monitoring · Secondary Treatment · Secondary Treatment · Ultrapure Water

同系列其他配置

型号
D3727E2TDigital Convertible Mounting Style Inductive (Electrodeless) Conductiv
D3705E2TDigital Sanitary Inductive (Electrodeless) Conductivity Sensor with Po
D3706E2TDigital Sanitary Inductive, Electrodeless Conductivity Sensor for Low
D3725E2TDigital Inductive Conductivity Sensor, Convertible Body Style
D3726E2TDigital Electrodeless Conductivity Sensor, Convertible Body Style, PVD
D3728E2TDigital Convertible Mounting Style Inductive, Electrodeless Conductivi

资料下载

7 models · click to collapse