Overview
The HI-1332D is a gel filled, plastic body pH electrode with double reference junction that can be used with most hand-held pH/mV meters. Ideal for Chemicals Field applications Quality control testing Supplied fitted with DIN connector 1m cable Tips for the most accurate measurements Keep electrode hydrated Ideally, pH electrodes should be kept in a storage solution when not in use. Placing the electrode in a small glass filled with storage solution is suitable.
An option for pocket meters is to place a small piece of sponge into the meter’s cap and pour storage solution (HI-70300) into the cap to wet the sponge. Alternatively, purchase HI-740200 specialised electrode caps for storage. Pouring off any excess solution beforehand, the cap can then be placed on the meter.
Clean electrodes before use Clean the junction of your electrodes once a day or at least once a week to prevent junction clogging and to maintain accuracy. Immerse the electrode in the proper cleaning solution for at least 15 to 20 minutes. Hanna offers a wide range of cleaning solutions for general purpose and specific applications.
Replace electrode once a year If your electrode takes too long to stabilise a reading, or readings fluctuate wildly, it is most likely time to replace the electrode. The typical life span of any pH electrode is from 6 months to 1.5 years.
Specifications
| Parameter | Value |
|---|---|
| SKU | HI1332D |
| Sku | HI-1332B, HI-1332P, HI-1332D |
| Description | pH electrode |
| Reference | double, Ag/AgCl |
| Junction/flow rate | ceramic, single / 15-20 µL/h |
| Electrolyte | KCl3.5M |
| Max Pressure | 0.1 bar |
| Range | pH: 1 to 12 |
| Recommended Operating Temp. | 0 to 70°C |
| Glass Type | LT (Low Temperature) |
| Tip/Shape | Spheric (dia: 7.5mm) |
| Temperature Sensor | No |
| Amplifier | No |
| Body Material | PEI |
| Cable | Coaxial; 1m |
| Recommended Use | Chemicals, field applications, quality control, aquariums |
| Connection | HI-1332B BNC HI-1332P BNC + pin*, HI-1332D DIN |
Accessories
- HI-7004L: HI-7004L pH 4.01 Buffer Solution, 500 mL bottle
- HI-7007L: HI-7007L pH 7.01 Buffer Solution, 500 mL bottle HI7007
- HI-7010L: HI-7010L pH 10.01 Buffer Solution, 500 mL bottle
- HI-70300L: HI-70300L pH Electrode Storage Solution, 500 mL
- HI-7061L: HI-7061L Electrode Cleaning Solution - General Purpose, 500mL
Related Products
HANNA InstrumentsSpare pH module for HI-7609829/HI-7629829 series of probes
HANNA InstrumentsT-Type Glass Body pH Electrode
HANNA InstrumentsAmpHel® ORP Electrode with Replaceable Battery
HANNA InstrumentsDirect Soil Probe
HANNA InstrumentspH Electrode with Ground Sleeve
HANNA InstrumentspH EC TDS and Temperature Electrode
HANNA InstrumentspH electrode with PTFE junction. High temperature glass bulb
HANNA Instruments