Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

EZ1102.9700140T

EZ1102.9700140T
Product family: EZ sc Nutrients Analyzers
EZ1102sc Ammonium Colorimetric Analyzer, 1 stream, 4x mA Output
HACH EZ1102.9700140T
Reliable, Scalable Systems Designed for Peak Performance

产品概述

Reliable, Scalable Systems Designed for Peak Performance

The EZ1102sc Ammonium analyzer with SC4500 Controller offers precise, reliable ammonium monitoring for industrial and wastewater applications. The SC4500 simplifies installation, reduces extra components, and improves usability. Its streamlined design ensures accuracy and easy setup, helping operators optimize performance. Built for efficiency, compliance, and reduced downtime, this advanced analyzer integrates smoothly into existing systems, even in tough environments. Key Features Flexible Customisation The EZ1102sc Ammonium analyzer is designed to meet your needs. The base model supports one stream but can be configured for four or eight streams, with options for 4x or 8x mA outputs. It offers robust connectivity, including Modbus RTU, Modbus TCP/IP, Profibus DP, and Ethernet/IP, ensuring easy integration with your facility for efficient data flow and process optimization. Reliable and Easy to Use Built for precision and durability, the EZ1102sc Ammonium analyzer provides accurate ammonium measurements with minimal maintenance. Its user-friendly interface simplifies operation and reduces training time. Designed for tough industrial environments, this analyzer ensures reliable performance, letting operators focus on critical tasks.. Explore Tailored Solutions Every facility faces unique challenges, and the EZ1102sc Ammonium analyzer is designed to meet them head-on. Reach out today to discuss your specific needs and discover how this versatile solution can be customized to optimize your processes, improve efficiency, and ensure compliance with confidence.

主要特点

  • Trusted Equipment for Critical Operations:The EZ sc Series combines robust engineering with high-quality materials to ensure reliability 24/7. Designed for long-term performance, EZ sc Series minimizes breakdowns and supports sustained operational success.
  • Scalable Solutions for Growing Needs:Built to evolve alongside your needs, online analysers from the EZ sc Series offer adaptable parameters and easier integration. They ensure your facility remains future proof without requiring expensive upgrades. Easily adjust parameter testing ranges and expand sample streams to accommodate innovation and growth.
  • Real-Time Data for Faster Decisions:The EZ sc Series offers advanced connectivity and real-time monitoring, delivering immediate access to actionable insights. Facilities utilising this technology benefit from improved reaction times, empowering more informed decision-making and quicker responses during critical situations.
  • Simplified Troubleshooting with Expert Support:Built-in diagnostics combined with secure remote data access make the EZ sc Series simplify issue resolution. Supported by a highly responsive team of experts, it ensures uninterrupted functionality and minimises operational downtime. By addressing and identifying issues early, EZ sc Series of online analysers helps prevent costly disruptions and repairs while maintaining system efficiency.
  • Clear Workflows for Predictable Outcomes:Standardized tools and guided workflows ensure consistent implementation and operation. The EZ sc Series minimizes miscommunication, sets clear expectations, and reduces process inconsistencies, driving greater efficiency.

技术参数

参数名称说明
Ambient Temperature50 - 86°F (10 - 30 °C) ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analog OutputsActive 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard: 4, optional: 8
Automatic cleaningYes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
CalibrationAutomatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
CertificationsCE, ETL certified to UL and CSA safety standards, UKCA
Cycle TimeNo Dilution Default: 30 min. Continuous: 29 min. Internal Dilution Default: 32 min. Continuous: 32 min.
Demineralized WaterInternal dilution option: Rinse water & Dilution water
Digital OutputsProfibus DP or Modbus RTU
Dimensions (H x W x D)688 mm x 460 mm x 340 mm
DrainAtmospheric pressure, vented, min. Ø 32 mm
Earth ConnectionDry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument AirDry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
InterferencesAmino acids, hydrazine (H₄N₂) and urea (H₂NCONH₂). Large quantities of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar interfere. Possible negative interference from alkalinity ˃ 800 mg/L (as CaCO₃, Mg ˃ 10 mg/L)
Lower Limit of Detection (LOD)No Dilution 0.005 - 0.1 mg/L NH₃-N: 0.005 mg/L 0.01 - 0.25 mg/L NH₃-N: 0.01 mg/L 0.01 - 0.5 mg/L NH₃-N: 0.01 mg/L 0.025 - 1.0 mg/L NH₃-N: 0.025 mg/L Internal Dilution 0.125 - 5 mg/L NH₃‑N: 0.125 mg/L 0.25 - 10 mg/L NH₃‑N: 0.25 mg/L 1 - 25 mg/L NH₃‑N: 1 mg/L 1.25 - 50 mg/L NH₃‑N: 1.25 mg/L 2 - 75 mg/L NH₃‑N: 2 mg/L 2.5 - 100 mg/L NH₃‑N: 2.5 mg/L
MaterialHinged part: Thermoform ABS, door: PMMA Wall section: Galvanized steel, powder coated
Measurement methodColorimetric measurement, Adapted method, based on standard method APHA4500-NH3 F
Measurement MethodColorimetric
Number of sample streamsStandard: 1 Optional accessory for 2, 4, or 8 sample streams
OutputsOptional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP
ParameterAmmonium
Power100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA
PrecisionBetter than 2% full scale range for standard test solutions
Protection ClassIP44
RangeNo Dilution 0.005 - 0.1 mg/L NH₃-N 0.01 - 0.25 mg/L NH₃-N 0.01 - 0.5 mg/L NH₃-N 0.025 - 1.0 mg/L NH₃-N Internal Dilution 0.125 - 5 mg/L NH₃‑N 0.25 - 10 mg/L NH₃‑N 1 - 25 mg/L NH₃‑N 1.25 - 50 mg/L NH₃‑N 2 - 75 mg/L NH₃‑N 2.5 - 100 mg/L NH₃‑N
Reagent RequirementsKeep between 50 - 86 °F (10 - 30 °C)
Relays5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)
Sample Flow Rate100 - 300 mL/min
Sample PressureBy external overflow vessel (open to atmospheric pressure)
Sample QualityMaximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature50 - 86 °F (10 - 30 °C)
ValidationAutomatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Warranty12 months
WeightMax. 77 lb (35 kg)
What's included?EZ1102sc Ammonium (Berthelot Method) Analyzer, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 2 x empty 2.5L Glass Reagent Containers with Fittings (for Reagent A & Reagent B Solution)

应用领域

Municipal Water

同系列其他配置

型号
EZ1028.9700140TEZ1028sc Nitrite Colorimetric Analyzer, 1 stream, 4x mA Output
EZ1031.9700140TEZ1031sc Phosphate Colorimetric Analyzer, 1 stream, 4x mA Output
EZ1032.9700140TEZ1032sc Phosphate Colorimetric Analyzer, 1 stream, 4x mA Output
EZ2732.970B140TEZ2732sc Total Phosphorus Colorimetric Analyzer, 1 stream, 4x mA Outpu
EZ1029.9750140TEZ1029sc Nitrate Colorimetric Analyzer, 1 stream, 4x mA Output
EZ2729.97VB140TEZ2729sc Total Nitrogen Colorimetric Analyzer, 1 stream, 4x mA Output

资料下载

7 models · click to collapse