产品概述
Turbidity and suspended solids Immersion probe specially designed for use in aggressive fluids in the industrial sector
The TSS Titanium2 sc is a digital process probe for the extremely accurate measurement of turbidity and suspended solids specially designed for use in aggressive fluids (e.g. organic and inorganic acids, lyes and chloride-containing solutions). This immersion probe is installed by tank-side mounting and can measure both on-line suspended solids and turbidity in one instrument. The two channel 90° scattered light turbidity method is in accordance with DIN EN ISO 7027. For the determination of suspended solids a scattered light measurement method at 120 ° is integrated. This eight channel measurement system, with an integrated bubble and temperature compensating software, enables the instrument to have a wide measuring range. The TSS Titanium2 sc probes are compatible with the Hach SC200 and SC1000 controller platform. Furthermore, they are made of titanium with a scratch resistant and easy to clean sapphire window design to withstand harsh environments. Please contact Hach for further information. Request for a quote today!
主要特点
- Unique Multi-Beam Pulsed Infrared Light System:TSS Titanium2 sc probes have a double optical system with two pulsating infrared LEDs and four receivers. This combined multiple beam alternating light method with beam focusing facilitates accurate color independent measurement of turbidity and suspended solids.
- Automatic Compensation for Air Bubbles and Temperature Swings for Accuracy:Special software enables the system to recognize gas bubbles or temperature swings for a more accurate suspended solids or turbidity measurement.
- Standard Compliant and Precise - Ease of Calibration:The turbidity measurement complies with the standard ISO 7027. Calibration is not necessary. If the probe is used to measure the solids content, a one-point calibration suffices. For special applications, curves can be defined using several calibration points.
- Titanium Probe Designed for the Most Difficult Environments:While offering the same high quality measuring properties, its special titanium body enables it to perform in aggressive fluids in the industrial sector
- Compatible with Hach SC controller platform:Measurement data is displayed and processed with the help of a digital Hach SC200 or SC1000 controller.
- Two parameters in one instrument--Suspended Solids and Turbidity:
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | Turbidity up to 1000 NTU: < 5% of the measured value ±0.01 NTU |
| Ambient Temperature | 0 - 60 °C (briefly 80 °C) |
| Application | Aggressive media |
| Automatic Wiper | No |
| Cable Length | 10 m (optional extension cables available) |
| Cable Length | 10 m (optional extension cables available) |
| Calibration | Turbidity (TRB): Factory calibrated Solids (TS): To be calibrated by customer on site Zero point: Permanently calibrated in the factory |
| Calibration Method | Turbidity (TRB): Factory calibrated Solids (TS): To be calibrated by customer on site Zero point: Permanently calibrated in the factory |
| Controller Compatibility | SC200, SC1000, SC4500. All controllers sold separately |
| Diameter | 40 mm |
| Flow | Max. 3 m/s (the presence of air bubbles affects the measurement) |
| Includes | Turbidity & Suspended Solids sensor, user manual |
| Installation style | User Defined |
| Length | Tank sensor: 330 mm |
| Maintenance Interval | 1 h/month |
| Material | Body and Optics: Ti2 (3.7035, UNS R50400), gasket: FKM |
| Max Temperature | 60 |
| Measurement Method | Combined multiple beam alternating light method with infrared diode system and beam focusing. Turbidity (TRB): 2-channel 90º scattered light measurement in accordance with DIN EN ISO 7027, wavelength = 860 nm Solids (TSS): 90º and 120º scattered light measurement, eight channel multi-angle measurement system, with an integrated bubble and temperature compensating, wavelength = 860 nm |
| Model | TSS Titanium sc |
| Mounting Configurations | Immersion |
| Operating Temperature Range | 0 - 60 °C |
| Parameter | Turbidity, Suspended Solids |
| Pressure Range | |
| Range | Turbidity (TRB): 0.001 - 9999 NTU Solids (TSS): 0.001 - 500 g/L |
| Repeatability | TSS content: < 4 % Turbidity: < 3 % |
| Response Time | 1 s < T90 < 300 s (adjustable) |
| Response Time T90 | 1 - 300 s adjustable |
| Special Notes | Installation note: Distance sensor–wall > 50 cm (Turbidity) > 10 cm (TSS) |
| Warranty | 12 months |
| Weight | approx. 1.6 kg |
| What's included? | Turbidity & Suspended Solids sensor, user manual |
应用领域
Pre-Treatment
同系列其他配置
| 型号 | |
|---|---|
| LXV329.99.20002 | TSS TITANIUM2 sc Turbidity and Suspended Solids TriClamp Insertion Pro |
| LXV330.99.20002 | TSS TITANIUM7 sc Turbidity and Suspended Solids TriClamp Insertion Pro |
| LXV330.99.10002 | TSS TITANIUM7 sc Turbidity and Suspended Solids Immersion Probe, Titan |
资料下载
- Case Study: SOLITAX Sensors, LDO Probes; Accurate On-Line DO and TSS Monitoring Helps Optimize Petrochem Effluent Treatment
- Beer Recovery from Surplus Yeast Using the TSS Vari sc by Hach
- Measuring Turbidity and Suspended Solids in Power Plants
- Introduction to Hach Services
- Complete Water Analysis Solutions : Upstream and Oil & Gas Industry
- Complete Water Analysis Solutions - Midstream and Liquefied Natural Gas Production
- Care Advantage Flyer : Complete service solution for all your instruments
- Care Advantage Brochure : Complete service solution for all your instruments
Related Products
HACHSonatax sc Sensor
HACHGelex Secondary Turbidity Standard, Stray Light
HACHGasket for mounting flange, Solitax sc, TSS sc
HACHWelded branch DN125 for TSS sc TriClamp
HACHBench Service Partnership, TSS Portable, 1 Service/Year
HACHTurbidity Standard, Stablcal®, 400.0 NTU, 500 mL
HACHWiper Set TSS sc
HACH