产品概述
The smart choice for accurate and reliable online process pH measurement
Hach’s patented differential electrode measurement technique ensures more accurate pH readings, providing increased reliability to reduce downtime and maintenance for measuring pH over traditional pH sensors. Our DPD1P1 digital pH probe is Convertible Mount complete with a 10m cable and compatible with Hach SC controllers. This instrument connects to Claros, Hach’s innovative Water Intelligence System, enabling you to seamlessly connect and manage instruments, data, and process – anywhere, anytime. The result is greater confidence in your data and improved efficiency in your operations. To unlock the full potential of Claros, insist on Claros Enabled instruments.
主要特点
- Differential pH Sensor With Encapsulated Preamp: Using three electrodes versus the traditional two increases accuracy in this digital pH probe. Reference junction potential is reduced along with elimination of sensor ground loops.:This field-proven technique uses three electrodes instead of the two normally used in conventional pH sensors. Process and reference electrodes measure the pH differentially with respect to a third ground electrode. The end result is unsurpassed measurement accuracy, reduced reference junction potential, and elimination of sensor ground loops. These process pH sensors provide greater reliability, resulting in less downtime and maintenance.
- Extended pH Sensor Life: Replaceable salt bridge holds higher volumes of buffer which extends pH sensor life and protects the junction reference electrode.:The double junction salt bridge creates a barrier to contamination which minimizes the dilution of the internal standard cell solution. The result is lower maintenance needs and a longer time period between calibrations.
- Durable Material Build For Aggressive Solutions: PEEK ® is used in construction of the body for chemical compatibility with most process applications for in line pH monitoring.:The unique, replaceable salt bridge holds an extraordinary volume of buffer to extend the working life of the sensor by protecting the reference electrode from harsh process conditions. The salt bridge simply threads onto the end of the sensor if replacement is needed.
- Versatile Mounting Options: pH sensors are available in four mounting styles – convertible, insertion, immersion, and sanitary for real time pH monitoring. The DPD1P1 is convertible build. Order our digital pHD today!:Encapsulated construction protects the sensor’s built-in preamp from moisture and humidity, ensuring reliable sensor operation. The preamp in the pHD analog sensor produces a strong signal, enabling the sensor to be located up to 1000 m (3280 ft.) from the analyzer.
- Patented Technology:The former GLI, now a Hach Company brand, invented the Differential Electrode Technique for pH measurement in 1970. The pHD sensor series (U.S. Patent Number 6395158B1, dated May 28, 2002) takes this field-proven technology to a new level.
- Versatile Mounting Styles:Sensors are available in four mounting styles - convertible, insertion, immersion, and sanitary
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | ± 0.02 pH |
| Body material | PEEK |
| Cable Connection | Digital |
| Cable Length | 10m (33 ft.) |
| Calibration Method | Two point automatic, one point automatic, two point manual, one point manual |
| Communication | Modbus |
| Compliance | For non-hazardous and non-maritime applications only |
| Drift | 0.03 pH per 24 hours, non-cumulative |
| Electrode Type | General Purpose |
| Flow Rate | 3 m (10 ft.) per second, maximum |
| Length | 271.3 mm |
| Mounting | Convertible |
| Operating Temperature Range | SS pHD: 0 - 50 °C (32 - 122 °F) pHD, ORP Digital Sensors: 5 - 70 °C (23 - 158 °F) pHD, ORP Analog Sensors: -5 - 105 °C (23 - 221 °F) Before initial pH calibration, calibrate the temperature measurement when the sensor is in water or buffer which is at approximately the same temperature as the pH buffers (matches current recommendation) Please Note: When the sensor is placed into the application sample, if that sample is more than 10 °C (18 °F) different than the previous temperature/pH calibration, then it is recommended to recalibrate the temperature while the sensor is in the sample to maintain the ±0.5 °C (±0.9 °F) temperature accuracy specification. |
| Pressure Range | Maximum 10.7 bar. 6.9 bar for Digital Sensor at 70 °C, and 6.9 bar for Analog Sensor at 105 °C. |
| Range | -2.0 to 14.0 pH |
| Repeatability | ± 0.05 pH |
| Sensitivity | ± 0.01 pH |
| Sensor Cable | 10 m (33 ft.) polyurethane, 4-conductor cable with one shield, rated to 105°C (221°F) |
| Sensor Thread | 1" NPT |
| Storage Conditions | 4 - 70 °C, 0 - 95% relative humidity (non-condensing) |
| Temperature Accuracy | ± 0.5 °C (± 0.9 °F) |
| Temperature Compensation | Automatic with NTC 300 Ω thermistor, or manually fixed at a user-entered temperature, additional selectable temperature correction factors (ammonia, morpholine, or user-defined pH/°C linear slope) available for pure water automatic compensation 0.0 - 50 °C |
| Temperature Sensor | NTC 300 Ω thermistor for automatic temperature compensation and analyzer temperature readout |
| Transmission Distance | 100 meters (328 ft.), combined length of both senor and add-on cable |
| Warranty | 12 months |
| Weight | 0.316 kg |
| Wetted Materials | PEEK or PPS, salt bridge of matching material with PVDF junction, glass process electrode, titanium ground electrode, and FKM/FPM O-ring seals (pH sensor with optional HF-resistant glass process electrode has 316 stainless steel ground electrode, and perfluoroelastomer wetted O-rings; consult factory for other available wetted O-ring materials) |
应用领域
Membrane Protection · Nutrient Monitoring · Nutrient Monitoring · Secondary Treatment · Secondary Treatment · Ultrapure Water
同系列其他配置
| 型号 | |
|---|---|
| DPD1R1.1 | Digital pH Sensor, PPS, Convertible, Non-Hazardous Environment |
| DPD1P3.1 | Digital pH Sensor, PEEK, Convertible, Non-Hazardous Environment |
| DPD1R3.1 | Digital pH Sensor, PPS, Convertible, Non-Hazardous Environment |
| DPD2P1.1 | Digital pH Sensor, PEEK, Insertion, Non-Hazardous Environment |
| DPS1.1 | Digital pH Sensor, Stainless Steel, Non-Hazardous Environment |
资料下载
- Temperature Compensation with pH Measurement
- Complete Water Analysis for the UPSTREAM OIL & GAS INDUSTRY
- Complete Water Analysis for the Food Industry
- Differential pH Design Overcomes Common pH Sensor Challenges
- Water Analysis Support for Exhaust Gas Cleaning Systems
- Cyanide Waste Treatment. This note provides basic information about treatment of extremely toxic cyanide wastes commonly found in the plating industry.
- pH Control Systems. This bulletin describes six basic types of pH control systems and outlines the operating principles of each
- Measure pH & Conductivity-Metal Working-Plating Industries
Related Products
HACHRadiometer Analytical PHC3105-9 Combination pH Electrode (epoxy body, gel electrolyte, screw cap)
HACHRadiometer Analytical PHC3031-9 Combination pH Electrode for Penetration Applications (glass body, Ag/AgCl ref., screw cap)
Radiometer Analytical PHC4006-9 Combination Calomel pH Electrode (l=158 mm, d=6.5 mm, glass body, screw cap)
Radiometer Analytical PHG200-8 Glass pH Electrode for Microsamples (d=4 mm, BNC)
HACHRadiometer Analytical PHG201-8 Glass pH Electrode (d=7.5 mm, BNC)
HACHpH Electrode Storage Solution, 500 mL
HACH