Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

LXV434.99.00001

LXV434.99.00001
9187sc Chlorine Dioxide Amperometric Sensor
HACH LXV434.99.00001
Smart amperometric Chlorine Dioxide Analyzer: 9187sc

产品概述

Smart amperometric Chlorine Dioxide Analyzer: 9187sc

The 9187 sc Chlorine Dioxide (ClO2) Sensor is an online, single-channel industrial sensor that measures ClO2 in drinking water treatment plants, distribution networks, and other applications that require monitoring ClO2 at the ppb and ppm levels. This instrument uses an amperometric method to measure ClO2 concentration. A membrane allows the selective diffusion of ClO2 molecules to the amperometric sensor. The measurement is compensated for temperature.

主要特点

  • Selective Membrane Avoids Interference from other species:The Hach 9187sc Amperometric Chlorine Dioxide sensor uses a membrane that is selective for chlorine dioxide species. There are no interferences from bromine, chlorine or hydrogen peroxide. The only potential interference is from ozone, which is rarely present in water treated with chlorine dioxide.
  • Easy Setup and Maintenance:The all-inclusive 9187sc ClO2 system comes preassembled on a panel. The membrane system of the sensor means there are no chemicals used for measurement. Maintenance is minimal and two years of typical maintenance items are included with the system.
  • Compatible with Hach's "Plug and Play" Digital Controllers:The 9187sc ClO2 sensor is designed to be used with Hach SC digital controllers. Set up is fast and easy, and the system is compatible with the full line of Hach digital sensors.
  • :
  • :

技术参数

参数名称说明
Height10.6 in
Width6.1 in
Accuracy5% or ±10 ppb ClO2, whichever is greater
Calibration Interval2 months
Calibration MethodZero Calibration: Electrically or with ozone-free water Calibration: Comparison of lab method with process sample
Enclosure RatingIP66 / NEMA 4X
Flow14 L/h (200 - 250 mL/min) auto-regulated by flow through cell
InterferencesOzone
Lower Limit of Detection (LOD)0.01 mg/L ClO2
Maintenance Interval6 months for membrane and electrolyte (measurement cell)
Measuring IntervalContinuous
Measuring PrincipleAmperometric/Membrane (electrode, membrane, electrolyte)
MountingFlat, vertical surface (panel, stand, etc.)
Operating Temperature Range0 - 45 °C
ParameterChlorine Dioxide
Pressure Range0.1 - 2 bar in flow cell
Range0 - 2 mg/L ClO2
Relative Humidity0 - 90% non-condensing
Response Time90% in less than 90 seconds
Sample Input0.25 " outer diameter
Sample Output0.5 " inner diameter
Sample Temperature2 - 45 °C (35.6 - 113 °F)
Standard deviation1.5
Storage Conditions-20 - 60 °C (-4 - 140 °F)
Temperature CompensationAutomatic over sample temperature range
Temperature CompensationAutomatic over sample temperature range
Warranty12 months
Weight6.5 kg (14.33 lb)
What's included?Hach 9187sc Chlorine Dioxide sensor panel, 4 spare membranes, and Manual.

应用领域

Discharge · Groundwater · Influent for Wastewater · Membrane Protection · Municipal Water · Regulations · Surface Water

资料下载