产品概述
Online, automatic titration of Chloride in water monitoring applications
State of the art titration platform The EZ4000 Series Analyzers are single-parameter titrators built on an industrial analytical platform. High precision dispensers, robust peristaltic pumps and carefully designed liquid pathways all add up to the highest performance for industrial and environmental analysis needs. Unique flexibility in titration methods Every application starts with the basics: the right titrimetric technique for the parameter of interest, the measuring range and the water matrix. Depending on the change in the specific variables, the EZ4000 Series run either acid-base, redox, precipitation or photocolorimetric titration allowing for a unique flexibility. The EZ4006 Chloride Analyzers use potentiometric titration. They combine unparalleled expertise in online titration with a set of unique analysis, control and communication features in a compact footprint: - Wide standard measuring range of 0-500 mg/L Chloride - Smart automatic features - Control and communication via industrial panel PC - Analogue and digital output options - Multiple stream analysis (up to 8 streams) There are many additional options available. Please contact Hach for more details.
技术参数
| 参数名称 | 说明 |
|---|---|
| Alarm | 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts |
| Ambient Temperature | 10 - 30 °C ± 4 °C deviation (50 - 86 °F ± 7.2 °F deviation) at 5 - 95% relative humidity (non-condensing) |
| Analog Outputs | Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option) |
| Calibration | N.A. |
| Certifications | CE compliant / UL certified |
| Cycle Time | 10 - 15 minutes |
| Demineralized Water | For rinsing / dilution |
| Digital Outputs | Optional: Modbus (TCP/IP, RS485) |
| Dimensions (H x W x D) | 690 mm x 465 mm x 330 mm |
| Drain | Atmospheric pressure, vented, min. Ø 32 mm |
| Earth Connection | Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |
| Instrument Air | Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air |
| Interferences | Bromide, sulfide, iodide ions may interfere. Ferricyanide causes high results and must be removed. Chromate and dichromate interfere and should be reduced to chromic state or removed. Ferric iron interferes if present in an amount substantially higher than the amount of chloride. Chromic ion, ferrous ion and phosphate do not interfere. Fats, oil, proteins, surfactants and tar. |
| Lower Limit of Detection (LOD) | 5 mg/L |
| Material | Hinged part: Thermoform ABS, door: PMMA Wall section: Galvanized steel, powder coated |
| Measurement Method | Potentiometric titration with silver nitrate (AgNO₃), conform with standard method APHA 4500-Cl (D) |
| Method Name | Titrimetric |
| Model | EZ4006 |
| Number of sample streams | 1, 2, 4 or 8 |
| Parameter | Chloride |
| Power | 100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA |
| Power Supply | 100 - 240 VAC, 50/60 Hz |
| Precision | Better than 2% full scale range for standard test solutions |
| Protection Class | Protection class: Analyser cabinet: IP44 / Panel PC: IP65 |
| Range | 25 - 500 mg/L Cl- Optional: 5 - 50 mg/L Cl- 5 - 125 mg/L Cl- 5 - 250 mg/L Cl- Internal dispenser dilution (max. factor 100) |
| Range | 25 ‐ 500 mg/L |
| Reagent Requirements | Keep between 10 - 30 °C (50 - 86 °F) |
| Sample Flow Rate | 100 - 300 mL/min |
| Sample Pressure | By external overflow vessel |
| Sample Quality | Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU |
| Sample Temperature | 10 - 30 °C (50 - 86 °F) |
| Validation | Automatic; frequency freely programmable |
| Warranty | 12 months |
| Weight | 25 kg (55 lbs.) |
| What's included? | EZ4006 Chloride Analyzer, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x mV electrode, 2 x empty 10L Reagent Containers with Fittings (for Cleaning & Buffer Solution) and 1 x empty 10L Reagent Dark Container with Fittings (for AgNO₃ Solution) |
应用领域
Municipal Water · Ultrapure Water
资料下载
- EZ Series Analyzer Service
- Complete Water Analysis Solutions - Midstream and Liquefied Natural Gas Production
- Care Advantage Flyer : Complete service solution for all your instruments
- Care Advantage Brochure : Complete service solution for all your instruments
- Analyzer Seri EZ Online Solusi Lengkap
- EZ Series Online Analyzers
- Analizadores en continuo de la serie EZ
- Analisadores on-line da série EZ
Related Products
HACHNP6000sc Phosphate Analyzers
HACHSC1000 relay module for DIN-rail mounting, 4 Relays, max. 240V
Stainless Steel Chain Mount Kit
Mercury Thermometer, Celsius, -10 to 200°C
Mercury Thermometer, Fahrenheit, -30 to 120°F
COVER-SERVICE REPLACEMENT
HACH